Workflow and data
Data output
What the software produces after a run, how to read calls and confidence, and how to keep your data records in order.
Synthetic data, not from a real run.
Overview
What the software produces after a run, how to read calls and confidence, and how to keep your data records in order.
After a run the software shows spectra, calls, plots and plate views. This section explains what each one is for, how laboratories review them and what to keep on record. Every figure here is synthetic or comes from a published study; none is from a real run.
In the public pages we checked on 24 November 2026, the manufacturer does not list real file types or report formats.
The manufacturer says analysis and reporting happen in Typer software, which shows qualitative results as Detected or Not Detected.[1]
In this section
Spectra
The most direct output is the spectrum for each spot: signal on the vertical axis, mass on the horizontal axis. A genotype is read from which peaks stand above the baseline and how tall they are. Published work shows spectra of relative intensity against mass as standard.[2]
The chemistry page explains where each peak comes from.
Four synthetic spectra: homozygous, homozygous, heterozygous, and no extension.
Text version of this figure
Each panel shows the unextended primer peak. Homozygous samples add one allele peak, a heterozygous sample adds two of similar height, and a failed reaction shows the primer and noise only.
Synthetic data, not from a real run.
Genotype calls and confidence
The software turns each spectrum into a call and attaches a confidence label. One published study reported five tiers from conservative to low probability, plus calls made by a person.[3] Low-frequency work uses signal-to-noise and z-score thresholds to separate calls from no-calls.[4]
Calls and confidence explains the tiers and lets you move a stringency slider on a synthetic data set.
Cluster plots
An allelic-discrimination or cluster plot puts one assay across many samples. Samples fall into tight groups by genotype; points between groups are judged by eye.[2]
Open the cluster plot explorer to click a point and see its schematic spectrum.
Synthetic cluster plot: 96 samples, one assay.
Text version of this figure
Three tight clusters (two homozygous, one heterozygous), a few points between them and a few failures near the origin.
Synthetic data, not from a real run.
Plate views
A plate view shows the whole plate at a glance, with each well tinted by the share of successful calls. A problem with a row, a column or a corner shows up as a pattern.
One laboratory's rule, not a specification Red under 15%, yellow from 15% to under 50%, light green from 50% to under 85%, dark green at 85% or more.[2]
Synthetic 96-well plate view in one laboratory's four color bands, with letters and patterns so color is not the only cue.
Text version of this figure
A 96-well grid. Most wells are in the highest band and a few are lower. The wells are coded by letter as well as by shade.
Synthetic data, not from a real run.
Mutation matrices
In one published UltraSEEK study the output is a table with one row per sample and one column per mutation. A cell holds T for a high-level call, L for a low-level call, or nothing.[4]
| Sample | Mutation 1 | Mutation 2 | Mutation 3 | Mutation 4 | Mutation 5 | Mutation 6 |
|---|---|---|---|---|---|---|
| Sample A | T | - | - | - | L | - |
| Sample B | - | - | - | - | - | - |
| Sample C | - | L | - | T | - | - |
| Sample D | - | - | - | - | - | L |
Synthetic data, not from a real run.
Methylation grids
For methylation work the output is a percentage for each CpG unit in each sample: the methylated peak area divided by the methylated plus unmethylated area.[5]
| Sample | CpG unit 1 | CpG unit 2 | CpG unit 3 | CpG unit 4 | CpG unit 5 | CpG unit 6 |
|---|---|---|---|---|---|---|
| Sample A | 12% | 8% | 70% | 64% | 31% | 5% |
| Sample B | 15% | 10% | 22% | 18% | 29% | 7% |
| Sample C | 11% | 9% | 81% | 77% | 35% | 4% |
Synthetic data, not from a real run.
Calls and confidence
How calls are made from spectra and how confidence shows up, with a synthetic cluster plot to explore.
A call is the software’s reading of one spectrum: which allele or alleles it sees, or that it sees none. Most software also says how sure it is. This page uses a published study to show what confidence tiers look like, then lets you change a stringency setting on synthetic data to see calls turn into no-calls.
The manufacturer says Typer software shows qualitative results as Detected or Not Detected.[1] For genotyping, the output is a call per sample and assay with a confidence label.
Confidence tiers in one published study
A peer-reviewed marker study using Typer reported genotype confidence in five tiers (conservative, moderate, aggressive, low probability, user call) and a group with no alleles. The table gives that study’s shares. It covers one species and one marker set and is not a performance specification.[3]
Read the tier names as an ordered ladder, from the calls the software is most sure of to those it is least sure of. The exact criteria for each tier belong to the software and your method, and are not stated here.
| Tier in that study | How to read the name | Share of genotypes |
|---|---|---|
| Conservative | The calls the software is most sure of. | 54% |
| Moderate | Still confident, below the top tier. | 15% |
| Aggressive | Called on looser criteria than the two tiers above. | 9% |
| Low probability | A call was made, and the software rates it as unlikely to be reliable. | 18% |
| No alleles | No allele signal was found. | 3% |
| User call | A person made or confirmed the call after looking at the spectrum. | 1% |
The tier ladder with the user-call branch and the no-call at the bottom. Tier names as used in one published study.
Text version of this figure
From top to bottom: conservative, moderate, aggressive, low probability, then no call. A side branch shows a user call made after review. An annotation gives the trade-off between a lenient and a strict setting as an average call rate of 90.0% against 69.9%.
What counts as a failed assay
The same study defined a failed assay as one where most genotypes were low-probability or where the samples did not cluster by genotype.[3] Both signs are visible on a cluster plot before any number is computed, which is why laboratories look at the plot as well as the call rate.
Low-frequency calls: one study’s rules
An UltraSEEK breast-panel study of 39 mutations in 5 genes used 10 ng of DNA and Typer v4.0.26.74. A mutant call needed a signal-to-noise ratio of at least 6 and a z-score of at least 7. A z-score from 7 to 10 was labeled L (low-level) and above 10 was labeled T (high-level). The output was a per-sample, per-mutation matrix of T and L codes, plus spectra.[4]
These are the authors’ rules for their panel, not manufacturer specifications. Your thresholds come from your own validation.
Three spectra with the same baseline: not called, called at low level, called at high level.
Text version of this figure
A spectrum with signal-to-noise 3 is not called. One with signal-to-noise 8 is called at low level (L). One with signal-to-noise 20 is called at high level (T).
Synthetic data, not from a real run.
The stringency trade-off
Stricter filtering removes doubtful calls and also removes some good ones. In a published study of an iPLEX identity panel, duplicate typing gave an average call rate of 90.0% with the default settings, falling to 69.9% when stricter filters removed every inconsistency. The same study found 10 ng or more of template optimal, a relatively high dropout risk at 1 ng or less, and lower call rates from degraded DNA.[11]
That is the trade you are making when you set a threshold: fewer wrong calls against more no-calls. The explorer below shows the same effect on synthetic points.
Cluster plot explorer
Move the stringency slider from permissive to strict and watch calls turn into no-calls. Points between clusters are called only at the most permissive settings, which is where a wrong call is most likely.
96 synthetic points, one assay.
Synthetic data, not from a real run.
- Called homozygous, allele 1
- Called heterozygous
- Called homozygous, allele 2
- No call
Calls accepted down to low probability. 92 of 96 points are called; 4 are no-calls.
| Group | Points | Called | No-call |
|---|---|---|---|
| Cluster 1 (homozygous, allele 1) | 28 | 28 | 0 |
| Cluster 2 (heterozygous) | 40 | 40 | 0 |
| Cluster 3 (homozygous, allele 2) | 20 | 20 | 0 |
| In-between points | 4 | 4 | 0 |
| Failed, little signal | 4 | 0 | 4 |
Inspect a point
Click a point on the plot, or choose a sample number here.
- Signal, allele 1 and 2
- 0.88 and 0.18
- Nearest cluster
- homozygous, allele 1
- Tier (demonstration rule)
- conservative
- At this setting
- Called
Schematic spectrum from the point’s two signals. Synthetic data, not from a real run.
No-call handling belongs to the SOP
What you do with a no-call, or with a low-probability call, is a decision for your laboratory: repeat the well, review the spectrum, or record it as not determined. Write that rule in the procedure before the run, not during review. See the run record and manual review pages.
QC metrics
Which quality metrics to track per run and per plate, with a worksheet that leaves every limit for you to set.
Quality control on a MassARRAY run asks two questions: did the run behave, and did each sample behave? Published validations describe the same kinds of check. No numeric limit is published as a standard, so each laboratory takes its limits from its own validation.[6]
This page lists the metrics, shows one laboratory’s working rules as an example of what a written rule looks like, and gives you a worksheet with every limit left blank.
Metrics laboratories compute
The SNPflow quality-control tool for iPLEX output computes call rate, expected against observed replicates, minor allele frequency, discordant replicates, discordance rate and a Hardy-Weinberg test, and has a sex-determination module. It sets no numeric cut-offs.[7]
| Metric | What it asks | Where it is described |
|---|---|---|
| Call rate per sample | How many assays gave a call for each sample. | SNPflow[7] |
| Call rate per assay | How many samples gave a call for each assay. | SNPflow[7] |
| Negative-control extension | Whether the no-template wells show signal. | Published validations[6] |
| Reference sample concordance | Whether a sample of known genotype gives its known result. | Published validations[6] |
| Replicate concordance | Whether duplicates of the same sample agree. | SNPflow, published validations[7][6] |
| Hardy-Weinberg test | Whether genotype frequencies fit expectation, where the study design supports the test. | SNPflow[7] |
| Identity or sex marker | Whether the sample is the one you meant. | SNPflow, published validations[7][6] |
One laboratory’s working rules
- The plate positive control matches the consensus genotype in all assays.
- No-template-control extension is under 10%, with at least 4 of 5 no-template wells in the lowest success band.
- Sample wells fall in the upper success bands, with at most one marginal well per plate.
- The assay success target is at least 85%.
- Calls with more than 75% unextended primer are rejected.
- Very small peaks, under 5 intensity units, are kept only if isolated and well shaped.
- Every allelic-discrimination plot is reviewed.
The same validation reported an average mass resolution of about 850 (m/Δm, full width at half maximum), which is the width we use for the illustrative peaks on the chemistry page.[13]
Plate view
A plate view tints each well by the share of successful calls so that row, column and edge problems show as patterns. One laboratory's rule, not a specification Red is under 15%, yellow 15% to under 50%, light green 50% to under 85% and dark green 85% or more.[2]
Synthetic 96-well plate view in one laboratory's four bands, coded with letters and patterns as well as color.
Text version of this figure
A 96-well grid with most wells in the top band and a few wells, grouped at one edge, in lower bands.
Synthetic data, not from a real run.
Methylation examples
EpiTYPER examples in the literature use triplicate measurements (126 per 384-well plate), a replicate standard deviation cut-off of 0.1 and 80% success across CpG units. The default analysis window is 1.5 to 7 kDa. The window and the thresholds are laboratory or software defaults, not manufacturer limits.[14]
Daily controls where CLIA applies
In the United States, 42 CFR 493.1256 sets daily minimum controls: one positive and one negative for qualitative tests, two controls for tests with an extraction step (one able to detect extraction errors), and for molecular amplification two controls plus an inhibition control where inhibition is a significant cause of false negatives.[15]
Whether that applies to you depends on your regulatory framework. It is a floor for some laboratories, not a description of what a validation needs.
General information; not legal, regulatory or clinical advice.
QC worksheet
Fill in your own limit for each metric before the run, enter what you observe, and record the result. Print it, or export a CSV or Markdown file.
Limits are yours to set.
Use your own codes. Do not enter patient, sample or personal data.
| Metric | What it asks | Limit set by my laboratory | Observed value | Result | Notes | Remove |
|---|---|---|---|---|---|---|
| Call rate per sample | Share of assays with a call for each sample | |||||
| Call rate per assay | Share of samples with a call for each assay | |||||
| Negative-control (no-template) extension | Signal in the no-template wells | |||||
| Reference sample concordance | Calls against the known genotype | |||||
| Replicate concordance | Agreement between duplicates | |||||
| Hardy-Weinberg check | Genotype frequencies against expectation, where your design supports it | |||||
| Identity or sex marker | Marker result against the record for the sample | |||||
| Allelic-discrimination plots reviewed | Whether every plot was looked at, and by whom |
Manual review
When and how to review a call by hand, and what to record when you override the software.
Software makes the first call. A person still looks at the evidence behind it. This page explains why that remains normal practice in published MassARRAY work, what reviewers look at, and what a review record should hold.
Why manual review persists
A published validation of a multi-gene panel needed significant development time, modified the manufacturer’s protocol and reviewed every assay plate manually.[8]
A published iPLEX HS study of lung tumor samples found that some mutations were weakly called or not called by the software and were judged from the raw spectra.[9]
Cluster plots with tight bands and in-between points that are judged by hand are standard in published MassARRAY work.[2] Published validations also describe second-person review as part of typical quality control.[6]
The point is not that the software is poor. It is that the software gives a call and the laboratory owns the result.
What reviewers look at
Three views, used together. Each one catches a problem the others hide. All three below are synthetic.
Spectra: is there a real peak, at the right mass, above the noise?
Text version of this figure
Four small spectra: homozygous, homozygous, heterozygous and a failed reaction.
Synthetic data, not from a real run.
Cluster plot: does the sample sit inside a cluster, or between two?
Text version of this figure
Three clusters of points, a few points between them, and a few near the origin.
Synthetic data, not from a real run.
Plate view: is the problem one well, or a row, a column or an edge?
Text version of this figure
A 96-well grid with a few low-scoring wells grouped together.
Synthetic data, not from a real run.
To try this yourself, open the cluster plot explorer and click the points that sit between two clusters.
A review policy outline
An outline for your SOP. It sets no acceptance criteria and no counts: those come from your validation.
Step 1: Say in the SOP what must be reviewed
List the cases that always go to a person: every control, every low-confidence call, every no-call, and a plot review for every assay on every plate.
Step 2: Give the reviewer the whole picture
Spectrum, cluster plot and plate view side by side, with the plate map, so the reviewer can see a pattern and not only a point.
Step 3: Decide from a short list
Accept the software’s call, change it, or mark it as no-call. Anything that does not fit goes to the person your SOP names.
Step 4: Record the decision where the next person will look
Use the record fields in the table below. A change with no reason written down is the first thing anyone reviewing the record will ask about.
Step 5: Get a second reviewer for the calls that changed
A second person looks at every changed call and every control failure, and signs.
What to record when a call changes
| Record | Why it is on the record |
|---|---|
| Plate or run code, sample position, assay | So anyone can find the exact spectrum again. |
| The software's call and its confidence label | What the software said before a person touched it. |
| The reviewer's call, or no-call | What was decided. |
| The reason, in a short coded list plus free text | A reviewer a year later can tell a judgment from a slip. |
| Evidence reference | A saved spectrum or plot image or a file path, so the evidence travels with the decision. |
| Reviewer, second reviewer, date | Review by two people is the usual control on a human decision. |
| Software version and settings in use | The same data can read differently under different settings. |
The records page has a checklist for keeping these files, and the run log template has a block for review and sign-off.
Typer software and records
Records and data-integrity habits around the Typer software, with a checklist for your own procedures.
The analysis software, the files it makes and the records around them are what a laboratory shows when someone asks “how do you know?” This page sorts out the names you will meet, says what this site does not state about the software, and gives you a checklist for your own record-keeping.
Typer naming and aliases
Papers and older documents print the analysis software under several names. This site treats them as aliases of one product line, so a reference to any of them in a record means the same family of software; the version still matters for your records.[10]
| Name as printed | Where it appears |
|---|---|
| Typer | General name for the analysis software. |
| TYPER 4.0 | Announced on 18 September 2007, alongside iPLEX Gold. |
| SpectroTYPER v4.0 | Printed in published papers. |
| Typer v4.0.26.74 | Printed in a published study. |
| MassARRAY Typer Analyzer v4.0.26.73 | Printed in published papers. |
| TyperAnalyzer | Short form in published papers. |
“Typer” is not on the manufacturer’s published list of marks, so this site writes it without a symbol and makes no statement about who owns the name.[16]
Interpretation software from other providers
The manufacturer has described a reporting network that pairs the mass spectrometer with interpretation software from other providers. The first was announced on 19 September 2017 and others were added later. The network’s status in 2026 is unknown.[17]
If your results flow into interpretation software, record its name and version in the run log beside the analysis software, as part of the same change control.
What this site does not state
This site makes no statement about these topics for any version of the software:
- electronic-record and e-signature rules
- audit trails
- laboratory information system connections
- export file formats
- operating-system and PC requirements
- software support dates
Ask the manufacturer for the documents that cover your software version, and keep a copy in your records. If it helps, ask us to confirm the answer with you at an installation or service visit.
We do not reproduce manufacturer manuals, procedures or error codes. Ask the manufacturer or use your customer portal.
Records and backups
A common test for a laboratory record asks whether it is attributable (who made it), legible, contemporaneous, original and accurate. Run data are no different from a paper notebook in that respect.
Data backup, networking and waste disposal are on the manufacturer’s list of customer responsibilities.[18] So the backup of the analysis computer is yours to plan: what is backed up, how often, where the copy lives, and who has tested a restore.
Keep the plate map, the run log and the result files together, and record the software version used. The run log template, the manual review record fields and the PC and OS change page all feed the same record.
Data-integrity checklist
An original prompt list for your own procedures. It is not a statement about any software and sets no requirement.
For your own procedures.
0 of 12 checks assessed. A status of No is a finding to follow up, not a failure of this tool.
Sources
[1]
WFL-01Agena presents the MassARRAY System as an open platform in 96- and 384-well formats: end-point PCR with single-base extension, automated sample transfer and MALDI-TOF detection, then analysis and reporting in Typer software, which shows qualitative results as Detected / Not Detected.
Agena Bioscience: MassARRAY System product page, agenabio.com/products/massarray-system/
Checked on 24 November 2026Manufacturer-reportedmedium confidence
[2]
DAT-09Cluster (allelic-discrimination) plots with tight bands and in-between points judged manually, spectra of relative intensity against mass, and plate views tinted by share of successful calls are standard in published MassARRAY work. One lab's plate color bands: red under 15%, yellow 15 to under 50%, light green 50 to under 85%, dark green 85% or more.
Journal of Applied Laboratory Medicine 2020;5(3):454-466, iPLEX PGx validation study (external)
Checked on 24 November 2026Peer-reviewedmedium confidence
[3]
DAT-01A peer-reviewed marker study using Typer reported genotype confidence tiers (conservative, moderate, aggressive, low probability, user call): 54% conservative, 15% moderate, 9% aggressive, 18% low probability, 3% no alleles and 1% user calls. It defined a failed assay as one where most genotypes were low-probability or samples did not cluster by genotype. One species and marker set, not a performance specification.
PLoS ONE genotyping study using Typer software (journal.pone.0049533) (external)
Checked on 24 November 2026Peer-reviewedmedium confidence
[4]
DAT-05UltraSEEK breast panel study (39 mutations in 5 genes, 10 ng, Typer v4.0.26.74): a mutant call needed signal-to-noise of at least 6 and a z-score of at least 7; z 7-10 is labeled L (low-level) and above 10 T (high-level). Output is a per-sample, per-mutation matrix with T/L codes plus spectra of relative intensity against mass-to-charge. These are the authors' rules, not Agena specifications.
UltraSEEK breast panel study (PubMed Central PMC8558178) (external)
Checked on 24 November 2026Peer-reviewedmedium confidence
[5]
CHM-09Published EpiTYPER protocol: bisulfite conversion, PCR with a T7 promoter tag, shrimp alkaline phosphatase, T7 transcription, RNase A cleavage after U positions (T-reaction), a clean-up step, then MALDI-TOF. A methylated CpG unit gives a fragment about 16 Da heavier; percent methylation is the methylated peak area over methylated plus unmethylated area.
Checked on 24 November 2026Peer-reviewedmedium confidence
[6]
DAT-10Published validations describe typical QC practice: reserved negative-control wells, reference samples of known genotype, duplicates to check discordance, an identity or sex marker, and second-person review. Acceptance numbers come from each laboratory's own validation; no numeric limit is published as a standard.
Checked on 24 November 2026Peer-reviewedmedium confidence
[7]
WFL-11The SNPflow QC tool for iPLEX output computes call rate, expected versus observed replicates, minor allele frequency, discordant replicates, discordance rate and a Hardy-Weinberg test, checks genotype frequencies against reference data and has a sex-determination SNP module. It sets no numeric cut-offs.
Checked on 24 November 2026Peer-reviewedmedium confidence
[8]
DAT-04A clinical laboratory's published account of validating a pharmacogenomic panel (J Appl Lab Med 2020;5(3):454-466). It needed significant development time, modified the manufacturer's protocol and reviewed every assay plate manually.
Journal of Applied Laboratory Medicine 2020;5(3):454-466 (PubMed 32445367) (external)
Checked on 24 November 2026Peer-reviewedmedium confidence
[9]
DAT-06A published iPLEX HS study of lung tumor samples reported that some mutations were weakly called or not called by the software and were judged from the raw spectra.
iPLEX HS lung tumor study (PubMed Central PMC5604939) (external)
Checked on 24 November 2026Peer-reviewedmedium confidence
[10]
DAT-07Typer naming: Sequenom announced TYPER 4.0 on 18 September 2007 alongside iPLEX Gold. Papers print Typer software, SpectroTYPER v4.0, Typer v4.0.26.74, MassARRAY Typer Analyzer v4.0.26.73 and TyperAnalyzer, which this site treats as aliases.
Sequenom launches TYPER 4.0 genotyping software (BioSpace); PubMed Central article PMC9022865, biospace.com/sequenom-inc-launches-typer-4-0-genotyping-software
Checked on 24 November 2026Third partymedium confidence
[11]
WFL-09An iPLEX Sample ID Plus Panel study found 10 ng or more of template optimal and a relatively high dropout risk at 1 ng or less; duplicate typing gave a 90.0% average call rate, falling to 69.9% when stricter filters removed all inconsistencies, and degraded casework DNA gave lower call rates. The authors judged the panel inappropriate for crime-scene trace samples.
Checked on 24 November 2026Peer-reviewedmedium confidence
[12]
DAT-02One clinical laboratory's working rules in an iPLEX PGx 74 plus custom panel validation: plate positive control matches the consensus genotype in all assays; no-template-control extension under 10% with at least 4 of 5 NTC wells in the lowest success band; sample wells in the upper success bands with at most one marginal well per plate; assay success target of at least 85%; calls with over 75% unextended primer rejected; peaks under 5 intensity units kept only if isolated and well shaped; every allelic-discrimination plot reviewed. Not Agena defaults.
Journal of Applied Laboratory Medicine 2020;5(3):454-466, iPLEX PGx validation study (external)
Checked on 24 November 2026Peer-reviewedmedium confidence
[13]
DAT-02aIn the same validation as DAT-02, average mass resolution was about 850 (m/delta m, FWHM).
Journal of Applied Laboratory Medicine 2020;5(3):454-466, iPLEX PGx validation study (external)
Checked on 24 November 2026Peer-reviewedmedium confidence
[14]
CHM-10EpiTYPER examples: an unmethylated fragment at 4484.85 Da and the same fragment with two methylated CpGs at 4516.84 Da (difference 31.99 Da, about 2 x 16); amplicons 100-600 bp (typically 250-450); default analysis window 1.5-7 kDa; 126 triplicate measurements per 384-well plate; example QC cut-offs of 0.1 replicate SD and 80% CpG-unit success. Window and thresholds are lab or software defaults.
Checked on 24 November 2026Peer-reviewedmedium confidence
[15]
LAB-05CLIA 42 CFR 493.1256: daily minimum controls are one positive and one negative for qualitative tests, two controls for tests with an extraction step (one able to detect extraction errors), and for molecular amplification two controls plus an inhibition control where inhibition is a significant cause of false negatives.
42 CFR 493.1256 Standard: Control procedures (eCFR) (external)
Checked on 24 November 2026Regulator
[16]
MAN-06Agena's website terms list EpiTYPER, iPLEX, iSEQ, LungCarta, MassARRAY, MassEXTEND and SpectroCHIP as registered marks, and ColoCarta, OncoCarta, OncoFOCUS, UltraSEEK, MassCLEAVE, SensiPLEX, MelaCarta, LungFUSION, IMPACT Dx, Hemo ID, Agena and Agena Bioscience as trademarks. Typer is not on the list.
Agena Bioscience: Terms and Conditions (website terms), agenabio.com/terms-and-conditions/
Checked on 24 November 2026Manufacturer-reportedmedium confidence
[17]
DAT-08MassARRAY Insights is Agena's reporting network pairing the mass spectrometer with interpretation software from outside companies. Molecular Health was announced on 19 September 2017 as the first, with Molecular Match, Pathagility, Translational Software and N-of-One added later. The network's status in 2026 is unknown.
Agena Bioscience announces Molecular Health as the first interpretation software provider for MassARRAY Insights (BioSpace), biospace.com/agena-biosciences-announces-molecular-health-as-the-first-interpretation-software-provider-for-massarray-insights
Checked on 24 November 2026Manufacturer-reportedmedium confidence
[18]
WAR-03The warranty page makes these customer responsibilities: periodic maintenance and calibration, replacing hardware components such as dispense pins, liquid-handler tips and clean resin tools, data backup, networking and waste disposal. Lasers are excepted, so laser replacement is not a customer task.
Agena Bioscience: System Warranty, agenabio.com/support/system-warranty/
Checked on 24 November 2026Manufacturer-reportedmedium confidence
Research-use context. No diagnostic or clinical claims. Read the research-use notice