T-02 Validation plan

Version 1. Topic: SOPs and plans.

Template v1, created by Agena Bio. Not a manufacturer document. It contains no acceptance criteria or procedures. Take values from your current manufacturer documents and your own validation. General guidance, not regulatory advice.

A plan outline for an assay validation. It names the questions to answer and leaves every value for your laboratory to set before the study starts.

Version

VersionDateChange
12026-11-24First release.

Plan control

Plan sections

No.SectionPromptYour text or reference
1Purpose and scope as set by the laboratoryState whether the assay is for research use or laboratory-defined use, as set by the laboratory director, and what it covers.
2Regulatory and quality-system basisList the rules, standards and contracts your laboratory works under and which parts apply to this assay.
3RolesName who plans, performs, reviews and signs off each part of the study.
4Instrument qualificationRefer to the qualification records for each instrument, or state what is still open.
5Assay design and contentDescribe the content (targets and versions) and how it was chosen. Record any change from the manufacturer's protocol.
6Reference materials and sample panelList reference materials and samples, their sources, how their expected results are known, and why they suit this assay.
7Performance studiesFor each performance characteristic, state the design, the number of samples and runs you chose and why, and the criterion you set before starting.
8QC strategyState the controls on each plate, how they are trended, and what happens when one is not met.
9Data analysis, review and disposition of results under the laboratory's own regulatory frameworkDescribe analysis settings, review steps, manual review rules and disposition, in the terms your framework uses.
10Data integrity and securityDescribe access, backup, record retention and change tracking for study data.
11Deviations and revalidation triggersState how deviations are recorded and which changes require a new study or part of one.
12Summary report and sign-offState what the summary report contains and who signs it.

Instrument qualification (DQ, IQ, OQ, PQ)

StageSource of requirementMethodAcceptance criterion (set by the manufacturer's document or by the laboratory)Evidence fileDate and signerStatus
DQ
IQ
OQ
PQ

Performance characteristics: qualitative germline genotyping

Names only. Study design and criteria are set by the laboratory before the study starts.

CharacteristicStudy designAcceptance criterion (laboratory-set)ResultReviewer
Accuracy against a comparator method
Precision: repeatability
Precision: intermediate precision or reproducibility
Analytical specificity, including interfering substances
Input quantity range
Specimen types
Call rate and handling of no-calls
Contamination and carryover
Robustness to small changes in the procedure

Performance characteristics: quantitative somatic detection

Names only. Use this table where the assay detects variants at low allele fractions.

CharacteristicStudy designAcceptance criterion (laboratory-set)ResultReviewer
Analytical sensitivity (lowest level detected)
Accuracy against a comparator method
Precision: repeatability
Precision: intermediate precision or reproducibility
Analytical specificity and false-positive handling
Quantitation range (if the result is quantitative)
Input quantity range
Specimen types and preservation
Contamination and carryover

Reference materials

MaterialSourceLotWhere expected results come fromUse in the studyStorage

Acceptance-criteria worksheet

Fix each criterion and its basis before data collection. This template gives no numbers.

ItemMetricAcceptance criterionBasis for the criterionSet byDate

Ongoing QC and revalidation

ItemWhat is recordedFrequency (set by the laboratory)OwnerRevalidation trigger
Plate controls
Reference sample trend
Instrument performance verification
External assessment
Lot changes
Software or PC changes

Sign-off

RoleNameSignatureDate
Laboratory director
Quality lead
Reviewer

Statement

General guidance, not regulatory advice. This template gives no pass or fail numbers because limits are set by the laboratory director.

Where the structure comes from

License and file check

Take values from your manufacturer documents and your own validation.

Free to use inside your own organization; redistribution not permitted.

SHA-256 of the Markdown file: 5388831bd1c5560f372d8d0d9285355aaef200f16ceeff433580ff599a68e870