Home

Independent instrument service for MassARRAY systems

Close view of a metal stage assembly with cables and connectors inside an instrument.

Repair and installation

Records, maintenance and questions

Service reports and records, preventive maintenance, PC and operating-system changes, and answers to common questions.

Checked on 24 November 2026

Covers stay closed. Work inside an instrument is for a qualified service engineer only. Safety statement

Triage, returns and the service lines are on Repair and installation.

On this page

Service reports and records

What a service report records, what it does not declare, and which records your laboratory should keep.

Anatomy of a service report

Twelve headings. Every report has all of them, even when a heading reads None.

Report and request
Report ID, the request reference, customer and site, dates and times on and off site, engineer names, and the customer representative who was present.
Instrument
Identifiers as printed and software and operating-system versions as recorded.
Decontamination
The declaration reference and the engineer's check against the visible condition of the unit.
Reported fault
The fault in the customer's words.
Findings
What was found, in plain statements.
Work performed
What was done, in order.
Parts replaced
Description, source according to the business's own policy, and lot where known.
Verification
The protocol the checks came from, and the results recorded. Recorded, not judged.
Open items and limitations
What was not done, what could not be checked, and why.
Recommendations
What we suggest, separate from what we did.
Impact assessment
A note for your own change control. The customer decides what it means.
Sign-offs and attachments
The engineer and the customer representative, and a list of attached files.
What the report says about itself

A service report records what was done and measured. It is not a declaration that an instrument is validated or fit for clinical use.

The records pack

Evidence belongs in one place, with an index that says where each piece is.

A records pack is the set of documents that lets someone else follow what happened to an instrument: the installation record, qualification records, each service report, maintenance log entries, change records and the declarations. The evidence index lists the document, where it is held, its version, its date and who reviewed it.

How long you keep records is your laboratory’s policy, so keep your own copy of each one. How long we keep request data is set out in the privacy notice.

How records support change control

Laboratories working under CLIA document maintenance and function checks, and equipment results must be within the manufacturer’s limits before patient testing[1]. A service report is evidence for that file. It is not a decision.

One vendor’s published pattern repeats operational qualification after reactive service, a software upgrade or preventive maintenance, and performance qualification after a qualified major repair[2]. Use the report’s impact note to open a change record, then decide which checks to repeat. See the pattern.

For a regular performance check, the performance-verification record template has blank metric and limit columns for you to set.

Verification after servicescope

Checks after repair against acceptance criteria that you set, recorded in the service report.

Regions: United States, Germany Lead time: Confirmed at triage.

Preventive maintenance

How a maintenance program is recorded and reviewed. Intervals and tasks come from your manufacturer's documents, not from us.

Customer side and engineer side

The manufacturer's own warranty draws the line. We follow it and add nothing.

The warranty makes periodic maintenance and calibration, replacing hardware components such as dispense pins and liquid-handler tips, data backup, networking and waste disposal the customer’s responsibility, and excepts lasers[3]. It does not list the tasks or their intervals in the public page, so we publish none.

Who does what, as the warranty page assigns it
TaskWhose it isNote
Periodic maintenance and calibrationCustomerTasks and intervals come from your manufacturer's guide. We name none.
Replacing dispense pins and liquid-handler tipsCustomerFollow your manufacturer's guide for how and when.
Data backupCustomerInclude the instrument PC. Check that a copy can be read.
NetworkingCustomerRecord who changes the network and when.
Waste disposalCustomerUnder your own biosafety rules.
Laser replacementEngineer onlyLasers are excepted from the customer tasks.
Anything inside the coversEngineer onlySee the safety statement.

Intervals come from your guide

Tasks and intervals are in your manufacturer’s own maintenance documents, and we do not reproduce them. Laboratories working under CLIA perform and document the manufacturer’s maintenance and function checks at least as often as the manufacturer specifies, and results must be within the manufacturer’s limits before patient testing; modified or non-manufacturer equipment needs the laboratory’s own written protocols[1].

The manufacturer sells service plans that list preventive maintenance; the page defines none of the terms[7]. If you hold one, the plan comes first.

Preventive maintenancescope

Scheduled maintenance to a documented schedule, with each visit entered in a maintenance log.

We work only from protocols and documents the customer supplies and holds lawfully; we do not supply manufacturer protocols.

Regions: United States, Germany Lead time: Confirmed at triage.

What a maintenance visit report contains

  • Instrument identifiers as printed, and software versions as recorded.
  • The document and version each task came from.
  • Each task performed, by whom, and the result recorded.
  • Parts replaced, with description, source and lot where known.
  • Verification performed, and a reference to the protocol used.
  • Open items and limitations.
  • An impact note for your own change control. You decide what it means.

After maintenance, one vendor’s published pattern repeats operational qualification[2]. See when qualification is usually repeated.

Build your maintenance log

Type each task and interval from your own documents. This page supplies none.

  1. Task 1
  2. Task 2
  3. Task 3

Nothing here is saved or sent. Close this page and the rows are gone, so download or print first.

Ask about a maintenance visit, or send your maintenance log for a second look. Request service

PC and operating-system changes

A change checklist for the instrument PC and operating system, so that a change is planned, backed up and recorded.

Why a PC change is not routine

The instrument PC runs the instrument. Changing it changes the system you qualified.

In the public pages we checked on 24 November 2026, we could not find a compatibility statement for operating-system or PC changes. The manufacturer’s service-plans page links a brochure titled Windows 10 upgrade and two others; we have not read their contents, so we can cite only the titles[4]. We therefore state nothing about which operating system or PC suits which software version.

Ask the manufacturer for a written compatibility statement for your software version before you change anything, and file the answer with the change record.

Four rules we follow

  1. Back up first. Data backup is the customer’s responsibility under the manufacturer’s warranty[3]. We ask for a backup before any work and ask you to confirm it can be read.
  2. List what is on the PC. The warranty excludes third-party hardware or software the manufacturer has not accepted in writing[5]. Know what you have installed and who decided to keep it.
  3. Plan the way back. A rollback plan with a named person is part of the change record.
  4. Verify, then sign off. After the change, run controls and a performance verification against criteria your laboratory set in advance.

License questions go to the manufacturer. We never describe ways around a license, a password or a protection on the instrument or its software.

Change checklist

Twelve lines in three groups. Your answers stay in this browser unless you download or print them.

0 of 12 lines have a status0 met0 not met

Before the change

  • Current software and operating-system names and versions recorded

    Copy them as shown on the PC. Do not abbreviate.

  • Written compatibility statement requested from the manufacturer

    Record the date you asked and the reference you were given.

  • License questions put to the manufacturer in writing

    License terms are the manufacturer's to answer. We never describe ways around them.

  • Third-party hardware or software listed and decided

    List anything on the PC that the manufacturer has not accepted in writing, and who decided to keep or remove it.

  • Backup of everything your manufacturer's guide lists

    Backup is a customer responsibility. Record where the copy is held and who made it.

  • Backup confirmed readable

    Record how you checked and who checked.

  • Rollback plan agreed

    Name who carries it out and what triggers it.

During the change

  • Network and security notes recorded

    Record who changed what, including accounts, firewall and remote-access settings.

  • Person making the change and date recorded

    Include the supplier's name if an outside person makes the change.

After the change

  • Verification run against your own criteria

    Controls and a performance verification run, with acceptance criteria that your laboratory set in advance.

  • Result reviewed and signed off by the person your procedure names

  • Change record opened and linked to the verification evidence

Software and PC support

Where we offer help with planning, backup and the change record, the scope is stated below.

Software and PC supportscope

Help with the instrument computer and software environment, within the licences you hold.

Regions: United States, Germany Lead time: Confirmed at triage.

Describe the change you are planning and the software you run today. Request service

Questions and answers

Twenty plain answers on warranty, qualification, declarations, parts, moves and what we do not offer.

Select a question to open the answer. Each answer says what it rests on.

Agena Bio is an independent service and validation-support provider. It is not affiliated with, sponsored by, endorsed by or certified by Agena Bioscience, Inc. or Mesa Laboratories, Inc., and this site does not claim Agena authorization.

The manufacturer’s warranty excludes a system repaired, modified or moved by anyone other than the manufacturer or a provider the page leaves undefined[6]. Read your own agreement before booking.

Read your plan first. The manufacturer’s plans are its own product and ours is separate. Its published tiers list maintenance, response and qualification items and define none of them[7], so we cannot tell you what yours allows. If in doubt, ask the manufacturer in writing before you book.

The manufacturer documents you hold lawfully, a completed site-readiness checklist, a signed decontamination declaration for any used unit, access arrangements, and a named customer representative who will be present. See the installation page.

We execute and document qualification tasks against your protocol or documents you supply. We do not set acceptance criteria and we do not declare an instrument validated or fit for clinical use.

For a laboratory, accuracy, precision and other performance characteristics are established or verified by the laboratory before it reports patient results[8].

Confirmed at triage.

Because the person who handles a unit has to know what they are handling. The manufacturer asks its customers for written notice of decontamination before a visit[9], and other instrument companies use a declaration form with some returns[10].

The manufacturer’s warranty assigns replacing hardware components such as dispense pins and liquid-handler tips to the customer[3]. Follow your manufacturer’s guide for how and when. Covers stay closed.

Possibly, after review of each request. Additional regulatory considerations may apply. The manufacturer states that apart from MassARRAY Dx, its products are for research use only[11].

No. We do not sell reagents, chips or compatible consumables, and we never advise reusing a chip. The manufacturer’s warranty excludes reused or refurbished chips[5].

From the manufacturer. Its support page names a customer portal for existing customers, with online manuals and safety data sheets[12].

We do not reproduce manufacturer manuals, procedures or error codes. Ask the manufacturer or use your customer portal.

Data backup is a customer responsibility under the manufacturer’s warranty[3]. We ask for a backup before any work. Software license questions go to the manufacturer. See PC and operating-system changes.

United States and Germany

Yes. See Move an instrument for the plan, the checklist and the warranty note.

Yes, as remote triage that classifies the problem. We do not guide anyone into opening a cover. See repair request and triage.

Research-use context. No diagnostic or clinical claims. Read the research-use notice

The manufacturer states that apart from MassARRAY Dx, its products are for research use only[11]. Your laboratory director decides, and the laboratory establishes or verifies performance before reporting patient results[8].

The manufacturer’s service-plans page links a brochure with that title. We have not read its contents, so we can cite only the title[4]. We cannot state compatibility. Ask the manufacturer in writing.

Ask us for details before you book a visit.

A quotation follows triage, so it fits your instrument and your site, and nothing starts before your go-ahead.

General information; not legal, regulatory or clinical advice.

Your question is not here? Ask it in the request form, or contact us. Request service

Notices

Research-use context. No diagnostic or clinical claims. Read the research-use notice

We do not reproduce manufacturer manuals, procedures or error codes. Ask the manufacturer or use your customer portal.

Ready to ask?

Describe the instrument and what you need. Leave out patient and sample data.

Request service

Sources

  1. [1]
    LAB-06

    CLIA 42 CFR 493.1254: for unmodified manufacturer-supported equipment the laboratory performs and documents the manufacturer's maintenance and function checks at least as often as the manufacturer specifies, with results within the manufacturer's limits before patient testing; modified or non-manufacturer equipment needs the laboratory's own written protocols.

    42 CFR 493.1254 Standard: Maintenance and function checks (eCFR) (external)

    Checked on 24 November 2026Regulator

  2. [2]
    LAB-17

    Another instrument vendor's public qualification pattern: IQ confirms installation to specification and safety rules, OQ confirms operation within operating specifications, and PQ follows a qualified major repair. IQ is repeated at install, after relocation and reinstallation and after environment change; OQ after reactive service, software upgrade or preventive maintenance.

    Instrument qualification services page (another instrument vendor), Other-vendor public page

    Checked on 24 November 2026Third partymedium confidence

  3. [3]
    WAR-03

    The warranty page makes these customer responsibilities: periodic maintenance and calibration, replacing hardware components such as dispense pins, liquid-handler tips and clean resin tools, data backup, networking and waste disposal. Lasers are excepted, so laser replacement is not a customer task.

    Agena Bioscience: System Warranty, agenabio.com/support/system-warranty/

    Checked on 24 November 2026Manufacturer-reportedmedium confidence

  4. [4]
    SUP-06

    Agena's service-plans page links three brochures titled Customer Support Solutions, Windows 10 upgrade and Performance Verification Kit. Their contents were not read, so only the titles can be cited.

    Agena Bioscience: Service Plans, agenabio.com/support/service-plans/

    Checked on 24 November 2026Manufacturer-reportedmedium confidence

  5. [5]
    WAR-04

    The warranty excludes reused or refurbished chips (SpectroCHIP Arrays are named), third-party hardware or software Agena has not accepted in writing, exposure to BSL-3/4 agents or radioactivity, malicious software, and acts of God.

    Agena Bioscience: System Warranty, agenabio.com/support/system-warranty/

    Checked on 24 November 2026Manufacturer-reportedmedium confidence

  6. [6]
    WAR-02

    The warranty page excludes a system that is repaired, modified, altered, assembled, disassembled or moved from its original installation location by anyone other than Agena or a service provider the page leaves undefined. It is a coverage exclusion, so customers should check their own agreement.

    Agena Bioscience: System Warranty, agenabio.com/support/system-warranty/

    Checked on 24 November 2026Manufacturer-reported

  7. [7]
    SUP-01

    Agena's service-plan page lists an Essential tier (preventive maintenance, on-site applications support, fast response) and a Plus tier that adds applications training, monthly performance verification kits, next-day response and qualification services. Both cover the system and the Chip Prep Module; no prices, hours or definitions are given.

    Agena Bioscience: Service Plans, agenabio.com/support/service-plans/

    Checked on 24 November 2026Manufacturer-reportedmedium confidence

  8. [8]
    LAB-04

    CLIA 42 CFR 493.1253: for an unmodified FDA-cleared test a laboratory verifies accuracy, precision and reportable range and confirms reference intervals; for modified tests and tests not subject to FDA clearance it also establishes analytical sensitivity and specificity (including interfering substances) and any other needed characteristic before reporting patient results.

    42 CFR 493.1253 Standard: Establishment and verification of performance specifications (eCFR) (external)

    Checked on 24 November 2026Regulator

  9. [9]
    WAR-05

    Before a service visit the customer must tell Agena in writing about any decontamination of the system, a customer representative must be on site throughout on-site service, and no product or component may be returned without Agena's prior consent.

    Agena Bioscience: System Warranty, agenabio.com/support/system-warranty/

    Checked on 24 November 2026Manufacturer-reportedmedium confidence

  10. [10]
    LAB-18

    Relocation and return patterns at other instrument companies: one describes relocation in four stages (pre-move consultation, insured transport, de-installation and re-installation by engineers, testing and calibration) and is silent on decontamination; another requires a decontamination declaration form with some returns, a reference number, item condition and a detailed fault description.

    How easy is it to relocate a mass spectrometer? (vendor community blog, 1 June 2017); goods-return page (another instrument vendor), Other-vendor public page

    Checked on 24 November 2026Third partymedium confidence

  11. [11]
    REG-01

    Agena states that, apart from MassARRAY Dx, all its products are for research use only and not for use in diagnostic procedures.

    Agena Bioscience: MassARRAY System product page, agenabio.com/products/massarray-system/

    Checked on 24 November 2026Manufacturer-reported

  12. [12]
    SUP-02

    Agena's support page names a Customer Support Portal for existing customers (complaint logging, online manuals, safety data sheets, product training) and regional help desks for the Americas, Australia/New Zealand, China and Asia Pacific, and Europe.

    Agena Bioscience: Support, agenabio.com/support/

    Checked on 24 November 2026Manufacturer-reportedmedium confidence