Validation and SOPs

SOPs

Structure and outlines for the assay SOP, the run record and change control, with a skeleton builder and no acceptance criteria.

Checked on 24 November 2026
On this page

Start here

Structure and templates for standard operating procedures: general guidance with no acceptance criteria and no manufacturer content.

A set of documents, not one long one

A MassARRAY laboratory usually ends up with six kinds of document. Keeping them apart keeps each one short enough to follow and easy to change.

The assay procedure changes rarely and needs careful review. The run record changes every day. The maintenance procedure follows the manufacturer’s guide, which has its own update cycle. Training records belong to people, not to assays. Putting them in one document means every small change re-opens the whole thing.

The outlines here are original structure. No manufacturer or laboratory SOP for a MassARRAY system was copied or used as a source for them. Document-control practice is covered in published guidance on laboratory documents.[1]

None of the outlines contains an acceptance criterion, an interval or an instrument procedure. Those come from your manufacturer’s current documents and your own validation, and you reference them from your SOP rather than copying them in.

A set of six SOP documentsThe assay procedure at the top with five companion documents beneath it: run record, maintenance and troubleshooting procedure, change control, training and competency record, and receiving and lot-change checks.Assay procedureRefers out to every document belowRun recordMaintenance and troubleshootingprocedureChange controlTraining and competency recordReceiving and lot-change checks
Text version: the assay procedure refers out to five companion documents: the run record, the maintenance and troubleshooting procedure, change control, the training and competency record, and receiving and lot-change checks.

The six documents

Build your own skeleton

Choose the sections your procedure needs. You get a Markdown file with a heading and a prompt for each.

Sections to include (21 of 21)
Skeleton previewMarkdown

Built in your browser. Nothing is sent to us.

An open binder of blank forms with a pen, beside a rack of empty glass tubes on a laboratory bench.

SOP structure

A recommended outline for an assay SOP, section by section, with notes on what belongs in each.

Start with document control

Before the content, the part an inspector reads first: which version is this, who owns it, and is it current?

  • Document title and ID
  • Version and effective date
  • Author, reviewer and document owner
  • Next review due, as your quality system sets it
  • Where the controlled copy lives

Published guidance covers developing and managing laboratory documents as a subject in its own right.[1]

The 21 sections

One outline for the assay procedure. Delete a section only if you say why.

The 21-section outline of an assay procedure SOP
No.SectionWhat belongs hereA prompt to start with
01PurposeOne short paragraph on why the procedure exists and what it controls.State what this procedure does, for which assay, in which laboratory.
02ScopeWhat is covered and what is not: instruments, assays, specimen types, plate formats, sites.List the instruments, assay content, specimen types and plate formats covered, and what is excluded.
03Responsibilities and competencyRoles, who may perform each task, and where training and competency are recorded.Name each role, what it is responsible for, and the training record that shows competency.
04DefinitionsTerms and abbreviations used in this document, written once.Define each term and abbreviation the way your laboratory uses it.
05Safety, waste and spillsPointers to your biosafety rules and the safety data sheets. Work inside an instrument is engineer-only.Refer to your site biosafety procedure and safety data sheets. State that covers stay closed and who to call.
06Sample and DNA acceptance as defined by the laboratoryWhat the laboratory accepts, how quality and quantity are judged, and what happens to a rejected sample.Write the acceptance rules your laboratory has set, how each is measured, and the route for rejection or recollection.
07Equipment and softwareInstrument identifiers, software names and versions as installed, and where maintenance and calibration status is recorded.List each instrument and software item with its identifier and version, and the record that shows its status.
08Reagents and consumablesItems, suppliers, storage as the supplier's insert states, lot and expiry recording, receiving checks.List each item, the storage condition from the insert, and how lot and expiry are recorded and checked.
09Facility layout and contamination controlAreas, one-way flow, and what the laboratory does to keep amplified material away from set-up areas.Describe the areas and the direction of work, and refer to the cleaning and monitoring records.
10Pre-run checksWhat must be true before a run starts: instrument ready, consumables on hand, maintenance current, controls prepared.List each check, who does it, and where it is recorded.
11Procedure by stageOne subsection per stage. Each opens with a reference to the current manufacturer's instructions, by document and version.For each stage write: 'Refer to the current manufacturer's instructions', the document reference, and any local variation with its validation reference.
12Controls and acceptance criteriaWhich controls run on each plate, where the acceptance criteria are written, and what happens when one is not met.Name the controls, point to the document where criteria are set by the laboratory director, and state the action when a criterion is not met.
13Data review and interpretationWho reviews, what is reviewed, how manual review of plots and spectra is recorded, and how exclusions are decided.State who reviews, what second review covers, and where exclusions and manual calls are recorded.
14Review and disposition of results under the laboratory's own regulatory frameworkHow a result is reviewed, held, repeated or passed on, in the terms your own framework uses.Describe review and disposition of results as your regulatory framework and your director require. This outline sets nothing here.
15Troubleshooting and escalationA pointer to the maintenance and troubleshooting procedure and the escalation route.Refer to the troubleshooting procedure and name who is called and when.
16Maintenance and calibration cross-referencesWhere the maintenance log, calibration records and service reports live.List the records and their locations. Do not copy intervals into this document.
17Records, retention and backupWhat is kept, where, for how long under your policy, and how backups are made and checked.List each record type, its location, your retention rule and the backup arrangement.
18Validation summary and change controlA pointer to the validation report and the change-control procedure.Refer to the validation summary, and state which changes trigger a change record.
19ReferencesDocuments this procedure depends on, each with version.List manufacturer documents, standards and internal procedures with their versions.
20AppendicesPlate map, run record, review checklist, reagent and lot log, deviation form.Attach or reference each form used with this procedure.
21Revision historyVersion, date, what changed, who made and reviewed the change.Record every revision.

Three sections that need care

Section 6: sample and DNA acceptance as defined by the laboratory

Your laboratory decides what it accepts and writes the rule here, with how each measure is taken. One university core facility publishes its own submission rules for DNA and plate set-up; they are that facility’s rules for its own service, not specifications, and yours will differ.[WFL-07]

Section 9: facility layout and contamination control

Published guidance recommends a one-way workflow that keeps work before amplification apart from work after it, using separate rooms or physical separation such as dead-air boxes or PCR workstations. It is general good practice, not a manufacturer requirement.[3]

Section 14: review and disposition of results

The heading is deliberately neutral. What a result is called, who may pass it on and to whom depend on the framework your laboratory works under. Write that framework in your words; this outline sets nothing. Published validations also describe second-person review as typical practice.[2]

What this outline deliberately leaves out

  • Step-by-step procedures for each stage. Section 11 points to the manufacturer's current instructions by document and version.
  • Acceptance criteria and limits. Section 12 says where your laboratory director set them.
  • Volumes, times, temperatures and instrument settings.
  • Maintenance tasks and intervals. Those belong in the maintenance and troubleshooting procedure.
  • Anything taken from a manufacturer manual, however short.

Download the outline

Want only some sections? The section picker builds a shorter skeleton in your browser.

Run record

What to capture in a run record so that a run can be reconstructed and reviewed later.

One sheet per plate or chip

A run record lets someone who was not there reconstruct what was used, who did what, what went wrong and what was decided.

Blocks of a run record
BlockWhat it capturesWhy it is there
HeaderRun ID, date, operator, reviewer, assay or panel with its content version, plate or chip ID, plate format.So the record can be found, and tied to one physical plate.
Instrument blockInstrument IDs, software name and version, analysis settings reference, thermal cycler and program name.A result is only reproducible if you know what produced it.
Reagent blockEach item with supplier, lot, expiry and the date it was opened.Lets you trace a problem to a lot, and shows nothing expired was used.
DNA blockSpecimen type, quantification method, input amount as your SOP defines it, a reference to the sample list.Records what went in without putting identities on the sheet.
Plate mapA 96 or 384 grid with each well marked: sample, no-template control, positive or reference, duplicate, identity or sex marker, or empty.Published validations describe reserved negative-control wells, reference samples, duplicates and an identity or sex marker as typical practice.[2]
Stage checkpointsStart and stop times, initials and a program or batch reference for each stage.Shows the order of work and who did each part.
Deviations and reworksWhat happened, when, the action taken, any rework and who reviewed it.A deviation written down at the time is a note. One reconstructed later is a story.
Acquisition summaryStart and end, wells acquired, operator notes.Separates what the instrument did from what the software then made of it.
Analysis summaryThe metrics your SOP defines, such as call rate per sample and per assay, negative-control result, reference-sample and replicate agreement, identity or sex check, and exclusions with reasons.[4]Each metric is recorded next to the criterion from your SOP, then marked met or not met.
DispositionThe decision, in the terms your SOP names, with the reason if a run is held or repeated.The record ends with a named person's decision, not a trailing off.
Review and sign-offReviewer and, if your SOP requires it, a second reviewer.Second-person review is typical practice in published validations.

Where the numbers come from

The analysis block has columns for criteria. They stay empty on the template.

One example framework sets minimum daily controls for qualitative and molecular tests, and more where extraction or amplification is involved.[5] Your SOP says which controls your assay uses and what counts as acceptable. The run record only holds the result and the comparison.

Published tools for checking genotyping output compute figures such as call rate, observed against expected replicates and discordance, without setting numeric cut-offs.[4] Cut-offs, where you use them, are a validation decision.

Get the sheet

To plan the plate before you pipette, use the plate and chip layout page, which has a click-to-assign plate map.

The assay procedure points to this record from its appendix. See SOP structure.

Change control

How to control changes to reagents, software, instruments and methods, and when a change triggers re-qualification.

Events that should open a change record

Most instrument trouble in a validated laboratory starts with a change nobody wrote down. The list is a starting point; add the events your own quality system recognizes.

Events that can trigger a re-check
EventWhy it can matterUsual ownerThe question to answer before going ahead
Reagent lot changeA new lot can behave differently from the one you validated with.Laboratory leadDoes the lot-change check meet the criteria in your SOP? See receiving and lot-change checks.
Chip lot changeSame reason as a reagent lot: a different batch of a consumable.Laboratory leadSame question: what check do you run, and against which criteria?
New variant or new assay contentNew targets are outside what your validation covered.Laboratory directorDoes your validation plan cover it, or does part of the plan need repeating?
New specimen typeThe range you validated may not include it.Laboratory directorWhich performance studies need evidence for this specimen type?
Software updateAcquisition or analysis behavior can change between versions.Instrument owner and quality leadWhat did you record before and after, and does your pattern call for a check?
PC or operating-system changeThe instrument software runs on it, and a backup and a rollback plan are yours to make.Instrument owner and ITDid you request a written compatibility statement? See PC and operating-system changes.
RelocationThe instrument is reinstalled in a different place.Instrument ownerWhich qualification stages repeat, and who signs them off?
Repair or preventive maintenanceWork was done on the instrument, so its behavior may have changed.Instrument ownerWhat verification will you run, and against which criteria, before returning it to use?
Regulatory changeThe framework your laboratory works under changes.Laboratory directorWhich SOPs and plans use the old wording, and who updates them?

For the instrument events, another vendor’s public pattern shows which qualification stages it repeats after relocation, a software upgrade, reactive service, preventive maintenance and a qualified major repair.[6] A pattern from another vendor; your SOP and the manufacturer's documents decide The same pattern appears, with the manufacturer’s plan wording beside it, on the instrument qualification page.[7]

A change record from start to close

Six steps. The order matters: the criteria are written before the results exist.

  1. Step 1: Raise the change

    Open a record before the change happens, say what is changing and why, and name the person who owns it.
  2. Step 2: Assess the impact

    Walk through assay performance, instrument qualification, procedures, training, records, suppliers and analysis settings. Write 'yes', 'no' or 'unsure' for each, with a note.
  3. Step 3: Decide what to repeat

    Name the verification and qualification activities, who does them, by when, and what evidence they leave. Do this before the change goes live.
  4. Step 4: Do the work and keep the evidence

    Record the results against the criteria you wrote in step 3. Do not change a criterion after seeing a result.
  5. Step 5: Sign off

    A named person decides whether the change goes live, and signs. Name that person in your SOP.
  6. Step 6: Close and update

    Update the documents the change touched, brief the people it affects, and close the record with the effective date.

Where manufacturer notices fit

A notice from the manufacturer about an instrument, a software version or a lot is a change trigger from outside. Log it, summarize it in your own words, and decide what it means for you. A register keeps these together.

See manufacturer notices and template T-18 Manufacturer-notice register.

Get the record

Pair it with the qualification stage skeleton when the change affects an instrument.

Continues on Maintenance, training and receiving: the maintenance and troubleshooting procedure, the training and competency record, and receiving and lot-change checks.

General information; not legal, regulatory or clinical advice.

Sources in reduced form

These sources are cited by identifier, journal and year only. This page gives none of their figures.

  • WFL-07

    SNP genotyping by MassARRAY (university genetics core resource page); Methods in Molecular Biology chapter on iPLEX Gold genotyping (2011), cores.arizona.edu/facilities/azgc/resources/snp-genotyping-mass-array

    Checked on 24 November 2026

Sources

  1. [1]
    LAB-13

    CLSI MM17 (2nd edition, 31 May 2018) covers validation and verification of qualitative multiplex nucleic acid assays for genotyping and pathogen detection, as good-practice recommendations rather than requirements; CLSI QMS02 (7th edition, 11 March 2024) is a guideline on developing and managing laboratory documents.

    CLSI MM17: Validation and Verification of Multiplex Nucleic Acid Assays; CLSI QMS02 (external)

    Checked on 24 November 2026Third party

  2. [2]
    DAT-10

    Published validations describe typical QC practice: reserved negative-control wells, reference samples of known genotype, duplicates to check discordance, an identity or sex marker, and second-person review. Acceptance numbers come from each laboratory's own validation; no numeric limit is published as a standard.

    SNPflow quality-control article (PLoS ONE, DOI 10.1371/journal.pone.0059508); Journal of Applied Laboratory Medicine 2020;5(3):454-466 (external)

    Checked on 24 November 2026Peer-reviewedmedium confidence

  3. [3]
    LAB-19

    APHL guidance on molecular laboratory space recommends unidirectional workflow separating pre-PCR stages, at least three enclosed rooms (reagents; sample preparation and assay setup; PCR and post-amplification) or physical separation using dead-air boxes or PCR workstations, and positive air pressure in reagent rooms. CDC's MMWR says CLIA requires unidirectional workflow with separate areas for specimen preparation, amplification and detection for open amplification systems; closed systems are exempt.

    APHL: Molecular Workflow (newborn screening resources); CDC MMWR Recommendations and Reports 58(RR-6), aphl.org/programs/newborn_screening/Pages/Molecular-Workflow.aspx

    Checked on 24 November 2026Third partymedium confidence

  4. [4]
    WFL-11

    The SNPflow QC tool for iPLEX output computes call rate, expected versus observed replicates, minor allele frequency, discordant replicates, discordance rate and a Hardy-Weinberg test, checks genotype frequencies against reference data and has a sex-determination SNP module. It sets no numeric cut-offs.

    SNPflow: a lightweight application for the processing, storing and automatic quality checking of genotyping assays (PLoS ONE) (external)

    Checked on 24 November 2026Peer-reviewedmedium confidence

  5. [5]
    LAB-05

    CLIA 42 CFR 493.1256: daily minimum controls are one positive and one negative for qualitative tests, two controls for tests with an extraction step (one able to detect extraction errors), and for molecular amplification two controls plus an inhibition control where inhibition is a significant cause of false negatives.

    42 CFR 493.1256 Standard: Control procedures (eCFR) (external)

    Checked on 24 November 2026Regulator

  6. [6]
    LAB-17

    Another instrument vendor's public qualification pattern: IQ confirms installation to specification and safety rules, OQ confirms operation within operating specifications, and PQ follows a qualified major repair. IQ is repeated at install, after relocation and reinstallation and after environment change; OQ after reactive service, software upgrade or preventive maintenance.

    Instrument qualification services page (another instrument vendor), Other-vendor public page

    Checked on 24 November 2026Third partymedium confidence

  7. [7]
    DIA-06

    Qualification after service is a common pattern: another vendor lists operational qualification after reactive service or preventive maintenance and performance qualification after a qualified major repair, while Agena's Plus plan names monthly performance verification kits and qualification services without defining them.

    Instrument qualification services page (another instrument vendor); Agena Bioscience: Service Plans, Other-vendor public page

    Checked on 24 November 2026Derivedmedium confidence