Plain-language definitions of the terms used across the MassARRAY workflow, with permalinks.

Checked on 24 November 2026

Terms

Definitions are short and plain. Where a definition rests on a source, a numbered marker leads to it. Definitions marked as textbook are general knowledge. Mark symbols are left out here; the trademarks section shows them.

Type part of a term or a word from a definition.

50 terms

  • 3-HPA matrix #

    The small organic compound, 3-hydroxypicolinic acid, mixed with DNA samples on the chip so that a laser pulse can lift the DNA off as ions. See also matrix.

    Textbook definition

    Used on: Chemistry

  • Adduct #

    A small ion, such as sodium or ammonium, that sticks to a molecule and adds a little mass. It shifts or widens a peak, and when two allele products are close in mass it can look like the wrong allele.

    Textbook definition

    Used on: Chemistry, Calls and confidence

  • Allele #

    One of the alternative forms of a DNA sequence at a position, for example the A or the G at a single-base variant.

    Textbook definition

    Used on: Chemistry

  • Amplicon #

    The stretch of DNA copied by PCR between a pair of primers.

    Textbook definition

    Used on: Stages

  • Assay design #

    Choosing the PCR and extension primers for a set of variants so that every expected peak lands at its own mass. Published studies describe using the manufacturer's assay-design software and adjusting primers to keep allele peaks apart.[1]

    Used on: Stages, Assay validation roadmap

  • Call rate #

    The share of attempted genotype calls that return a call. Quality-control tools such as SNPflow compute it with replicate and Hardy-Weinberg checks, and they set no numeric limits. Your limits come from your validation.[2]

    Used on: Qc metrics, Calls and confidence

  • Calls: conservative, moderate, aggressive, low probability, user call #

    Also written: call tiers; genotype confidence

    Confidence tiers that Typer software attached to genotype calls in one published study: conservative, moderate, aggressive and low probability, plus user calls entered by the reviewer. The tier names are that study's. Your protocol decides what you accept.[3]

    Used on: Calls and confidence

  • Chip Prep Module (CPM) #

    Also written: CPM; chip-prep unit

    A unit that automates dispensing and desalting of samples onto SpectroCHIP arrays and loading them onto the analyzer. It was announced on 24 August 2015 and is tied to the 96-well system.[4]

    Used on: Stages, Contamination control

  • CLIA #

    Also written: Clinical Laboratory Improvement Amendments

    The U.S. federal rules for clinical laboratory testing, in 42 CFR part 493. They set what a laboratory verifies or establishes before it reports patient results, how it uses controls and how it maintains equipment.[5]

    Used on: Regulatory context, Research-use notice

  • Cluster plot #

    Also written: allelic-discrimination plot

    A scatter plot of the signal for each allele across samples, in which each genotype forms its own group. Points between groups are judged by a person.

    Textbook definition

    Used on: Calls and confidence, Manual review

  • CpG unit #

    In an EpiTYPER read-out, a fragment that holds one or more CpG sites. In a published protocol each methylated CpG makes its fragment about 16 Da heavier, and percent methylation is the methylated peak area divided by the methylated plus unmethylated area.[6]

    Used on: Chemistry

  • ddNTP and dideoxy terminator #

    A nucleotide that lacks the chemical group needed to add the next base, so extension stops after one base. In single-base extension the terminator added to the primer reveals which allele was present.[7]

    Used on: Chemistry

  • Detected and Not Detected #

    The qualitative results that Typer software shows for an assay, as the manufacturer describes the platform.[8]

    Used on: Calls and confidence

  • DQ, IQ, OQ, PQ #

    Also written: design qualification; installation qualification; operational qualification; performance qualification

    Design, installation, operational and performance qualification: the stages in which USP general chapter <1058> frames analytical instrument qualification, with instruments sorted into groups A, B and C. A draft revision renames the chapter Analytical Instrument and System Qualification. How you run each stage is for your protocol and your manufacturer's documents.[9]

    Used on: Instrument qualification, Installation qualification

  • EpiTYPER #

    A methylation-analysis chemistry described in published protocols: bisulfite conversion, PCR with a T7 promoter tag, transcription, base-specific cleavage and MALDI-TOF detection. Whether it is currently offered is not stated on this site.[6]

    Used on: Chemistry

  • Extension primer #

    A short DNA primer that anneals directly next to the variant. The polymerase adds one terminator to it, and the mass of the extended primer shows which allele was present.[7]

    Used on: Chemistry

  • FFPE #

    Formalin-fixed, paraffin-embedded tissue: the way many tumor and archive samples are stored. DNA from it is often damaged or scarce.

    Textbook definition

    Used on: Reference materials

  • GeT-RM #

    The CDC's Genetic Testing Reference Material Coordination Program. It lists consensus genotypes for reference samples whose DNA comes from cell lines, for use as known-genotype controls.[10]

    Used on: Reference materials

  • Hardy-Weinberg #

    A statistical test of whether genotype frequencies in a group match the proportions expected from the allele frequencies. Large departures can point to a genotyping problem. SNPflow includes the test among its checks.[2]

    Used on: Qc metrics

  • iPLEX (Gold, Pro, HS) #

    The manufacturer's single-base-extension chemistry: locus-specific PCR, shrimp alkaline phosphatase treatment, extension with mass-modified terminators, then MALDI-TOF detection. Gold, Pro and HS are named versions. The chemistry page shows how they differ.[7]

    Used on: Chemistry

  • ISO 15189 #

    The international standard for the quality and competence of medical laboratories. The transition period for the 2022 edition ended in December 2025.[11]

    Used on: Regulatory context

  • LDT (laboratory-developed test) #

    A test designed, made and used within a single laboratory. FDA's 2024 final rule on LDTs was vacated in 2025; the regulatory context page gives the dated detail.[12]

    Used on: Regulatory context

  • LIMS #

    Also written: laboratory information management system

    Laboratory information management system: software that tracks samples, runs and results across a laboratory. This site makes no statement about how the platform's software connects to one.

    Textbook definition

    Used on: Software and records

  • MALDI #

    Also written: matrix-assisted laser desorption/ionization

    Matrix-assisted laser desorption/ionization: a laser pulse strikes a sample dried with a matrix on a chip, and the molecules leave as ions that can be measured by time of flight.

    Textbook definition

    Used on: Stages, Chemistry

  • Mass-modified terminator #

    A terminator with extra mass added so that the products for different alleles sit farther apart in the spectrum. The manufacturer states that the system distinguishes analytes 16 Da apart. In the public pages we checked on 24 November 2026, its exact mass increments are not listed.[13]

    Used on: Chemistry

  • MassARRAY Dx #

    The manufacturer's diagnostic-labeled product name. The manufacturer's own pages disagree on which product is the exception to research use only, and this site states no current status for it.[14][15]

    Used on: Research-use notice, Regulatory context

  • Matrix #

    In MALDI, the compound mixed with the sample and dried on the chip. It absorbs the laser energy and carries the analyte into the gas phase as ions. It also adds its own small peaks at low mass.

    Textbook definition

    Used on: Chemistry

  • Nanodispenser #

    A dispensing unit that places nanoliter-scale spots of sample onto matrix-loaded chip pads. The manufacturer announced a diagnostic-labeled version, the Dx Nanodispenser RS 1000, for Europe on 4 April 2016. How dispensing is done differs by unit; follow your manufacturer's documents.[15]

    Textbook definition

    Used on: Stages

  • Negative (no-template) control #

    Also written: NTC

    A well that goes through the whole process with no DNA, to show contamination or background signal. Published validations describe reserving such wells on every plate; how many is for your protocol to say.[16]

    Used on: Plate and chip layout, Qc metrics

  • No-call #

    A sample and assay for which the software or the reviewer returns no genotype.

    Textbook definition

    Used on: Calls and confidence

  • Peak area #

    The integrated signal under a peak. In EpiTYPER, percent methylation is worked out from peak areas rather than peak heights.[6]

    Used on: Chemistry

  • Performance verification #

    A run of known samples and controls that shows an instrument and assay still perform against the acceptance criteria you set. It records results; it does not set the criteria.

    Textbook definition

    Used on: Performance verification

  • Proficiency testing #

    External assessment in which a laboratory tests samples of undisclosed content and its results are compared with those of other laboratories or with the expected answer.

    Textbook definition

    Used on: Proficiency testing

  • QMSR #

    Also written: Quality Management System Regulation

    The revised 21 CFR Part 820, the Quality Management System Regulation for medical device makers. It took effect on 2 February 2026 and incorporates ISO 13485:2016 by reference.[18]

    Used on: Regulatory context

  • Research use only (RUO) #

    A label stating that a product is not for use in diagnostic procedures. The manufacturer states that apart from one diagnostic-labeled product all of its products carry it. A research-use label does not by itself settle how a laboratory may use the product.[19]

    Used on: Research-use notice, Regulatory context

  • Resin clean-up #

    Also written: desalting

    A step that removes salts from the extension reaction before spotting, because salt adducts and incomplete desalting can distort spectra and cause wrong calls. The manufacturer describes salts being removed before spotting, with desalting automated by the Chip Prep Module.[20]

    Used on: Stages, Chemistry

  • RMA #

    Also written: return number

    A reference number issued before an item is sent back for service, so that the receiver can match the package to a case. The manufacturer's warranty page says no product or component may be returned to it without its prior consent.[21]

    Used on: Turnaround and returns

  • SAP (shrimp alkaline phosphatase) #

    An enzyme added after PCR to neutralize leftover nucleotides so they do not take part in the extension step.[7]

    Used on: Chemistry

  • Sequenom #

    The company that built the MassARRAY platform before selling the business to Agena Bioscience, in a deal announced on 30 May 2014. It is why older units, papers and resale listings say Sequenom MassARRAY.[22][23]

    Used on: Used instruments, Trademarks and nominative use

  • Single-base extension #

    A reaction in which a polymerase adds exactly one terminator to a primer that sits next to the variant. The added base, read as a mass, gives the genotype.[7]

    Used on: Chemistry

  • SpectroCHIP array #

    Also written: chip

    The chip with matrix-loaded pads onto which samples are spotted before detection. Reused or refurbished chips are excluded from the manufacturer's warranty.[24]

    Used on: Plate and chip layout, Stages

  • Time of flight #

    Ions accelerated through the same field travel a field-free tube; lighter ions arrive sooner and heavier ones later, so arrival time gives mass-to-charge.

    Textbook definition

    Used on: Stages, Chemistry

  • Typer software #

    Also written: SpectroTYPER; TyperAnalyzer; TYPER 4.0

    The analysis and reporting software for the platform. Papers print it under several names, including Typer software, SpectroTYPER and TyperAnalyzer, and this site treats them as aliases. TYPER 4.0 was announced on 18 September 2007 alongside iPLEX Gold.[25]

    Used on: Software and records, Calls and confidence

  • UltraSEEK #

    A panel family for detecting mutations, listed on the manufacturer's panels page for lung, EGFR, colon and melanoma. The chemistry page explains how the approach works.[26]

    Used on: Chemistry

  • Unextended primer (UEP) #

    The extension primer that did not receive a terminator. Its peak always sits at a lower mass than any extended product, and a large one can mean an incomplete reaction.[7]

    Used on: Chemistry, Calls and confidence

  • Variant allele frequency (VAF) #

    The share of DNA molecules in a sample that carry a variant, often given as a percentage. It matters most in tumor samples, where the variant may be in only a small share of the DNA.

    Textbook definition

    Used on: Chemistry

  • WGA #

    Also written: whole-genome amplification

    Whole-genome amplification: copying all of a small DNA sample to get more material. Published work compares amplified with unamplified DNA; check any amplified DNA against your own acceptance criteria.

    Textbook definition

    Used on: Reference materials

  • z-score and signal-to-noise #

    Also written: S/N

    Signal-to-noise compares a peak's height with the background noise. A z-score says how many standard deviations a signal sits above the background distribution. Software and studies set their own cut-offs, so take yours from your validation.

    Textbook definition

    Used on: Calls and confidence

Sources

  1. [1]
    WFL-08

    A 2019 open-access study of Staphylococcus aureus isolates used Assay Design Suite v2.0, a 10-nucleotide tag on primary PCR primers to move unextended primers outside the 4500-9000 Da window, and extra bases on some extension primers to keep allele peaks apart. Per-SNP call rates ran 81% to 98.3% (about 89% average over nine usable SNPs), and the listed steps add up to 11.25 h.

    Open-access S. aureus MassARRAY genotyping study (PubMed Central PMC7471782) (external)

    Checked on 24 November 2026Peer-reviewedmedium confidence

  2. [2]
    WFL-11

    The SNPflow QC tool for iPLEX output computes call rate, expected versus observed replicates, minor allele frequency, discordant replicates, discordance rate and a Hardy-Weinberg test, checks genotype frequencies against reference data and has a sex-determination SNP module. It sets no numeric cut-offs.

    SNPflow: a lightweight application for the processing, storing and automatic quality checking of genotyping assays (PLoS ONE) (external)

    Checked on 24 November 2026Peer-reviewedmedium confidence

  3. [3]
    DAT-01

    A peer-reviewed marker study using Typer reported genotype confidence tiers (conservative, moderate, aggressive, low probability, user call): 54% conservative, 15% moderate, 9% aggressive, 18% low probability, 3% no alleles and 1% user calls. It defined a failed assay as one where most genotypes were low-probability or samples did not cluster by genotype. One species and marker set, not a performance specification.

    PLoS ONE genotyping study using Typer software (journal.pone.0049533) (external)

    Checked on 24 November 2026Peer-reviewedmedium confidence

  4. [4]
    WFL-05

    The Chip Prep Module (announced 24 August 2015) automates dispensing and desalting of samples onto SpectroCHIP Arrays and loading them onto the analyzer. It processes eight 96-pad arrays in under a day, has on-board cooling and storage for two more, runs largely unattended and is tied to the 96-well system.

    Agena Bioscience introduces front-end automation for its MassARRAY System for genetic analysis (press release, 24 August 2015); Technology Networks report, agenabio.com/company/press-release/agena-bioscience-introduces-front-end-automation-for-its-massarray-system-for-genetic-analysis/

    Checked on 24 November 2026Manufacturer-reported

  5. [5]
    LAB-04

    CLIA 42 CFR 493.1253: for an unmodified FDA-cleared test a laboratory verifies accuracy, precision and reportable range and confirms reference intervals; for modified tests and tests not subject to FDA clearance it also establishes analytical sensitivity and specificity (including interfering substances) and any other needed characteristic before reporting patient results.

    42 CFR 493.1253 Standard: Establishment and verification of performance specifications (eCFR) (external)

    Checked on 24 November 2026Regulator

  6. [6]
    CHM-09

    Published EpiTYPER protocol: bisulfite conversion, PCR with a T7 promoter tag, shrimp alkaline phosphatase, T7 transcription, RNase A cleavage after U positions (T-reaction), a clean-up step, then MALDI-TOF. A methylated CpG unit gives a fragment about 16 Da heavier; percent methylation is the methylated peak area over methylated plus unmethylated area.

    Frontiers in Genetics (2015) EpiTYPER methylation protocol article, DOI 10.3389/fgene.2015.00287 (external)

    Checked on 24 November 2026Peer-reviewedmedium confidence

  7. [7]
    CHM-01

    iPLEX is locus-specific PCR, then shrimp alkaline phosphatase treatment to neutralize leftover nucleotides, then single-base extension of a primer annealing directly next to the SNP using mass-modified dideoxynucleotide terminators, then MALDI-TOF. The mass of the extended primer shows which allele was present.

    SNP genotyping using the Sequenom MassARRAY iPLEX platform (Current Protocols in Human Genetics, 2009); PubMed Central article PMC5604939 (external)

    Checked on 24 November 2026Peer-reviewed

  8. [8]
    WFL-01

    Agena presents the MassARRAY System as an open platform in 96- and 384-well formats: end-point PCR with single-base extension, automated sample transfer and MALDI-TOF detection, then analysis and reporting in Typer software, which shows qualitative results as Detected / Not Detected.

    Agena Bioscience: MassARRAY System product page, agenabio.com/products/massarray-system/

    Checked on 24 November 2026Manufacturer-reportedmedium confidence

  9. [9]
    LAB-16

    USP general chapter <1058> frames analytical instrument qualification as DQ, IQ, OQ and PQ, with instruments sorted into groups A, B and C; a draft revision renames it Analytical Instrument and System Qualification.

    An Enhanced Approach to Analytical Instrument and System Qualification (Spectroscopy, 1 May 2025), spectroscopyonline.com/an-enhanced-approach-to-analytical-instrument-and-system-qualification

    Checked on 24 November 2026Third partymedium confidence

  10. [10]
    LAB-14

    CDC's GeT-RM page (10 September 2024) lists consensus genotypes for 363 samples across 9 studies covering 34 genes or loci, plus a list of 546 hard-to-detect pathogenic variants in 84 genes; DNA comes from Coriell cell lines. Characterization studies covered 107 DNAs for five loci (2010), 137 DNAs across 28 pharmacogenetic genes (2016) and 179 DNAs focused on CYP2D6 (2019).

    CDC Genetic Testing Reference Material Coordination Program (GeT-RM): reference materials (external)

    Checked on 24 November 2026Regulator

  11. [11]
    LAB-12

    The ISO 15189:2022 transition period ended in December 2025. UKAS flags clinical involvement in reviewing verification reports, and periodic review of examination methods (clauses 7.3.1 and 7.3.2), as common nonconformities.

    ISO 15189 transition update (UKAS, 6 June 2024) (external)

    Checked on 24 November 2026Third partymedium confidence

  12. [12]
    LAB-07

    FDA's 6 May 2024 final rule on laboratory-developed tests (89 FR 37286) was vacated by the U.S. District Court for the Eastern District of Texas on 31 March 2025 (ACLA v. FDA, No. 4:24-CV-479-SDJ); an FDA rule published 19 September 2025 restored the earlier text of 21 CFR 809.3(a).

    Laboratory Developed Tests (FDA); Federal Register rules of 6 May 2024 and 19 September 2025 (external)

    Checked on 24 November 2026Regulator

  13. [13]
    CHM-02

    Agena states the iPLEX Pro read window as about 4,500 to 9,000 Da and says the system distinguishes analytes 16 Da apart; a 2006 review gives gaps of at least 16 Da between the four terminator products and the same window.

    Agena Bioscience: Understanding Different MassARRAY Chemistries; 2006 review of MALDI-TOF DNA analysis (PubMed Central PMC1523240), agenabio.com/resources/articles/understanding-different-massarray-chemistries/

    Checked on 24 November 2026Manufacturer-reported

  14. [14]
    REG-02

    Agena's own pages disagree on the exception to research use only: the footer names MassARRAY Dx, while the website-terms body names the IMPACT Dx System. The point is unsettled.

    Agena Bioscience: Terms and Conditions; Agena Bioscience: Product Terms and Conditions, agenabio.com/terms-and-conditions/

    Checked on 24 November 2026Manufacturer-reported

  15. [15]
    REG-03

    On 4 April 2016 Agena announced a MassARRAY Dx Analyzer 4, a Dx Nanodispenser RS 1000 and Colon and Lung panels for diagnostic use in Europe only, not sold in the United States; a 29 June 2017 release added a CE-IVD system with the Chip Prep Module. Historical; current status is unverified.

    Agena Bioscience launches in vitro diagnostic tests for lung and colon cancer in Europe (4 April 2016); IVD genetic testing system launched in Europe (29 June 2017), agenabio.com/company/press-release/agena-bioscience-launches-vitro-diagnostic-tests-lung-colon-cancer-europe/

    Checked on 24 November 2026Manufacturer-reportedmedium confidence

  16. [16]
    DAT-10

    Published validations describe typical QC practice: reserved negative-control wells, reference samples of known genotype, duplicates to check discordance, an identity or sex marker, and second-person review. Acceptance numbers come from each laboratory's own validation; no numeric limit is published as a standard.

    SNPflow quality-control article (PLoS ONE, DOI 10.1371/journal.pone.0059508); Journal of Applied Laboratory Medicine 2020;5(3):454-466 (external)

    Checked on 24 November 2026Peer-reviewedmedium confidence

  17. [17]
    WFL-06

    Agena application notes from 2015 and 2017 say iPLEX genotyping runs in 24-, 96- or 384-well formats with up to 40 SNPs per well in about one day, and that the platform can interrogate 10 to 400 SNPs per sample, handle a few to hundreds of samples per day, take blood, plasma, buccal swab and FFPE material and work from as little as 5 ng DNA. Plex limits depend on the chemistry.

    Single nucleotide polymorphism detection with the iPLEX assay and the MassARRAY System (Agena application note, SelectScience 2015 and The Pathologist 2017), selectscience.net/resource/single-nucleotide-polymorphism-detection-with-the-iplex-r-assay-and-the-massarray-r-system

    Checked on 24 November 2026Manufacturer-reportedmedium confidence

  18. [18]
    LAB-02

    The revised 21 CFR Part 820, the Quality Management System Regulation (QMSR), took effect on 2 February 2026 and incorporates ISO 13485:2016 by reference.

    Quality Management System Regulation final rule (Federal Register, 2 February 2024, 2024-01709) (external)

    Checked on 24 November 2026Regulator

  19. [19]
    REG-01

    Agena states that, apart from MassARRAY Dx, all its products are for research use only and not for use in diagnostic procedures.

    Agena Bioscience: MassARRAY System product page, agenabio.com/products/massarray-system/

    Checked on 24 November 2026Manufacturer-reported

  20. [20]
    DIA-02

    Salt adducts and incomplete desalting can distort spectra and cause wrong calls (published patent, see CHM-05). Agena describes salts being removed before spotting, with desalting automated by the Chip Prep Module.

    US 7,888,127 B2 (published patent, Google Patents); Agena Bioscience: Understanding Different MassARRAY Chemistries (external)

    Checked on 24 November 2026Third partymedium confidence

  21. [21]
    WAR-05

    Before a service visit the customer must tell Agena in writing about any decontamination of the system, a customer representative must be on site throughout on-site service, and no product or component may be returned without Agena's prior consent.

    Agena Bioscience: System Warranty, agenabio.com/support/system-warranty/

    Checked on 24 November 2026Manufacturer-reportedmedium confidence

  22. [22]
    MAN-01

    Sequenom sold its MassARRAY business to Agena Bioscience, then backed by the investment firm Telegraph Hill, in a deal announced 30 May 2014; Agena stayed in San Diego. This is why older units and papers say Sequenom.

    Sequenom, Inc.'s Bioscience Business Acquired by Agena Bioscience (press release, 30 May 2014); BioSpace report of the sale, prnewswire.com/news-releases/sequenom-incs-bioscience-business-acquired-by-agena-bioscience-a-telegraph-hill-partners-portfolio-company-261295181.html

    Checked on 24 November 2026Manufacturer-reported

  23. [23]
    MAN-08

    Older sources and resale listings call the platform Sequenom MassARRAY or Sequenom MassARRAY Analyzer 4; Agena pages say MassARRAY System or MassARRAY Analyzer 4; FDA's 2024 letter abbreviates it MassARRAY4 System (MA4).

    Agena Bioscience, Inc. - 665159 - 03/21/2024 (FDA warning letter); LabWrench equipment listing for the Sequenom MassARRAY Analyzer 4 (external)

    Checked on 24 November 2026Third partymedium confidence

  24. [24]
    WAR-04

    The warranty excludes reused or refurbished chips (SpectroCHIP Arrays are named), third-party hardware or software Agena has not accepted in writing, exposure to BSL-3/4 agents or radioactivity, malicious software, and acts of God.

    Agena Bioscience: System Warranty, agenabio.com/support/system-warranty/

    Checked on 24 November 2026Manufacturer-reportedmedium confidence

  25. [25]
    DAT-07

    Typer naming: Sequenom announced TYPER 4.0 on 18 September 2007 alongside iPLEX Gold. Papers print Typer software, SpectroTYPER v4.0, Typer v4.0.26.74, MassARRAY Typer Analyzer v4.0.26.73 and TyperAnalyzer, which this site treats as aliases.

    Sequenom launches TYPER 4.0 genotyping software (BioSpace); PubMed Central article PMC9022865, biospace.com/sequenom-inc-launches-typer-4-0-genotyping-software

    Checked on 24 November 2026Third partymedium confidence

  26. [26]
    VAL-07

    Agena's current panels page lists 15 unique panels (VeriDose Core 16 genes, DPYD Plus 18 variants, HFE 6 variants, iPLEX Pro Sample ID 44 SNPs, iPLEX HS Lung, Colon and Melanoma, PIK3CA Breast, UltraSEEK Lung, EGFR, Colon and Melanoma, CFTR 100+, CYP2D6 CNV, Exome QC); none is labeled as a diagnostic product there, and CFTR 100+ was announced on 6 April 2026 as research use only.

    Agena Bioscience: Panels; Agena Bioscience announces launch of the CFTR 100+ panel on the MassARRAY System (6 April 2026), agenabio.com/applications/panels/

    Checked on 24 November 2026Manufacturer-reportedmedium confidence