Home

Independent instrument service for MassARRAY systems

Close view of a brushed metal sample plate with a grid of small wells, a few glowing amber.

Workflow and data

Chemistry and contamination control

The assay chemistry and mass spectrometry behind the workflow, and how to keep pre- and post-amplification work apart.

Checked on 24 November 2026

The overview and each stage are on Workflow.

MALDI-TOF and assay chemistry

A plain explainer of the assay chemistry and mass spectrometry, with a synthetic spectra explorer.

This page explains how the assay turns a DNA variant into a peak on a mass spectrum. It uses original diagrams and one computed example that you can change. The numbers in the example are hypothetical; the numbers from the manufacturer, papers and patents are attributed where they appear.

Where we describe a peak as “closer together and harder to separate reliably”, we mean exactly that: the peaks overlap more, and a call from them is less certain. We never say two peaks are “too close to separate”.

Principle: weighing molecules by their flight time

MALDI-TOF stands for matrix-assisted laser desorption/ionization, time of flight. The sample is mixed with a matrix and dried on a pad, where the two crystallize together. A short laser pulse turns some of the sample into ions. An electric field accelerates the ions into a long flight tube under vacuum, and a detector at the far end records when each ion arrives.

For ions that carry the same charge, flight time grows with the square root of mass. Lighter ions arrive first and heavier ones later, so the arrival times become a mass spectrum: a plot of signal against mass.

DNA suits this method only when it is short. Genotyping assays therefore do not measure the long PCR product. They measure a short extension primer whose mass changes by a known amount when one base is added.

Intact detection works for short single-stranded DNA of about 3 to 29 bases; in practice the limit is below about 100 bases because longer fragments break up during MALDI. That is a 2006 review, not a current manufacturer specification.[1]

Covers stay closed. Work inside an instrument is for a qualified service engineer only. Safety statement

MALDI-TOF measurement, schematicDiagram of a MALDI-TOF measurement: a laser pulse ionizes analyte on a matrix spot, ions are accelerated into a vacuum flight tube, lighter ions reach the detector first, producing a mass spectrum.magnified padanalyte co-crystallizedwith matrixsample chiplaser pulseaccelerating fieldfield-free flight tubeunder vacuumdetectorpulse tracetime of flightt1t2t3lighter ions arrive firstheavier ions arrive latert1 < t2 < t3mass spectrumflight time increases with the square root of mass-to-chargeSchematic. Not drawn from any specific instrument.

A MALDI-TOF measurement, left to right. Schematic; not drawn from any specific instrument.

Text version of this figure

A laser pulse ionizes analyte that has crystallized with matrix on a chip pad. Ions are accelerated by an electric field into a vacuum flight tube. Lighter ions reach the detector first and heavier ions later, which produces a mass spectrum.

iPLEX in five steps

Single-base extension chemistry in five stepsFive panels: locus-specific PCR, shrimp alkaline phosphatase treatment, single-base extension with four terminator tiles, desalting with resin, and dispensing onto a matrix-loaded chip pad to read a small spectrum.1Locus-specific PCRPrimers copy the shortregion that holds theSNP.2SAP treatmentShrimp alkalinephosphatase neutralizesunused nucleotides.3Single-base extensionOne terminator is addednext to the SNP.4DesaltingSalts are removed beforespotting.5Dispense and detectA nanoliter-scale spotgoes onto amatrix-loaded pad.SNPprimerprimerampliconSAPunused nucleotides grayedext. primerCTAGweight (qualitative)resin bead+--+small ions capturedspotchip pad with matrixprimer + allele peaks

The five panels of the iPLEX workflow. No volumes, cycling profiles or terminator structures are shown.

Text version of this figure
  1. Locus-specific PCR amplifies the region around the variant.
  2. Shrimp alkaline phosphatase neutralizes unused nucleotides.
  3. A primer next to the variant is extended by one mass-modified terminator.
  4. Salts are removed before spotting.
  5. The product is dispensed onto a chip pad, and the spectrum shows the unextended primer and one or two allele peaks.

iPLEX is a locus-specific PCR, then shrimp alkaline phosphatase treatment to neutralize leftover nucleotides, then single-base extension of a primer that anneals directly next to the variant using mass-modified dideoxynucleotide terminators, then MALDI-TOF. The mass of the extended primer shows which allele was present.[14]

Manufacturer-reported (2026) The manufacturer states the read window of one chemistry as about 4,500 to 9,000 Da and says the system distinguishes analytes 16 Da apart. A 2006 review gives gaps of at least 16 Da between the four terminator products and the same window.[3]

Why 16 Da matters: a worked example

Take an invented 20-mer extension primer, 5′-ACGTAGCTAGCATCGGATCA-3′, with six A, five C, five G and four T. Add the average residue masses (A 313.21, C 289.18, G 329.21, T 304.20 Da), subtract the 61.96 Da correction, and the primer weighs 6,126.05 Da. A plain dideoxy terminator adds its residue mass minus 16.00 Da, so the four possible products come out at 6,399.23 (C), 6,414.25 (T), 6,423.26 (A) and 6,439.26 Da (G).[2]

The six possible allele pairs then sit 9.01 to 40.03 Da apart. The A/T pair, 9.01 Da apart, is below the 16 Da the manufacturer states the system distinguishes, so with plain terminators its two peaks sit closer together and are harder to separate reliably. Mass-modified terminators exist to widen exactly this kind of gap. Their real mass shifts are not public, so the explorer below shows a schematic with the published 16 Da minimum spacing.[3]

Extension primer and allele peaks with plain terminatorsSix small spectra showing the extension primer peak and the two allele peaks for each possible SNP pair when plain dideoxy terminators are used; the A/T pair is only 9 Da apart.Shaded band: the 16 Da separation the manufacturer states the system distinguishes [C]Synthetic data, not from a real run.A / T6,423.26 and 6,414.25 Dagap 9.01 Da [K], below 16 Davalley about 72% of peak height [K]6,2006,400primerATC / T6,399.23 and 6,414.25 Dagap 15.02 Da [K], below 16 Davalley about 12% of peak height [K]6,2006,400primerCTA / G6,423.26 and 6,439.26 Dagap 16.00 Da [K], equal to 16 Davalley about 9% of peak height [K]6,2006,400primerAGA / C6,423.26 and 6,399.23 Dagap 24.03 Da [K], above 16 Davalley between peaks near the baseline6,2006,400primerACG / T6,439.26 and 6,414.25 Dagap 25.01 Da [K], above 16 Davalley between peaks near the baseline6,2006,400primerGTC / G6,399.23 and 6,439.26 Dagap 40.03 Da [K], above 16 Davalley between peaks near the baseline6,2006,400primerCG4,500 Da9,000 Dathis view: 6,100 to 6,460 DaRead window: about 4,500 to 9,000 Da [C]Hypothetical 20-mer primer (A6 C5 G5 T4) and average residue masses; the manufacturer's mass-modified terminator increments arenot public. Plain dideoxy terminators are shown to explain the problem mass modification solves. Peaks that sit closer togetherare harder to separate reliably.Peak width 7.5 Da [I], sigma 3.18 Da: illustrative width, not an instrument specification.

Six small spectra of the worked example, one for each allele pair. Peak width is illustrative, set near the average resolution one laboratory reported.

Text version of this figure

For each of the six allele pairs the figure shows the unextended primer at 6,126.05 Da and the two allele peaks. The A/T pair is 9.01 Da apart and overlaps strongly. The C/T pair is 15.02 Da apart. A/G is 16.00 Da apart. The A/C, G/T and C/G pairs are 24.03, 25.01 and 40.03 Da apart and are well separated.

Synthetic data, not from a real run.

Worked example: gaps between allele peaks with plain dideoxy terminators (computed)
Allele pairProduct masses (Da)Gap (Da)Against 16 DaValley between equal peaks
A/TT 6,414.25 and A 6,423.269.01Below 16 Daabout 72%
C/TC 6,399.23 and T 6,414.2515.02Below 16 Daabout 12%
A/GA 6,423.26 and G 6,439.2616.00Equal to 16 Daabout 9%
A/CC 6,399.23 and A 6,423.2624.03Above 16 Daunder 1%
G/TT 6,414.25 and G 6,439.2625.01Above 16 Daunder 1%
C/GC 6,399.23 and G 6,439.2640.03Above 16 Daunder 1%

Valley depth is the summed curve at the midpoint of two equal peaks as a share of the highest point, using the illustrative peak width of 7.5 Da (full width at half maximum, about 850 in m/Δm at 6,414 Da). One laboratory reported an average mass resolution of about 850 in its own validation, which is why we chose that width; it is not an instrument specification.[15]

I-13Spectra explorer

Computed, illustrative.

Synthetic spectrum explorerSynthetic spectrum for a heterozygous T/A sample. Unextended primer at 6,126.05 Da; allele peaks at 6,414.25 and 6,423.26 Da, 9.01 Da apart.6,1006,2006,3006,400primerTAm/z (Da), syntheticShaded band: 16 Da

Synthetic data, not from a real run. Illustrative peak width, not an instrument specification.

Terminators
Sample genotype

Readout

Unextended primer
6,126.05 Da
Allele T
6,414.25 Da
Allele A
6,423.26 Da
Gap between alleles
9.01 Da
Valley between equal peaks
about 72% of peak height

A gap of 9.01 Da is below 16 Da. The two allele peaks sit closer together and are harder to separate reliably. This is the problem that mass-modified terminators are used to solve.

I-13Oligo mass calculator

Illustrative average masses, not instrument values.

Other characters are ignored. 20 bases counted. Do not enter a real assay design or any sample data.

Arithmetic for the primer mass
BaseCountResidue mass (Da)Subtotal (Da)
A6313.211,879.26
C5289.181,445.90
G5329.211,646.05
T4304.201,216.80
Sum6,188.01
Minus correction−61.96
Primer mass6,126.05
Product mass with each plain dideoxy terminator
TerminatorAdds (Da)Product (Da)
A297.216,423.26
C273.186,399.23
G313.216,439.26
T288.206,414.25
Gaps between allele pairs, smallest first
PairGap (Da)Against 16 Da
A/T9.01Below 16 Da
C/T15.02Below 16 Da
A/G16.00Equal to 16 Da
A/C24.03Above 16 Da
G/T25.01Above 16 Da
C/G40.03Above 16 Da

The primer mass of 6,126.05 Da lies inside the 4,500 to 9,000 Da window the manufacturer states for one chemistry.

Genotypes as peaks

A homozygous sample shows the unextended primer plus one allele peak. A heterozygous sample shows two allele peaks of similar height. A reaction that failed, and a no-template control, show mostly the unextended primer and baseline noise.

Published MassARRAY work shows the same three views side by side: spectra of relative intensity against mass, cluster plots, and plate views. See data output.[4]

What genotype and failed-reaction spectra look likeFour schematic spectra: homozygous A with one allele peak, homozygous T-star with one peak 16 Da lower, heterozygous A and T-star with two peaks, and a failed or negative reaction with only the unextended primer peak.Synthetic data, not from a real run.a Homozygous A6,2006,3006,4006,500unextended primer 6,126.05A 6,423.26Homozygote: one extended peak.b Homozygous T*6,2006,3006,4006,500unextended primer 6,126.05T* 6,407.26Homozygote for the other allele: one extended peak, 16 Da lower.c Heterozygous A / T*6,2006,3006,4006,500unextended primer 6,126.05A 6,423.26T* 6,407.26Heterozygote: two extended peaks of similar height.d No extension (failed or negative control)6,2006,3006,4006,500unextended primer 6,126.05Failed or negative reaction: only the unextended primer.Schematic: spacing set at the published 16 Da minimum; the real mass-modified increments are not public. Heights are illustrative [I]; theprimer mass 6,126.05 Da is the computed example [K].

Four spectra: homozygous A, homozygous T*, heterozygous A/T*, and no extension. Schematic spacing set at the published 16 Da minimum.

Text version of this figure

Panel (a), homozygous A: the primer and one peak at 6,423.26 Da. Panel (b), homozygous T*: the primer and one peak 16 Da lower. Panel (c), heterozygous: both allele peaks at similar height. Panel (d), no extension: the primer peak and noise only.

Synthetic data, not from a real run.

Salts and adducts

As reported in a published patent: sodium adds about 22 Da to a DNA peak, ammonia 17 Da (mainly on G and T) and ammonia plus potassium 55 Da. The matrix itself gives peaks at 94, 138 and 188 Da. Ammonia adducts made one assay repeatedly call heterozygotes, and sodium adducts caused a false insertion call in the patent’s examples.[5]

Look at the drawing. A 17 Da shift lands about 1 Da from the neighboring allele peak, and 22 Da lands about 6 Da above it. Both are inside the width of a peak, so an adduct can pass for the other allele.

This is why desalting before spotting matters: salt adducts and incomplete desalting can distort spectra and cause wrong calls.[6] It is a vendor-reported example, not a general rate.

Salt adducts next to an allele peakA schematic pair of allele peaks with a sodium adduct 22 Da and an ammonia adduct 17 Da above the T-star peak; the adducts fall within the width of the A peak. A small inset shows matrix peaks at 94, 138 and 188 Da.Adducts of the T* peakAs reported in a published patent6,4006,4106,4206,4306,440T* 6,407.26A 6,423.26+17 6,424.26+22 6,429.26Daammonia adduct of T*: 1.00 Da from the A peak [K]sodium adduct of T*: 6.00 Da above the A peak [K]Both adducts land inside the width of the A peak (shaded; peak width 7.5 Da [I]), sothe reader cannot always tell them from a second allele.Low-mass strip: matrix peaks [C]94138188mass in DaShifts as reported in a published patent family (sodium +22, ammonia +17, ammonia plus potassium +55 Da; matrix peaks 94, 138, 188Da). Illustrative heights.

A T* peak with sodium and ammonia adducts beside the A peak. Shifts as reported in a published patent family; heights are illustrative.

Text version of this figure

The T* peak is accompanied by smaller peaks 17 and 22 Da higher. The 17 Da adduct falls about 1 Da from the A peak and the 22 Da adduct about 6 Da above it, both inside the A peak’s width. An inset shows matrix peaks at 94, 138 and 188 Da.

Synthetic data, not from a real run.

Rare alleles and iPLEX HS

As reported in a published patent Standard iPLEX uses equimolar terminators, so a minor allele below roughly 10% is hard to see. An Agena patent lowers the wild-type terminator relative to the mutant (about 0.5% to under 20% of the mutant terminator, or 0.01% to 10% in a narrower claim) and shows minor species detected down to 1.25%. The patent does not name iPLEX HS.[7]

Manufacturer-reported (2026) The manufacturer’s page describes iPLEX HS as a mutant-enriched single-base extension with about 1% variant allele frequency, as little as 10 ng of DNA, amplicons of 80 to 120 bp, about 1 hour hands-on and 8 hours in total. A peer-reviewed paper describes wild-type-terminator-depleted chemistry with about 1% variant allele frequency and DNA to result within 8 hours.[8]

UltraSEEK: capture chemistry

Manufacturer-reported (2026) UltraSEEK runs a multiplex PCR, then a mutation-specific single-base extension with one capture-tagged terminator, then bead capture, wash, elution, spotting on a SpectroCHIP array and MALDI-TOF. The manufacturer states a sensitivity of 0.1% variant allele frequency for the EGFR, Colon and Melanoma panels and 1% for the Lung panel, from as little as 10 ng.[9]

These are manufacturer statements, not performance specifications for any laboratory. Each laboratory establishes its own.

One published breast-panel study of 39 mutations in 5 genes (10 ng, Typer v4.0.26.74) called a mutation when signal-to-noise was at least 6 and the z-score at least 7, labeled a call L (low-level) at z of 7 to 10 and T (high-level) above 10. Those are that study’s rules, not manufacturer specifications.[10]

Low-frequency mutation calls from signal-to-noise and z-scoreThree mini spectra with the same baseline noise: not called, called at low level, and called at high level, with the published study's signal-to-noise and z-score rules and a five-step tagged-terminator inset.Synthetic data, not from a real run.Not calledS/N 3, z 4 [I]mass-to-chargeCalled, low level (L)S/N 8, z 8.5 [I]mass-to-chargeCalled, high level (T)S/N 20, z 14 [I]mass-to-chargeIn one published study (39-mutation breast panel): a call needs signal-to-noise at least 6 and z-score atleast 7; z 7 to 10 is labeled L; above 10 is labeled T. These are that study's rules, not manufacturerspecifications.Chemistry behind the call: one capture-tagged terminator, thenTagged terminatorBead captureWashEluteSpot on chip

Three mini spectra with the same baseline: not called, called at low level (L), called at high level (T). The rule panel shows one published study's thresholds.

Text version of this figure

A spectrum with signal-to-noise 3 and z-score 4 is not called. One with signal-to-noise 8 and z-score 8.5 is called at low level, labeled L. One with signal-to-noise 20 and z-score 14 is called at high level, labeled T.

Synthetic data, not from a real run.

EpiTYPER: measuring methylation

A published EpiTYPER protocol runs bisulfite conversion, PCR with a T7 promoter tag, shrimp alkaline phosphatase, T7 transcription, RNase A cleavage after U positions (the T-reaction), a clean-up step, then MALDI-TOF. A methylated CpG unit gives a fragment about 16 Da heavier. Percent methylation is the methylated peak area divided by the sum of the methylated and unmethylated areas.[11]

In one example an unmethylated fragment sits at 4,484.85 Da and the same fragment with two methylated CpGs at 4,516.84 Da, a difference of 31.99 Da, about 2 × 16. Amplicons run 100 to 600 bp, typically 250 to 450.[12]

EpiTYPER is not named in the public pages we checked; availability is unconfirmed, so ask the manufacturer before planning work around it.

Methylation shift between two fragment peaksTwo peaks 31.99 Da apart: an unmethylated fragment and a fragment with two methylated CpG units, each about 16 Da heavier, with an illustrative area split of 70 to 30 giving 70 percent methylation.One methylated CpG unit adds about 16 DaSynthetic data, not from a real run.4,4804,5004,5204,5404,484.854,516.84unmethylatedtwo methylated CpGsabout 16 Daabout 16 Da31.99 Da [C]Percent methylationmethylated area divided bythe sum of methylated andunmethylated areaareas 70 and 30 [I]70 / (70 + 30) = 70% [K]Described from a published protocol; the manufacturer's current pages we checked do not name this application, soavailability is unconfirmed.

Unmethylated and methylated fragments 31.99 Da apart, with illustrative areas of 30 and 70, giving 70% methylation.

Text version of this figure

Two peaks at 4,484.85 and 4,516.84 Da. With illustrative areas of 70 for the methylated and 30 for the unmethylated peak, percent methylation is 70 divided by 100, or 70%.

Synthetic data, not from a real run.

Published limits, with their age

These are statements from the sources named, and several are from 2006. They are context for design decisions, not current specifications.

Published mass and plexing figures
WhatStatementSource
Read window for one chemistryabout 4,500 to 9,000 DaManufacturer-reported (2026)[3]
Spacing the system distinguishesanalytes 16 Da apartManufacturer-reported (2026)[3]
Intact detection of short single-stranded DNAabout 3 to 29 bases (about 1,000 to 8,600 Da)2006 review[1]
Practical size limitbelow about 100 bases, because longer fragments break up during MALDI2006 review[1]
Plexing in one author's hands26 to 29 at about a 95% pass rate; about 93 allele trios could fit at 16 Da spacing across the window2006 review[16]

What is not shown here

The structures of the mass-modified terminators and their exact mass shifts, the materials and settings used in clean-up, how reactions are dispensed, and the acquisition settings of the instrument are not on this site. They belong to the manufacturer and to your validated method.

We do not reproduce manufacturer manuals, procedures or error codes. Ask the manufacturer or use your customer portal.

Contamination control

Keep pre- and post-amplification work apart: one-way flow, controls, and a floor-plan sketch tool.

PCR makes billions of copies of a short stretch of DNA. Every later stage of a MassARRAY run works with those copies, so the main contamination risk is carry-over from finished reactions into new ones. The defense is organizational: keep the work before PCR and the work after PCR apart, and let people, plates and tools move one way.

Everything here is general good practice for molecular laboratories. None of it is a manufacturer requirement, and none of it replaces your own SOP or your biosafety rules.

Why post-PCR product matters

A single reaction tube of amplified product holds far more target DNA than a sample does. A tiny aerosol or a contaminated glove can move enough of it into a new reaction to give a false signal.

The stages from SAP treatment onward handle amplified product. In MassARRAY terms that is stages 4 to 10 on the stages page: SAP, extension, clean-up, dispensing, acquisition and the plate itself afterwards. The no-template control is the wells that tell you whether it has happened.

One-way workflow and separate areas

APHL guidance on molecular laboratory space recommends unidirectional workflow that separates the pre-PCR stages, with at least three enclosed rooms (reagents; sample preparation and assay setup; PCR and post-amplification) or physical separation using dead-air boxes or PCR workstations, and positive air pressure in the reagent rooms.[13]

CDC’s MMWR says CLIA requires unidirectional workflow with separate areas for specimen preparation, amplification and detection for open amplification systems, and that closed systems are exempt.[13] Whether a given MassARRAY workflow counts as open or closed is a question for your laboratory director and your regulator. This site does not say.

Whatever layout you use, the rule is the same: a plate that has been to the post-PCR area does not come back, and neither do its tools, pens, racks or lab coats.

One-way workflow through three zonesA floor plan with three zones in order: reagent preparation; sample preparation and assay setup; PCR and post-amplification, with one-way arrows between them and no return path. An alternative below shows dead-air boxes or PCR workstations. General good practice, not a manufacturer requirement.Three enclosed rooms, one directionReagent preparationzone 1Sample preparation andassay setupzone 2PCR andpost-amplificationzone 3People and materials move forward onlyno return pathOr: dead-air boxes or PCR workstationsReagent boxSample boxPCR workstationGeneral good practice (APHL, CDC), not a manufacturer requirement. Layout is schematic; no airflow or pressure values areshown.

Three areas with one-way arrows, and the alternative of dead-air boxes or PCR workstations. General good practice (APHL, CDC), not a manufacturer requirement. No pressure values.

Text version of this figure

Floor plan with three zones in order: reagent preparation, sample preparation and assay setup, then PCR and post-amplification. Arrows run one way from the first to the third. An alternative shows the same order using dead-air boxes or PCR workstations in one room.

Sketch your own layout

The floor-plan sketch tool is on the tools page: sketch your own layout and read the one-way flow check.

Plate and tool handling

Keep a separate set of pipettes, tips, racks, pens and lab coats for each area, and mark them so they cannot drift. Move consumables into an area, not out of it.

  • Keep plates sealed except at the bench where the step is done, and open them only after a short spin.
  • Label plate orientation on the plate and the plate map, so that a plate turned around is caught before dispensing.
  • Bring items from pre-PCR to post-PCR in one direction only. If something has to go back, treat it as contaminated and follow your SOP.
  • Clean surfaces and equipment as your own contamination SOP states. This site gives no cleaning agent and no schedule.

Waste disposal is on the manufacturer’s list of customer responsibilities.[17] Handle specimens, reagents and waste under your laboratory’s own biosafety rules.

Negative-control placement

Published validations describe reserved negative-control wells as part of typical quality control.[18] One core facility's rule asks for at least two open wells on each 384-well plate for negative controls.[19]

Place no-template controls at more than one position on the plate, for example one near each end, so a local problem such as a splash or a bad seal shows up in more than one place. Carry them through every stage with the samples.

When a negative control shows signal, stop reviewing the plate as if it were clean. Record the finding, tell the person your SOP names, and decide what to repeat before any result from that plate is used.

The plate map builder reserves control wells, and QC metrics lists what laboratories track.

What the manufacturer stated about its chip-prep unit

Manufacturer-reported (2015) A 24 August 2015 release says the chip-prep unit is designed to reduce cross-contamination from manual sample handling and lets the system run largely unattended.[20]

That is a manufacturer claim from 2015 and gives no performance figure. It does not remove the need for the controls and the one-way workflow described above.

Sources

  1. [1]
    CHM-03

    In MALDI-TOF, intact detection works for short single-stranded DNA of about 3 to 29 bases (about 1,000 to 8,600 Da); the practical limit is below about 100 bases because longer DNA fragments during MALDI. A 2006 review, not a current Agena specification.

    2006 review of MALDI-TOF DNA analysis (PubMed Central PMC1523240) (external)

    Checked on 24 November 2026Peer-reviewedmedium confidence

  2. [2]
    CHM-04

    Worked example from average residue masses: an invented 20-mer extension primer (A6 C5 G5 T4) has a mass of 6,126.05 Da; plain ddC, ddT, ddA and ddG terminators give products at 6,399.23, 6,414.25, 6,423.26 and 6,439.26 Da, with gaps of 15.02 (C to T), 9.01 (T to A) and 16.00 Da (A to G). Hypothetical sequence and average masses, not Agena's mass-modified terminators.

    Oligonucleotide Properties Calculator (average residue masses); arithmetic computed for this site, biosyn.com/Gizmo/Tools/Oligo/Oligonucleotide%20Properties%20Calculator.htm

    Checked on 24 November 2026Derived

  3. [3]
    CHM-02

    Agena states the iPLEX Pro read window as about 4,500 to 9,000 Da and says the system distinguishes analytes 16 Da apart; a 2006 review gives gaps of at least 16 Da between the four terminator products and the same window.

    Agena Bioscience: Understanding Different MassARRAY Chemistries; 2006 review of MALDI-TOF DNA analysis (PubMed Central PMC1523240), agenabio.com/resources/articles/understanding-different-massarray-chemistries/

    Checked on 24 November 2026Manufacturer-reported

  4. [4]
    DAT-09

    Cluster (allelic-discrimination) plots with tight bands and in-between points judged manually, spectra of relative intensity against mass, and plate views tinted by share of successful calls are standard in published MassARRAY work. One lab's plate color bands: red under 15%, yellow 15 to under 50%, light green 50 to under 85%, dark green 85% or more.

    Journal of Applied Laboratory Medicine 2020;5(3):454-466, iPLEX PGx validation study (external)

    Checked on 24 November 2026Peer-reviewedmedium confidence

  5. [5]
    CHM-05

    As reported in a published patent family: sodium adds about 22 Da, ammonia 17 Da (mainly on G and T), ammonia plus potassium 55 Da, and 3-HPA matrix peaks appear at 94, 138 and 188 Da. Ammonia adducts made one assay repeatedly call heterozygotes, sodium adducts caused a false insertion call, and an ascorbic-acid additive gave 8% more calls in the example and cut adducts by up to 40%. Vendor-reported example, not a general rate.

    US 7,888,127 B2 and US 10,329,612 B2 (published patents, Google Patents) (external)

    Checked on 24 November 2026Manufacturer-reportedmedium confidence

  6. [6]
    DIA-02

    Salt adducts and incomplete desalting can distort spectra and cause wrong calls (published patent, see CHM-05). Agena describes salts being removed before spotting, with desalting automated by the Chip Prep Module.

    US 7,888,127 B2 (published patent, Google Patents); Agena Bioscience: Understanding Different MassARRAY Chemistries (external)

    Checked on 24 November 2026Third partymedium confidence

  7. [7]
    CHM-06

    Standard iPLEX uses equimolar terminators, so minor alleles below roughly 10% are hard to see. An Agena patent lowers the wild-type terminator relative to the mutant (about 0.5% to under 20% of the mutant terminator, narrower 0.01-10%) and shows minor species detected down to 1.25%. The patent does not name iPLEX HS.

    US 10,233,489 B2 (published patent, Google Patents) (external)

    Checked on 24 November 2026Manufacturer-reportedmedium confidence

  8. [8]
    CHM-07

    iPLEX HS (Agena page): mutant-enriched single-base extension, about 1% variant allele frequency, as little as 10 ng DNA, 80-120 bp amplicons, about 1 hour hands-on and 8 hours total. A peer-reviewed paper describes wild-type-terminator-depleted chemistry with about 1% VAF and DNA to result within 8 h.

    Agena Bioscience: Understanding Different MassARRAY Chemistries; PubMed Central article PMC5604939; Agena iPLEX HS launch (SelectScience), agenabio.com/resources/articles/understanding-different-massarray-chemistries/

    Checked on 24 November 2026Manufacturer-reportedmedium confidence

  9. [9]
    CHM-08

    UltraSEEK: multiplex PCR, then mutation-specific single-base extension with one capture-tagged terminator, bead capture, wash, elution, spotting on a SpectroCHIP Array and MALDI-TOF. The manufacturer states a sensitivity of 0.1% variant allele frequency for the EGFR, Colon and Melanoma panels and 1% for Lung, from as little as 10 ng. These are manufacturer statements, not performance specifications for any laboratory.

    Agena Bioscience: Understanding Different MassARRAY Chemistries; PubMed Central article PMC10487510; Journal of Molecular Diagnostics (2015), DOI 10.1016/j.jmoldx.2015.08.001, agenabio.com/resources/articles/understanding-different-massarray-chemistries/

    Checked on 24 November 2026Manufacturer-reportedmedium confidence

  10. [10]
    DAT-05

    UltraSEEK breast panel study (39 mutations in 5 genes, 10 ng, Typer v4.0.26.74): a mutant call needed signal-to-noise of at least 6 and a z-score of at least 7; z 7-10 is labeled L (low-level) and above 10 T (high-level). Output is a per-sample, per-mutation matrix with T/L codes plus spectra of relative intensity against mass-to-charge. These are the authors' rules, not Agena specifications.

    UltraSEEK breast panel study (PubMed Central PMC8558178) (external)

    Checked on 24 November 2026Peer-reviewedmedium confidence

  11. [11]
    CHM-09

    Published EpiTYPER protocol: bisulfite conversion, PCR with a T7 promoter tag, shrimp alkaline phosphatase, T7 transcription, RNase A cleavage after U positions (T-reaction), a clean-up step, then MALDI-TOF. A methylated CpG unit gives a fragment about 16 Da heavier; percent methylation is the methylated peak area over methylated plus unmethylated area.

    Frontiers in Genetics (2015) EpiTYPER methylation protocol article, DOI 10.3389/fgene.2015.00287 (external)

    Checked on 24 November 2026Peer-reviewedmedium confidence

  12. [12]
    CHM-10

    EpiTYPER examples: an unmethylated fragment at 4484.85 Da and the same fragment with two methylated CpGs at 4516.84 Da (difference 31.99 Da, about 2 x 16); amplicons 100-600 bp (typically 250-450); default analysis window 1.5-7 kDa; 126 triplicate measurements per 384-well plate; example QC cut-offs of 0.1 replicate SD and 80% CpG-unit success. Window and thresholds are lab or software defaults.

    2006 review of MALDI-TOF DNA analysis (PubMed Central PMC1523240); Frontiers in Genetics (2015) EpiTYPER methylation protocol article (external)

    Checked on 24 November 2026Peer-reviewedmedium confidence

  13. [13]
    LAB-19

    APHL guidance on molecular laboratory space recommends unidirectional workflow separating pre-PCR stages, at least three enclosed rooms (reagents; sample preparation and assay setup; PCR and post-amplification) or physical separation using dead-air boxes or PCR workstations, and positive air pressure in reagent rooms. CDC's MMWR says CLIA requires unidirectional workflow with separate areas for specimen preparation, amplification and detection for open amplification systems; closed systems are exempt.

    APHL: Molecular Workflow (newborn screening resources); CDC MMWR Recommendations and Reports 58(RR-6), aphl.org/programs/newborn_screening/Pages/Molecular-Workflow.aspx

    Checked on 24 November 2026Third partymedium confidence

  14. [14]
    CHM-01

    iPLEX is locus-specific PCR, then shrimp alkaline phosphatase treatment to neutralize leftover nucleotides, then single-base extension of a primer annealing directly next to the SNP using mass-modified dideoxynucleotide terminators, then MALDI-TOF. The mass of the extended primer shows which allele was present.

    SNP genotyping using the Sequenom MassARRAY iPLEX platform (Current Protocols in Human Genetics, 2009); PubMed Central article PMC5604939 (external)

    Checked on 24 November 2026Peer-reviewed

  15. [15]
    DAT-02a

    In the same validation as DAT-02, average mass resolution was about 850 (m/delta m, FWHM).

    Journal of Applied Laboratory Medicine 2020;5(3):454-466, iPLEX PGx validation study (external)

    Checked on 24 November 2026Peer-reviewedmedium confidence

  16. [16]
    CHM-11

    A 2006 review reports practical plexing of 26-29 at about 95% pass rate in the author's hands and estimates about 93 allele trios could fit at 16 Da spacing across 4,500-9,000 Da; spots were 15 nL on a silicon-based chip, ten chips per batch (3,840 reactions). Author's experience in 2006, not a current limit.

    2006 review of MALDI-TOF DNA analysis (PubMed Central PMC1523240) (external)

    Checked on 24 November 2026Peer-reviewedmedium confidence

  17. [17]
    WAR-03

    The warranty page makes these customer responsibilities: periodic maintenance and calibration, replacing hardware components such as dispense pins, liquid-handler tips and clean resin tools, data backup, networking and waste disposal. Lasers are excepted, so laser replacement is not a customer task.

    Agena Bioscience: System Warranty, agenabio.com/support/system-warranty/

    Checked on 24 November 2026Manufacturer-reportedmedium confidence

  18. [18]
    DAT-10

    Published validations describe typical QC practice: reserved negative-control wells, reference samples of known genotype, duplicates to check discordance, an identity or sex marker, and second-person review. Acceptance numbers come from each laboratory's own validation; no numeric limit is published as a standard.

    SNPflow quality-control article (PLoS ONE, DOI 10.1371/journal.pone.0059508); Journal of Applied Laboratory Medicine 2020;5(3):454-466 (external)

    Checked on 24 November 2026Peer-reviewedmedium confidence

  19. [19]
    WFL-07

    One university core facility's own submission rules, not Agena specifications: elute DNA in water rather than TE because the instrument is salt-sensitive; 260/280 about 1.7-2.0; 1 uL at 10 ng/uL per reaction; at least two open wells per 384-well plate for negative controls; up to 36-plex (25-30 typical); amplicons about 100 bp; three unlabeled primers of 15-25 bases per SNP; no known polymorphism within 20 bp of the SNP.

    SNP genotyping by MassARRAY (university genetics core resource page); Methods in Molecular Biology chapter on iPLEX Gold genotyping (2011), cores.arizona.edu/facilities/azgc/resources/snp-genotyping-mass-array

    Checked on 24 November 2026Third partylow confidence

  20. [20]
    WFL-14

    Agena's 24 August 2015 Chip Prep Module release says the module is designed to reduce cross-contamination from manual sample handling and lets the system run largely unattended. Manufacturer claim from 2015; no performance figure is given.

    Agena Bioscience introduces front-end automation for its MassARRAY System for genetic analysis (press release, 24 August 2015), agenabio.com/company/press-release/agena-bioscience-introduces-front-end-automation-for-its-massarray-system-for-genetic-analysis/

    Checked on 24 November 2026Manufacturer-reportedmedium confidence