Regulatory context

A dated, neutral summary of the regulatory framework a laboratory works under. General information, not advice.

Checked on 24 November 2026
On this page

Dated events

Status as of 24 November 2026

Each row says where a topic stood on the date we last checked. Regulatory positions move, so check the source before you rely on a row.

Regulatory status as of 24 November 2026
TopicWhere it stands
FDA rule on laboratory-developed tests (LDTs)FDA’s final rule of 6 May 2024 (89 FR 37286) was vacated by the U.S. District Court for the Eastern District of Texas on 31 March 2025. An FDA rule published 19 September 2025 restored the earlier text of 21 CFR 809.3(a)[1].
FDA and commercially distributed componentsOne law-firm advisory from July 2026 says FDA does not regulate the laboratory service itself but still asserts oversight of commercially distributed components, with research-use and investigational-use labeling as an enforcement tool. This rests on a single secondary source [6].
Congress: H.R. 8890The Enhancing CLIA Act of 2026 was introduced on 19 May 2026. It would place LDT oversight under an expanded CLIA framework. We found no later action [4].
CLIA request for informationCMS and CDC published a request for information on 16 July 2026. Comments closed on 14 September 2026. A request for information is not a proposed rule [5].
CLIA: performance specificationsFor an unmodified FDA-cleared test a laboratory verifies accuracy, precision and reportable range and confirms reference intervals. For a modified test, or one not subject to FDA clearance, it also establishes analytical sensitivity and specificity, including interfering substances, and any other characteristic needed before reporting patient results[13].
CLIA: control proceduresDaily minimum controls are one positive and one negative for qualitative tests, and two controls for tests with an extraction step, one of which can detect extraction errors. For molecular amplification there are two controls plus an inhibition control where inhibition is a significant cause of false negatives [14].
CLIA: maintenance and function checksFor unmodified, manufacturer-supported equipment, the laboratory performs and documents the manufacturer’s maintenance and function checks at least as often as the manufacturer specifies, with results within the manufacturer’s limits before patient testing. Modified or non-manufacturer equipment needs the laboratory’s own written protocols [15].
New York StateA clinical laboratory permit and a test-specific review for laboratory-developed tests are required for specimens from New York. The review sits inside the permit process [16].
ISO 15189:2022The transition period ended in December 2025. UKAS flags clinical involvement in reviewing verification reports, and periodic review of examination methods (clauses 7.3.1 and 7.3.2), as common nonconformities [2].
CLSI guidanceCLSI MM17 (second edition, 31 May 2018) covers validation and verification of qualitative multiplex nucleic acid assays for genotyping and pathogen detection, as good-practice recommendations rather than requirements. CLSI QMS02 (seventh edition, 11 March 2024) is a guideline on developing and managing laboratory documents [17].

Research-use labeling

The manufacturer states that, apart from one diagnostic-labeled product, its products are for research use only[7]. Its own pages disagree on which product is the exception, so the point is unsettled[8]. A laboratory that uses a research-use instrument in a test it reports on patients establishes the performance of that test itself, as the CLIA rows above describe. The research-use notice has the plain-language version.

Servicing and remanufacturing

FDA separates servicing, meaning repair or routine maintenance that returns a device to the maker’s specifications and intended use, from remanufacturing, meaning acts that significantly change performance, safety specifications or intended use. It judges by what is actually done, not by what an entity calls itself[9].

The revised 21 CFR Part 820, the Quality Management System Regulation, took effect on 2 February 2026 and incorporates ISO 13485:2016 by reference [3]. Section 820.1 lists installing and servicing among covered activities for finished devices and names remanufacturers. FDA’s preamble says it kept that scope and declined to add third-party servicers and refurbishers, so whether a given servicer is in scope is a question for counsel[10].

This site presents service as process and records. We do not modify or upgrade instruments.

Diagnostic-labeled products

If a product is labeled for diagnostic use, the labeling, the supplier’s obligations and the laboratory’s verification duties differ from those for a research-use product. The manufacturer announced diagnostic-labeled configurations for Europe in 2016 and 2017. We cannot confirm their current status, so we state none[11].

The only Agena-related 510(k) record we found was filed by Sequenom, Inc., not by Agena: K132978, a Factor V Leiden and Factor II genotyping test, with a decision date of 13 June 2014. We draw no conclusion from it about any current product [12].

We also work on diagnostic-labeled units. Additional regulatory considerations apply, so tell us the exact product when you request service.

What a service record is

A laboratory that must document maintenance and function checks can use a service report as one input to its own records. It does not discharge the laboratory’s duties, and it does not show that an assay or an instrument is suitable for any use.

A service report records what was done and measured. It is not a declaration that an instrument is validated or fit for clinical use.

Sources

  1. [1]
    LAB-07

    FDA's 6 May 2024 final rule on laboratory-developed tests (89 FR 37286) was vacated by the U.S. District Court for the Eastern District of Texas on 31 March 2025 (ACLA v. FDA, No. 4:24-CV-479-SDJ); an FDA rule published 19 September 2025 restored the earlier text of 21 CFR 809.3(a).

    Laboratory Developed Tests (FDA); Federal Register rules of 6 May 2024 and 19 September 2025 (external)

    Checked on 24 November 2026Regulator

  2. [2]
    LAB-12

    The ISO 15189:2022 transition period ended in December 2025. UKAS flags clinical involvement in reviewing verification reports, and periodic review of examination methods (clauses 7.3.1 and 7.3.2), as common nonconformities.

    ISO 15189 transition update (UKAS, 6 June 2024) (external)

    Checked on 24 November 2026Third partymedium confidence

  3. [3]
    LAB-02

    The revised 21 CFR Part 820, the Quality Management System Regulation (QMSR), took effect on 2 February 2026 and incorporates ISO 13485:2016 by reference.

    Quality Management System Regulation final rule (Federal Register, 2 February 2024, 2024-01709) (external)

    Checked on 24 November 2026Regulator

  4. [4]
    LAB-09

    H.R. 8890, the Enhancing CLIA Act of 2026, was introduced on 19 May 2026 by Rep. Neal Dunn and would place laboratory-developed-test oversight under an expanded CLIA framework; no later action was found.

    H.R. 8890 (119th Congress), Enhancing CLIA Act of 2026, as introduced (external)

    Checked on 24 November 2026Regulator

  5. [5]
    LAB-10

    CMS and CDC jointly published a CLIA Request for Information on 16 July 2026 (Federal Register document 2026-14358, CMS-3485-NC); comments closed on 14 September 2026. An RFI is not a proposed rule.

    Request for Information: Clinical Laboratory Improvement Amendments of 1988 (CLIA) Regulations (Federal Register, 16 July 2026) (external)

    Checked on 24 November 2026Regulator

  6. [6]
    LAB-08

    A July 2026 law-firm advisory says FDA does not regulate the laboratory service itself but still asserts oversight of commercially distributed components, with research-use and investigational-use labeling as an enforcement tool. Single secondary source.

    Oversight of Laboratory-Developed Tests One Year After ACLA v. FDA (law-firm advisory, July 2026), arnoldporter.com/en/perspectives/advisories/2026/07/oversight-of-laboratory-developed-tests-one-year-after-acla-v-fda

    Checked on 24 November 2026Third partymedium confidence

  7. [7]
    REG-01

    Agena states that, apart from MassARRAY Dx, all its products are for research use only and not for use in diagnostic procedures.

    Agena Bioscience: MassARRAY System product page, agenabio.com/products/massarray-system/

    Checked on 24 November 2026Manufacturer-reported

  8. [8]
    REG-02

    Agena's own pages disagree on the exception to research use only: the footer names MassARRAY Dx, while the website-terms body names the IMPACT Dx System. The point is unsettled.

    Agena Bioscience: Terms and Conditions; Agena Bioscience: Product Terms and Conditions, agenabio.com/terms-and-conditions/

    Checked on 24 November 2026Manufacturer-reported

  9. [9]
    LAB-01

    FDA separates servicing (repair or routine maintenance that returns a device to the maker's specifications and intended use) from remanufacturing (acts that significantly change performance, safety specifications or intended use), judging by what is actually done rather than what an entity calls itself.

    Remanufacturing and Servicing of Medical Devices (FDA) (external)

    Checked on 24 November 2026Regulator

  10. [10]
    LAB-03

    21 CFR 820.1 lists installing and servicing among covered activities for finished devices, names remanufacturers and excludes makers of components or parts. FDA's QMSR preamble says it kept that scope and declined to add third-party servicers and refurbishers; whether a given servicer is in scope is a question for counsel.

    21 CFR 820.1 Scope (eCFR); Quality Management System Regulation final rule (Federal Register, 2024-01709) (external)

    Checked on 24 November 2026Regulator

  11. [11]
    REG-03

    On 4 April 2016 Agena announced a MassARRAY Dx Analyzer 4, a Dx Nanodispenser RS 1000 and Colon and Lung panels for diagnostic use in Europe only, not sold in the United States; a 29 June 2017 release added a CE-IVD system with the Chip Prep Module. Historical; current status is unverified.

    Agena Bioscience launches in vitro diagnostic tests for lung and colon cancer in Europe (4 April 2016); IVD genetic testing system launched in Europe (29 June 2017), agenabio.com/company/press-release/agena-bioscience-launches-vitro-diagnostic-tests-lung-colon-cancer-europe/

    Checked on 24 November 2026Manufacturer-reportedmedium confidence

  12. [12]
    REG-06

    The only Agena-related 510(k) found is K132978, the IMPACT DX Factor V Leiden and Factor II Genotyping Test, applicant Sequenom, Inc. (not Agena), decision date 13 June 2014, Substantially Equivalent, product code PHJ, 21 CFR 864.7280.

    FDA 510(k) Premarket Notification database entry K132978 (external)

    Checked on 24 November 2026Regulator

  13. [13]
    LAB-04

    CLIA 42 CFR 493.1253: for an unmodified FDA-cleared test a laboratory verifies accuracy, precision and reportable range and confirms reference intervals; for modified tests and tests not subject to FDA clearance it also establishes analytical sensitivity and specificity (including interfering substances) and any other needed characteristic before reporting patient results.

    42 CFR 493.1253 Standard: Establishment and verification of performance specifications (eCFR) (external)

    Checked on 24 November 2026Regulator

  14. [14]
    LAB-05

    CLIA 42 CFR 493.1256: daily minimum controls are one positive and one negative for qualitative tests, two controls for tests with an extraction step (one able to detect extraction errors), and for molecular amplification two controls plus an inhibition control where inhibition is a significant cause of false negatives.

    42 CFR 493.1256 Standard: Control procedures (eCFR) (external)

    Checked on 24 November 2026Regulator

  15. [15]
    LAB-06

    CLIA 42 CFR 493.1254: for unmodified manufacturer-supported equipment the laboratory performs and documents the manufacturer's maintenance and function checks at least as often as the manufacturer specifies, with results within the manufacturer's limits before patient testing; modified or non-manufacturer equipment needs the laboratory's own written protocols.

    42 CFR 493.1254 Standard: Maintenance and function checks (eCFR) (external)

    Checked on 24 November 2026Regulator

  16. [16]
    LAB-11

    New York State requires a clinical laboratory permit and a test-specific review for laboratory-developed tests on specimens from New York; that review sits inside the permit process.

    Wadsworth Center (New York State Department of Health): clinical laboratory permit and test review information (external)

    Checked on 24 November 2026Regulator

  17. [17]
    LAB-13

    CLSI MM17 (2nd edition, 31 May 2018) covers validation and verification of qualitative multiplex nucleic acid assays for genotyping and pathogen detection, as good-practice recommendations rather than requirements; CLSI QMS02 (7th edition, 11 March 2024) is a guideline on developing and managing laboratory documents.

    CLSI MM17: Validation and Verification of Multiplex Nucleic Acid Assays; CLSI QMS02 (external)

    Checked on 24 November 2026Third party