Workflow and data

Data output

What the software produces after a run, how to read calls and confidence, and how to keep your data records in order.

Checked on 24 November 2026

Synthetic data, not from a real run.

Overview

What the software produces after a run, how to read calls and confidence, and how to keep your data records in order.

After a run the software shows spectra, calls, plots and plate views. This section explains what each one is for, how laboratories review them and what to keep on record. Every figure here is synthetic or comes from a published study; none is from a real run.

In the public pages we checked on 24 November 2026, the manufacturer does not list real file types or report formats.

The manufacturer says analysis and reporting happen in Typer software, which shows qualitative results as Detected or Not Detected.[1]

In this section

Spectra

The most direct output is the spectrum for each spot: signal on the vertical axis, mass on the horizontal axis. A genotype is read from which peaks stand above the baseline and how tall they are. Published work shows spectra of relative intensity against mass as standard.[2]

The chemistry page explains where each peak comes from.

What genotype and failed-reaction spectra look likeFour schematic spectra: homozygous A with one allele peak, homozygous T-star with one peak 16 Da lower, heterozygous A and T-star with two peaks, and a failed or negative reaction with only the unextended primer peak.Synthetic data, not from a real run.a Homozygous A6,2006,3006,4006,500unextended primer 6,126.05A 6,423.26Homozygote: one extended peak.b Homozygous T*6,2006,3006,4006,500unextended primer 6,126.05T* 6,407.26Homozygote for the other allele: one extended peak, 16 Da lower.c Heterozygous A / T*6,2006,3006,4006,500unextended primer 6,126.05A 6,423.26T* 6,407.26Heterozygote: two extended peaks of similar height.d No extension (failed or negative control)6,2006,3006,4006,500unextended primer 6,126.05Failed or negative reaction: only the unextended primer.Schematic: spacing set at the published 16 Da minimum; the real mass-modified increments are not public. Heights are illustrative [I]; theprimer mass 6,126.05 Da is the computed example [K].

Four synthetic spectra: homozygous, homozygous, heterozygous, and no extension.

Text version of this figure

Each panel shows the unextended primer peak. Homozygous samples add one allele peak, a heterozygous sample adds two of similar height, and a failed reaction shows the primer and noise only.

Synthetic data, not from a real run.

Genotype calls and confidence

The software turns each spectrum into a call and attaches a confidence label. One published study reported five tiers from conservative to low probability, plus calls made by a person.[3] Low-frequency work uses signal-to-noise and z-score thresholds to separate calls from no-calls.[4]

Calls and confidence explains the tiers and lets you move a stringency slider on a synthetic data set.

Cluster plots

An allelic-discrimination or cluster plot puts one assay across many samples. Samples fall into tight groups by genotype; points between groups are judged by eye.[2]

Open the cluster plot explorer to click a point and see its schematic spectrum.

Synthetic cluster plot, 96 samples, one assayA scatter plot of 96 synthetic samples for one assay: three genotype clusters (AA, AB, BB), four in-between points and four failures near the origin. Axes are signal allele 1 and signal allele 2 in arbitrary units. Synthetic data, not from a real run.Synthetic data, not from a real run.000.50.511signal allele 1 (arbitrary units)signal allele 2 (arbitrary units)AAABBBfailuresAAABBBIn betweenFailed96 points, seed20261124

Synthetic cluster plot: 96 samples, one assay.

Text version of this figure

Three tight clusters (two homozygous, one heterozygous), a few points between them and a few failures near the origin.

Synthetic data, not from a real run.

Plate views

A plate view shows the whole plate at a glance, with each well tinted by the share of successful calls. A problem with a row, a column or a corner shows up as a pattern.

One laboratory's rule, not a specification Red under 15%, yellow from 15% to under 50%, light green from 50% to under 85%, dark green at 85% or more.[2]

Synthetic plate view tinted by share of successful callsA 96-well plate drawn as rows A to H and columns 1 to 12, each well tinted by share of successful calls in four bands: under 15 percent, 15 to under 50, 50 to under 85, and 85 percent or more. Bands also carry a letter and a pattern. One laboratory's scheme. Synthetic data, not from a real run.Share of successful calls per wellSynthetic data, not from a real run.123456789101112ABCDEFGHcddddddddddccadddddddddccdddddddddddcdddddddddddcdaddddddddccddddddddddccdddddddddddccdddcdcccabBandsaunder 15%b15% to under 50%c50% to under 85%d85% and aboveOne laboratory's scheme, not a manufacturer setting.Each well shows its band letter, so the plate reads without color.

Synthetic 96-well plate view in one laboratory's four color bands, with letters and patterns so color is not the only cue.

Text version of this figure

A 96-well grid. Most wells are in the highest band and a few are lower. The wells are coded by letter as well as by shade.

Synthetic data, not from a real run.

Mutation matrices

In one published UltraSEEK study the output is a table with one row per sample and one column per mutation. A cell holds T for a high-level call, L for a low-level call, or nothing.[4]

Synthetic mutation matrix: T high-level, L low-level, dash not called
SampleMutation 1Mutation 2Mutation 3Mutation 4Mutation 5Mutation 6
Sample AT---L-
Sample B------
Sample C-L-T--
Sample D-----L

Synthetic data, not from a real run.

Methylation grids

For methylation work the output is a percentage for each CpG unit in each sample: the methylated peak area divided by the methylated plus unmethylated area.[5]

Synthetic methylation grid: percent methylation per CpG unit
SampleCpG unit 1CpG unit 2CpG unit 3CpG unit 4CpG unit 5CpG unit 6
Sample A12%8%70%64%31%5%
Sample B15%10%22%18%29%7%
Sample C11%9%81%77%35%4%

Synthetic data, not from a real run.

Calls and confidence

How calls are made from spectra and how confidence shows up, with a synthetic cluster plot to explore.

A call is the software’s reading of one spectrum: which allele or alleles it sees, or that it sees none. Most software also says how sure it is. This page uses a published study to show what confidence tiers look like, then lets you change a stringency setting on synthetic data to see calls turn into no-calls.

The manufacturer says Typer software shows qualitative results as Detected or Not Detected.[1] For genotyping, the output is a call per sample and assay with a confidence label.

Confidence tiers in one published study

A peer-reviewed marker study using Typer reported genotype confidence in five tiers (conservative, moderate, aggressive, low probability, user call) and a group with no alleles. The table gives that study’s shares. It covers one species and one marker set and is not a performance specification.[3]

Read the tier names as an ordered ladder, from the calls the software is most sure of to those it is least sure of. The exact criteria for each tier belong to the software and your method, and are not stated here.

Tier names and shares of genotypes in one published study
Tier in that studyHow to read the nameShare of genotypes
ConservativeThe calls the software is most sure of.54%
ModerateStill confident, below the top tier.15%
AggressiveCalled on looser criteria than the two tiers above.9%
Low probabilityA call was made, and the software rates it as unlikely to be reliable.18%
No allelesNo allele signal was found.3%
User callA person made or confirmed the call after looking at the spectrum.1%
Genotype confidence tiers, from conservative to low probabilityA ladder of four call tiers: conservative, moderate, aggressive and low probability, with a user call side branch and a no-call outcome at the bottom. A bracket notes that average call rate was 90.0 percent by default and 69.9 percent with stricter filters in one study.Call tierstier names as used in one published studyConservativeModerateAggressiveLow probabilityUser callset by a personNo callStringency trade-off [C]Average call rate 90.0% by default versus69.9% with stricter filters, in oneduplicate-typing study. Not a performancespecification.

The tier ladder with the user-call branch and the no-call at the bottom. Tier names as used in one published study.

Text version of this figure

From top to bottom: conservative, moderate, aggressive, low probability, then no call. A side branch shows a user call made after review. An annotation gives the trade-off between a lenient and a strict setting as an average call rate of 90.0% against 69.9%.

What counts as a failed assay

The same study defined a failed assay as one where most genotypes were low-probability or where the samples did not cluster by genotype.[3] Both signs are visible on a cluster plot before any number is computed, which is why laboratories look at the plot as well as the call rate.

Low-frequency calls: one study’s rules

An UltraSEEK breast-panel study of 39 mutations in 5 genes used 10 ng of DNA and Typer v4.0.26.74. A mutant call needed a signal-to-noise ratio of at least 6 and a z-score of at least 7. A z-score from 7 to 10 was labeled L (low-level) and above 10 was labeled T (high-level). The output was a per-sample, per-mutation matrix of T and L codes, plus spectra.[4]

These are the authors’ rules for their panel, not manufacturer specifications. Your thresholds come from your own validation.

Low-frequency mutation calls from signal-to-noise and z-scoreThree mini spectra with the same baseline noise: not called, called at low level, and called at high level, with the published study's signal-to-noise and z-score rules and a five-step tagged-terminator inset.Synthetic data, not from a real run.Not calledS/N 3, z 4 [I]mass-to-chargeCalled, low level (L)S/N 8, z 8.5 [I]mass-to-chargeCalled, high level (T)S/N 20, z 14 [I]mass-to-chargeIn one published study (39-mutation breast panel): a call needs signal-to-noise at least 6 and z-score atleast 7; z 7 to 10 is labeled L; above 10 is labeled T. These are that study's rules, not manufacturerspecifications.Chemistry behind the call: one capture-tagged terminator, thenTagged terminatorBead captureWashEluteSpot on chip

Three spectra with the same baseline: not called, called at low level, called at high level.

Text version of this figure

A spectrum with signal-to-noise 3 is not called. One with signal-to-noise 8 is called at low level (L). One with signal-to-noise 20 is called at high level (T).

Synthetic data, not from a real run.

The stringency trade-off

Stricter filtering removes doubtful calls and also removes some good ones. In a published study of an iPLEX identity panel, duplicate typing gave an average call rate of 90.0% with the default settings, falling to 69.9% when stricter filters removed every inconsistency. The same study found 10 ng or more of template optimal, a relatively high dropout risk at 1 ng or less, and lower call rates from degraded DNA.[11]

That is the trade you are making when you set a threshold: fewer wrong calls against more no-calls. The explorer below shows the same effect on synthetic points.

Cluster plot explorer

Move the stringency slider from permissive to strict and watch calls turn into no-calls. Points between clusters are called only at the most permissive settings, which is where a wrong call is most likely.

I-17Cluster plot explorer

96 synthetic points, one assay.

Synthetic cluster plot96 synthetic points for one assay in three genotype clusters, with a few in-between points and failures near the origin. 92 points are called at the current setting.000.250.250.50.50.750.7511signal allele 1 (arbitrary units)signal allele 2 (arbitrary units)

Synthetic data, not from a real run.

  • Called homozygous, allele 1
  • Called heterozygous
  • Called homozygous, allele 2
  • No call

Calls accepted down to low probability. 92 of 96 points are called; 4 are no-calls.

What happens to each group of points
GroupPointsCalledNo-call
Cluster 1 (homozygous, allele 1)28280
Cluster 2 (heterozygous)40400
Cluster 3 (homozygous, allele 2)20200
In-between points440
Failed, little signal404

Inspect a point

Click a point on the plot, or choose a sample number here.

Signal, allele 1 and 2
0.88 and 0.18
Nearest cluster
homozygous, allele 1
Tier (demonstration rule)
conservative
At this setting
Called
m/z, schematic

Schematic spectrum from the point’s two signals. Synthetic data, not from a real run.

No-call handling belongs to the SOP

What you do with a no-call, or with a low-probability call, is a decision for your laboratory: repeat the well, review the spectrum, or record it as not determined. Write that rule in the procedure before the run, not during review. See the run record and manual review pages.

QC metrics

Which quality metrics to track per run and per plate, with a worksheet that leaves every limit for you to set.

Quality control on a MassARRAY run asks two questions: did the run behave, and did each sample behave? Published validations describe the same kinds of check. No numeric limit is published as a standard, so each laboratory takes its limits from its own validation.[6]

This page lists the metrics, shows one laboratory’s working rules as an example of what a written rule looks like, and gives you a worksheet with every limit left blank.

Metrics laboratories compute

The SNPflow quality-control tool for iPLEX output computes call rate, expected against observed replicates, minor allele frequency, discordant replicates, discordance rate and a Hardy-Weinberg test, and has a sex-determination module. It sets no numeric cut-offs.[7]

Quality metrics and where each is described
MetricWhat it asksWhere it is described
Call rate per sampleHow many assays gave a call for each sample.SNPflow[7]
Call rate per assayHow many samples gave a call for each assay.SNPflow[7]
Negative-control extensionWhether the no-template wells show signal.Published validations[6]
Reference sample concordanceWhether a sample of known genotype gives its known result.Published validations[6]
Replicate concordanceWhether duplicates of the same sample agree.SNPflow, published validations[7][6]
Hardy-Weinberg testWhether genotype frequencies fit expectation, where the study design supports the test.SNPflow[7]
Identity or sex markerWhether the sample is the one you meant.SNPflow, published validations[7][6]

One laboratory’s working rules

  • The plate positive control matches the consensus genotype in all assays.
  • No-template-control extension is under 10%, with at least 4 of 5 no-template wells in the lowest success band.
  • Sample wells fall in the upper success bands, with at most one marginal well per plate.
  • The assay success target is at least 85%.
  • Calls with more than 75% unextended primer are rejected.
  • Very small peaks, under 5 intensity units, are kept only if isolated and well shaped.
  • Every allelic-discrimination plot is reviewed.

The same validation reported an average mass resolution of about 850 (m/Δm, full width at half maximum), which is the width we use for the illustrative peaks on the chemistry page.[13]

Plate view

A plate view tints each well by the share of successful calls so that row, column and edge problems show as patterns. One laboratory's rule, not a specification Red is under 15%, yellow 15% to under 50%, light green 50% to under 85% and dark green 85% or more.[2]

Synthetic plate view tinted by share of successful callsA 96-well plate drawn as rows A to H and columns 1 to 12, each well tinted by share of successful calls in four bands: under 15 percent, 15 to under 50, 50 to under 85, and 85 percent or more. Bands also carry a letter and a pattern. One laboratory's scheme. Synthetic data, not from a real run.Share of successful calls per wellSynthetic data, not from a real run.123456789101112ABCDEFGHcddddddddddccadddddddddccdddddddddddcdddddddddddcdaddddddddccddddddddddccdddddddddddccdddcdcccabBandsaunder 15%b15% to under 50%c50% to under 85%d85% and aboveOne laboratory's scheme, not a manufacturer setting.Each well shows its band letter, so the plate reads without color.

Synthetic 96-well plate view in one laboratory's four bands, coded with letters and patterns as well as color.

Text version of this figure

A 96-well grid with most wells in the top band and a few wells, grouped at one edge, in lower bands.

Synthetic data, not from a real run.

Methylation examples

EpiTYPER examples in the literature use triplicate measurements (126 per 384-well plate), a replicate standard deviation cut-off of 0.1 and 80% success across CpG units. The default analysis window is 1.5 to 7 kDa. The window and the thresholds are laboratory or software defaults, not manufacturer limits.[14]

Daily controls where CLIA applies

In the United States, 42 CFR 493.1256 sets daily minimum controls: one positive and one negative for qualitative tests, two controls for tests with an extraction step (one able to detect extraction errors), and for molecular amplification two controls plus an inhibition control where inhibition is a significant cause of false negatives.[15]

Whether that applies to you depends on your regulatory framework. It is a floor for some laboratories, not a description of what a validation needs.

General information; not legal, regulatory or clinical advice.

QC worksheet

Fill in your own limit for each metric before the run, enter what you observe, and record the result. Print it, or export a CSV or Markdown file.

I-18QC worksheet

Limits are yours to set.

Use your own codes. Do not enter patient, sample or personal data.

Quality metrics for one run
MetricWhat it asksLimit set by my laboratoryObserved valueResultNotesRemove
Call rate per sampleShare of assays with a call for each sample
Call rate per assayShare of samples with a call for each assay
Negative-control (no-template) extensionSignal in the no-template wells
Reference sample concordanceCalls against the known genotype
Replicate concordanceAgreement between duplicates
Hardy-Weinberg checkGenotype frequencies against expectation, where your design supports it
Identity or sex markerMarker result against the record for the sample
Allelic-discrimination plots reviewedWhether every plot was looked at, and by whom

Manual review

When and how to review a call by hand, and what to record when you override the software.

Software makes the first call. A person still looks at the evidence behind it. This page explains why that remains normal practice in published MassARRAY work, what reviewers look at, and what a review record should hold.

Why manual review persists

A published validation of a multi-gene panel needed significant development time, modified the manufacturer’s protocol and reviewed every assay plate manually.[8]

A published iPLEX HS study of lung tumor samples found that some mutations were weakly called or not called by the software and were judged from the raw spectra.[9]

Cluster plots with tight bands and in-between points that are judged by hand are standard in published MassARRAY work.[2] Published validations also describe second-person review as part of typical quality control.[6]

The point is not that the software is poor. It is that the software gives a call and the laboratory owns the result.

What reviewers look at

Three views, used together. Each one catches a problem the others hide. All three below are synthetic.

What genotype and failed-reaction spectra look likeFour schematic spectra: homozygous A with one allele peak, homozygous T-star with one peak 16 Da lower, heterozygous A and T-star with two peaks, and a failed or negative reaction with only the unextended primer peak.Synthetic data, not from a real run.a Homozygous A6,2006,3006,4006,500unextended primer 6,126.05A 6,423.26Homozygote: one extended peak.b Homozygous T*6,2006,3006,4006,500unextended primer 6,126.05T* 6,407.26Homozygote for the other allele: one extended peak, 16 Da lower.c Heterozygous A / T*6,2006,3006,4006,500unextended primer 6,126.05A 6,423.26T* 6,407.26Heterozygote: two extended peaks of similar height.d No extension (failed or negative control)6,2006,3006,4006,500unextended primer 6,126.05Failed or negative reaction: only the unextended primer.Schematic: spacing set at the published 16 Da minimum; the real mass-modified increments are not public. Heights are illustrative [I]; theprimer mass 6,126.05 Da is the computed example [K].

Spectra: is there a real peak, at the right mass, above the noise?

Text version of this figure

Four small spectra: homozygous, homozygous, heterozygous and a failed reaction.

Synthetic data, not from a real run.

Synthetic cluster plot, 96 samples, one assayA scatter plot of 96 synthetic samples for one assay: three genotype clusters (AA, AB, BB), four in-between points and four failures near the origin. Axes are signal allele 1 and signal allele 2 in arbitrary units. Synthetic data, not from a real run.Synthetic data, not from a real run.000.50.511signal allele 1 (arbitrary units)signal allele 2 (arbitrary units)AAABBBfailuresAAABBBIn betweenFailed96 points, seed20261124

Cluster plot: does the sample sit inside a cluster, or between two?

Text version of this figure

Three clusters of points, a few points between them, and a few near the origin.

Synthetic data, not from a real run.

Synthetic plate view tinted by share of successful callsA 96-well plate drawn as rows A to H and columns 1 to 12, each well tinted by share of successful calls in four bands: under 15 percent, 15 to under 50, 50 to under 85, and 85 percent or more. Bands also carry a letter and a pattern. One laboratory's scheme. Synthetic data, not from a real run.Share of successful calls per wellSynthetic data, not from a real run.123456789101112ABCDEFGHcddddddddddccadddddddddccdddddddddddcdddddddddddcdaddddddddccddddddddddccdddddddddddccdddcdcccabBandsaunder 15%b15% to under 50%c50% to under 85%d85% and aboveOne laboratory's scheme, not a manufacturer setting.Each well shows its band letter, so the plate reads without color.

Plate view: is the problem one well, or a row, a column or an edge?

Text version of this figure

A 96-well grid with a few low-scoring wells grouped together.

Synthetic data, not from a real run.

To try this yourself, open the cluster plot explorer and click the points that sit between two clusters.

A review policy outline

An outline for your SOP. It sets no acceptance criteria and no counts: those come from your validation.

  1. Step 1: Say in the SOP what must be reviewed

    List the cases that always go to a person: every control, every low-confidence call, every no-call, and a plot review for every assay on every plate.

  2. Step 2: Give the reviewer the whole picture

    Spectrum, cluster plot and plate view side by side, with the plate map, so the reviewer can see a pattern and not only a point.

  3. Step 3: Decide from a short list

    Accept the software’s call, change it, or mark it as no-call. Anything that does not fit goes to the person your SOP names.

  4. Step 4: Record the decision where the next person will look

    Use the record fields in the table below. A change with no reason written down is the first thing anyone reviewing the record will ask about.

  5. Step 5: Get a second reviewer for the calls that changed

    A second person looks at every changed call and every control failure, and signs.

What to record when a call changes

Review record fields
RecordWhy it is on the record
Plate or run code, sample position, assaySo anyone can find the exact spectrum again.
The software's call and its confidence labelWhat the software said before a person touched it.
The reviewer's call, or no-callWhat was decided.
The reason, in a short coded list plus free textA reviewer a year later can tell a judgment from a slip.
Evidence referenceA saved spectrum or plot image or a file path, so the evidence travels with the decision.
Reviewer, second reviewer, dateReview by two people is the usual control on a human decision.
Software version and settings in useThe same data can read differently under different settings.

The records page has a checklist for keeping these files, and the run log template has a block for review and sign-off.

Typer software and records

Records and data-integrity habits around the Typer software, with a checklist for your own procedures.

The analysis software, the files it makes and the records around them are what a laboratory shows when someone asks “how do you know?” This page sorts out the names you will meet, says what this site does not state about the software, and gives you a checklist for your own record-keeping.

Typer naming and aliases

Papers and older documents print the analysis software under several names. This site treats them as aliases of one product line, so a reference to any of them in a record means the same family of software; the version still matters for your records.[10]

Names under which the analysis software is printed
Name as printedWhere it appears
TyperGeneral name for the analysis software.
TYPER 4.0Announced on 18 September 2007, alongside iPLEX Gold.
SpectroTYPER v4.0Printed in published papers.
Typer v4.0.26.74Printed in a published study.
MassARRAY Typer Analyzer v4.0.26.73Printed in published papers.
TyperAnalyzerShort form in published papers.

“Typer” is not on the manufacturer’s published list of marks, so this site writes it without a symbol and makes no statement about who owns the name.[16]

Interpretation software from other providers

The manufacturer has described a reporting network that pairs the mass spectrometer with interpretation software from other providers. The first was announced on 19 September 2017 and others were added later. The network’s status in 2026 is unknown.[17]

If your results flow into interpretation software, record its name and version in the run log beside the analysis software, as part of the same change control.

What this site does not state

This site makes no statement about these topics for any version of the software:

  • electronic-record and e-signature rules
  • audit trails
  • laboratory information system connections
  • export file formats
  • operating-system and PC requirements
  • software support dates

Ask the manufacturer for the documents that cover your software version, and keep a copy in your records. If it helps, ask us to confirm the answer with you at an installation or service visit.

We do not reproduce manufacturer manuals, procedures or error codes. Ask the manufacturer or use your customer portal.

Records and backups

A common test for a laboratory record asks whether it is attributable (who made it), legible, contemporaneous, original and accurate. Run data are no different from a paper notebook in that respect.

Data backup, networking and waste disposal are on the manufacturer’s list of customer responsibilities.[18] So the backup of the analysis computer is yours to plan: what is backed up, how often, where the copy lives, and who has tested a restore.

Keep the plate map, the run log and the result files together, and record the software version used. The run log template, the manual review record fields and the PC and OS change page all feed the same record.

Data-integrity checklist

An original prompt list for your own procedures. It is not a statement about any software and sets no requirement.

I-18Data-integrity checklist

For your own procedures.

0 of 12 checks assessed. A status of No is a finding to follow up, not a failure of this tool.

Records of the run
  • The plate map used at PCR is the same one used at spotting and in the analysis software.

  • Each run has a run log with reagent lots, instrument and software identifiers, operator and date.

  • Deviations and reruns are recorded with a reason and who decided.

Data files
  • Raw and result files are kept unchanged, apart from any edited or exported copies.

  • File names follow one written convention that ties each file to a plate and a run.

  • It is written down who may edit, move or delete run data, and accounts are not shared.

Review and change
  • Manual calls and overrides record who changed what, when and why.

  • A second person reviews results, as your SOP requires.

  • The software version and analysis settings used are recorded for each run, and changes go through change control.

Backup and retention
  • Backups of run data are made on a schedule your SOP sets, and a restore has been tested.

  • A retention period is defined by your policy and the rules that apply to you.

  • Archived runs can be found again and read years later, including the plate map and run log.

Sources

  1. [1]
    WFL-01

    Agena presents the MassARRAY System as an open platform in 96- and 384-well formats: end-point PCR with single-base extension, automated sample transfer and MALDI-TOF detection, then analysis and reporting in Typer software, which shows qualitative results as Detected / Not Detected.

    Agena Bioscience: MassARRAY System product page, agenabio.com/products/massarray-system/

    Checked on 24 November 2026Manufacturer-reportedmedium confidence

  2. [2]
    DAT-09

    Cluster (allelic-discrimination) plots with tight bands and in-between points judged manually, spectra of relative intensity against mass, and plate views tinted by share of successful calls are standard in published MassARRAY work. One lab's plate color bands: red under 15%, yellow 15 to under 50%, light green 50 to under 85%, dark green 85% or more.

    Journal of Applied Laboratory Medicine 2020;5(3):454-466, iPLEX PGx validation study (external)

    Checked on 24 November 2026Peer-reviewedmedium confidence

  3. [3]
    DAT-01

    A peer-reviewed marker study using Typer reported genotype confidence tiers (conservative, moderate, aggressive, low probability, user call): 54% conservative, 15% moderate, 9% aggressive, 18% low probability, 3% no alleles and 1% user calls. It defined a failed assay as one where most genotypes were low-probability or samples did not cluster by genotype. One species and marker set, not a performance specification.

    PLoS ONE genotyping study using Typer software (journal.pone.0049533) (external)

    Checked on 24 November 2026Peer-reviewedmedium confidence

  4. [4]
    DAT-05

    UltraSEEK breast panel study (39 mutations in 5 genes, 10 ng, Typer v4.0.26.74): a mutant call needed signal-to-noise of at least 6 and a z-score of at least 7; z 7-10 is labeled L (low-level) and above 10 T (high-level). Output is a per-sample, per-mutation matrix with T/L codes plus spectra of relative intensity against mass-to-charge. These are the authors' rules, not Agena specifications.

    UltraSEEK breast panel study (PubMed Central PMC8558178) (external)

    Checked on 24 November 2026Peer-reviewedmedium confidence

  5. [5]
    CHM-09

    Published EpiTYPER protocol: bisulfite conversion, PCR with a T7 promoter tag, shrimp alkaline phosphatase, T7 transcription, RNase A cleavage after U positions (T-reaction), a clean-up step, then MALDI-TOF. A methylated CpG unit gives a fragment about 16 Da heavier; percent methylation is the methylated peak area over methylated plus unmethylated area.

    Frontiers in Genetics (2015) EpiTYPER methylation protocol article, DOI 10.3389/fgene.2015.00287 (external)

    Checked on 24 November 2026Peer-reviewedmedium confidence

  6. [6]
    DAT-10

    Published validations describe typical QC practice: reserved negative-control wells, reference samples of known genotype, duplicates to check discordance, an identity or sex marker, and second-person review. Acceptance numbers come from each laboratory's own validation; no numeric limit is published as a standard.

    SNPflow quality-control article (PLoS ONE, DOI 10.1371/journal.pone.0059508); Journal of Applied Laboratory Medicine 2020;5(3):454-466 (external)

    Checked on 24 November 2026Peer-reviewedmedium confidence

  7. [7]
    WFL-11

    The SNPflow QC tool for iPLEX output computes call rate, expected versus observed replicates, minor allele frequency, discordant replicates, discordance rate and a Hardy-Weinberg test, checks genotype frequencies against reference data and has a sex-determination SNP module. It sets no numeric cut-offs.

    SNPflow: a lightweight application for the processing, storing and automatic quality checking of genotyping assays (PLoS ONE) (external)

    Checked on 24 November 2026Peer-reviewedmedium confidence

  8. [8]
    DAT-04

    A clinical laboratory's published account of validating a pharmacogenomic panel (J Appl Lab Med 2020;5(3):454-466). It needed significant development time, modified the manufacturer's protocol and reviewed every assay plate manually.

    Journal of Applied Laboratory Medicine 2020;5(3):454-466 (PubMed 32445367) (external)

    Checked on 24 November 2026Peer-reviewedmedium confidence

  9. [9]
    DAT-06

    A published iPLEX HS study of lung tumor samples reported that some mutations were weakly called or not called by the software and were judged from the raw spectra.

    iPLEX HS lung tumor study (PubMed Central PMC5604939) (external)

    Checked on 24 November 2026Peer-reviewedmedium confidence

  10. [10]
    DAT-07

    Typer naming: Sequenom announced TYPER 4.0 on 18 September 2007 alongside iPLEX Gold. Papers print Typer software, SpectroTYPER v4.0, Typer v4.0.26.74, MassARRAY Typer Analyzer v4.0.26.73 and TyperAnalyzer, which this site treats as aliases.

    Sequenom launches TYPER 4.0 genotyping software (BioSpace); PubMed Central article PMC9022865, biospace.com/sequenom-inc-launches-typer-4-0-genotyping-software

    Checked on 24 November 2026Third partymedium confidence

  11. [11]
    WFL-09

    An iPLEX Sample ID Plus Panel study found 10 ng or more of template optimal and a relatively high dropout risk at 1 ng or less; duplicate typing gave a 90.0% average call rate, falling to 69.9% when stricter filters removed all inconsistencies, and degraded casework DNA gave lower call rates. The authors judged the panel inappropriate for crime-scene trace samples.

    Forensic Science International: Genetics (2013), iPLEX Sample ID Plus Panel evaluation, DOI 10.1016/j.fsigen.2013.04.009 (external)

    Checked on 24 November 2026Peer-reviewedmedium confidence

  12. [12]
    DAT-02

    One clinical laboratory's working rules in an iPLEX PGx 74 plus custom panel validation: plate positive control matches the consensus genotype in all assays; no-template-control extension under 10% with at least 4 of 5 NTC wells in the lowest success band; sample wells in the upper success bands with at most one marginal well per plate; assay success target of at least 85%; calls with over 75% unextended primer rejected; peaks under 5 intensity units kept only if isolated and well shaped; every allelic-discrimination plot reviewed. Not Agena defaults.

    Journal of Applied Laboratory Medicine 2020;5(3):454-466, iPLEX PGx validation study (external)

    Checked on 24 November 2026Peer-reviewedmedium confidence

  13. [13]
    DAT-02a

    In the same validation as DAT-02, average mass resolution was about 850 (m/delta m, FWHM).

    Journal of Applied Laboratory Medicine 2020;5(3):454-466, iPLEX PGx validation study (external)

    Checked on 24 November 2026Peer-reviewedmedium confidence

  14. [14]
    CHM-10

    EpiTYPER examples: an unmethylated fragment at 4484.85 Da and the same fragment with two methylated CpGs at 4516.84 Da (difference 31.99 Da, about 2 x 16); amplicons 100-600 bp (typically 250-450); default analysis window 1.5-7 kDa; 126 triplicate measurements per 384-well plate; example QC cut-offs of 0.1 replicate SD and 80% CpG-unit success. Window and thresholds are lab or software defaults.

    2006 review of MALDI-TOF DNA analysis (PubMed Central PMC1523240); Frontiers in Genetics (2015) EpiTYPER methylation protocol article (external)

    Checked on 24 November 2026Peer-reviewedmedium confidence

  15. [15]
    LAB-05

    CLIA 42 CFR 493.1256: daily minimum controls are one positive and one negative for qualitative tests, two controls for tests with an extraction step (one able to detect extraction errors), and for molecular amplification two controls plus an inhibition control where inhibition is a significant cause of false negatives.

    42 CFR 493.1256 Standard: Control procedures (eCFR) (external)

    Checked on 24 November 2026Regulator

  16. [16]
    MAN-06

    Agena's website terms list EpiTYPER, iPLEX, iSEQ, LungCarta, MassARRAY, MassEXTEND and SpectroCHIP as registered marks, and ColoCarta, OncoCarta, OncoFOCUS, UltraSEEK, MassCLEAVE, SensiPLEX, MelaCarta, LungFUSION, IMPACT Dx, Hemo ID, Agena and Agena Bioscience as trademarks. Typer is not on the list.

    Agena Bioscience: Terms and Conditions (website terms), agenabio.com/terms-and-conditions/

    Checked on 24 November 2026Manufacturer-reportedmedium confidence

  17. [17]
    DAT-08

    MassARRAY Insights is Agena's reporting network pairing the mass spectrometer with interpretation software from outside companies. Molecular Health was announced on 19 September 2017 as the first, with Molecular Match, Pathagility, Translational Software and N-of-One added later. The network's status in 2026 is unknown.

    Agena Bioscience announces Molecular Health as the first interpretation software provider for MassARRAY Insights (BioSpace), biospace.com/agena-biosciences-announces-molecular-health-as-the-first-interpretation-software-provider-for-massarray-insights

    Checked on 24 November 2026Manufacturer-reportedmedium confidence

  18. [18]
    WAR-03

    The warranty page makes these customer responsibilities: periodic maintenance and calibration, replacing hardware components such as dispense pins, liquid-handler tips and clean resin tools, data backup, networking and waste disposal. Lasers are excepted, so laser replacement is not a customer task.

    Agena Bioscience: System Warranty, agenabio.com/support/system-warranty/

    Checked on 24 November 2026Manufacturer-reportedmedium confidence