Validation and SOPs

Assay validation and reference materials

A roadmap of questions to answer when validating an assay, and what to know about reference materials; worksheets leave every value for you to set.

Checked on 24 November 2026
On this page

Assay validation roadmap

A roadmap of questions to answer when validating an assay, with worksheets that leave every value for you to set.

A roadmap, not a protocol

This page lists the questions an assay validation has to answer. Your laboratory writes the answers, and sets every limit.

No pass or fail numbers are given here because limits are set by the laboratory director. The worksheets below have blank columns where those numbers go, and you fill them before you collect any data.

Whether an assay is for research use or for a laboratory-defined use is also the director’s decision, made under your own regulatory framework. See Regulatory context for a dated summary of that landscape.

Twelve sections of a validation plan

Write the plan before the study. Each section has a prompt.

  1. Step 1: Purpose and scope as set by the laboratory

    State whether the assay is for research use or laboratory-defined use, as set by the laboratory director, and what it covers.
  2. Step 2: Regulatory and quality-system basis

    List the rules, standards and contracts your laboratory works under and which parts apply to this assay.
  3. Step 3: Roles

    Name who plans, performs, reviews and signs off each part of the study.
  4. Step 4: Instrument qualification

    Refer to the qualification records for each instrument, or state what is still open.
  5. Step 5: Assay design and content

    Describe the content (targets and versions) and how it was chosen. Record any change from the manufacturer's protocol.
  6. Step 6: Reference materials and sample panel

    List reference materials and samples, their sources, how their expected results are known, and why they suit this assay.
  7. Step 7: Performance studies

    For each performance characteristic, state the design, the number of samples and runs you chose and why, and the criterion you set before starting.
  8. Step 8: QC strategy

    State the controls on each plate, how they are trended, and what happens when one is not met.
  9. Step 9: Data analysis, review and disposition of results under the laboratory's own regulatory framework

    Describe analysis settings, review steps, manual review rules and disposition, in the terms your framework uses.
  10. Step 10: Data integrity and security

    Describe access, backup, record retention and change tracking for study data.
  11. Step 11: Deviations and revalidation triggers

    State how deviations are recorded and which changes require a new study or part of one.
  12. Step 12: Summary report and sign-off

    State what the summary report contains and who signs it.

Published guidance on validating and verifying qualitative multiplex assays for genotyping exists as good-practice recommendations rather than requirements.[1]

Performance characteristics to consider

Names and questions only. Which ones apply, how many samples and what counts as good enough are decisions for your plan.

One framework describes the range of characteristics: accuracy, precision and reportable range for an unmodified test, with analytical sensitivity and specificity, including interfering substances, added for modified tests and tests not subject to clearance.[2]

Qualitative germline genotyping
CharacteristicThe question to answer
Accuracy against a comparator methodDo the calls agree with a method you trust? Which comparator, and how do you know it is right?
Precision: repeatabilityDoes the same sample give the same call when the same person repeats it within one session?
Precision: intermediate precision or reproducibilityDoes it still agree across days, operators, plates, lots and instruments?
Analytical specificity, including interfering substancesDoes anything in the sample or the preparation produce a wrong call?
Input quantity rangeOver what range of input does the assay work, and what happens outside it?
Specimen typesWhich specimen types and preservation methods have you actually tested?
Call rate and handling of no-callsHow are failed or ambiguous calls recorded, repeated and counted?
Contamination and carryoverDo no-template controls stay clean across a run and across runs?
Robustness to small changes in the procedureWhich small variations did you test on purpose, and what happened?
Quantitative somatic detection
CharacteristicThe question to answer
Analytical sensitivity (lowest level detected)What is the lowest level of a variant the assay detects reliably, and how was that shown?
Accuracy against a comparator methodDo the calls agree with a method you trust? Which comparator, and how do you know it is right?
Precision: repeatabilityDoes the same sample give the same call when the same person repeats it within one session?
Precision: intermediate precision or reproducibilityDoes it still agree across days, operators, plates, lots and instruments?
Analytical specificity and false-positive handlingHow are false positives found, reviewed and prevented?
Quantitation range (if the result is quantitative)If the result is a fraction or a level, over what range is it reliable?
Input quantity rangeOver what range of input does the assay work, and what happens outside it?
Specimen types and preservationWhich specimen types and preservation methods have you actually tested?
Contamination and carryoverDo no-template controls stay clean across a run and across runs?

Acceptance-criteria worksheet

Pick the characteristics your plan covers, write the study design and the criterion before the study, then record the result.

Performance characteristics and acceptance criteria
Assay typeCharacteristicStudy designAcceptance criterion (laboratory-set)Basis for the criterionResultReviewerRow actions

Your entries are kept in this browser only. Nothing is sent to us.

Sample planner

Turn the counts you chose into wells and plates. It suggests no minimum.

A whole number you chose

A whole number you chose

A whole number you chose

As your SOP reserves them

Arithmetic onlyno suggested minimums

Enter the four counts to see the arithmetic. Whole numbers only; zero is allowed.

This tool does not say how many samples, replicates or runs a study needs. Your plan and your laboratory director decide that.

After validation: ongoing QC and revalidation

A validation describes the assay on the day it was done. These items keep describing it.

Ongoing QC and revalidation
ItemWhat is recordedFrequencyRevalidation trigger
Plate controlsNegative controls, a reference sample, duplicates and an identity or sex marker, as your SOP defines them.[3]Set by your laboratoryA control failing, or changing how controls are used.
Reference sample trendCalls and any signal metric your SOP tracks, run after run.Set by your laboratoryA sustained shift you decided in advance counts as one.
Instrument performance verificationSee performance verification.Set by your laboratoryA result outside the limits you set, or a repair.
External assessmentSee proficiency testing.Set by your laboratoryAn unsatisfactory result or a corrective action.
Lot and software changesSee change control.At each changeYour impact assessment says so.

One example framework sets daily minimum controls for qualitative and molecular tests, with more where extraction or amplification is involved.[4] A UK assessment body lists the review of verification reports and periodic review of examination methods among its common findings.[6]

A reality check on effort

Reference materials

What reference materials are for, how to choose and track them, and where to look for sources.

Samples whose answer you already know

A reference material is DNA with an expected result you can trace to something other than your own assay.

Laboratories use them to check accuracy during validation, to trend an assay from run to run, and to compare one reagent lot with the next. The value of a reference material depends on how well its expected result is known, and on whether it resembles your real samples.

A reference material is not a proficiency sample. You know the answer to the first; the second is sent blind by an external provider to test how your laboratory performs. See proficiency testing.

Public sources to look at

Starting points, not recommendations. Check each against your own panel before relying on it.

  • CDC GeT-RM

    Genotype reference DNA

    A US public-health program that lists consensus genotypes for a large set of cell-line DNAs across many genes, plus a list of hard-to-detect pathogenic variants. The DNA comes from a public cell-line repository.[5]

  • A commercial somatic standard

    Multiplex FFPE standard, example only

    One supplier sells a multiplex FFPE reference standard with variants at a range of allele fractions, confirmed per batch by a second method. It covers only some variants, so check the list against your panel. It is labeled for in vitro diagnostic use and is available only where licensed.[7]

  • iPLEX reference standards

    Manufacturer-announced, 2016

    In April 2016 the manufacturer and a reference-standard supplier announced reference standards for verifying oncology assay performance. Current availability is unknown, so ask before you plan around them.[8]

Comparator methods

Where no reference material exists for a variant, laboratories compare against a second method: a different genotyping platform, sequencing, or digital PCR for low allele fractions. The comparator needs its own evidence of accuracy, and any disagreement needs an investigation, not a vote.

Keep a register

One row per reference material. The register answers 'where did this expected result come from?' years later.

Reference-material register
MaterialSourceLotExpiryWhere the expected results come fromStorageUsed forRow actions

Your entries are kept in this browser only. Nothing is sent to us.

Published examples

Papers that show how others planned or reported a validation. This list gives titles and identifiers only, with no results.

Showing 4 of 4 references.

Published examples: reference, journal and year, identifier, and what each illustrates
ReferenceJournal and yearIdentifierWhat it illustratesTopic
Tang NY and colleaguesValidation of a Large Custom-Designed Pharmacogenomics Panel on an Array Genotyping PlatformJournal of Applied Laboratory Medicine, 2021; 6:1505-1516DOI 10.1093/jalm/jfab056 (external)How one laboratory planned accuracy and precision studies against sequencing and genotyping comparators. It used a different platform, so read it as a design template, not as evidence about any MassARRAY configuration.Validation design
Jensen L, Børsting C, Dalhoff K, Morling NEvaluation of the iPLEX ADME PGx Pro Panel and allele frequencies of pharmacogenetic markers in DanesClinical Biochemistry, 2016; 49:1299-1301DOI 10.1016/j.clinbiochem.2016.07.014 (external)A commercial panel assessed by a group independent of the manufacturer, which is one way to see how assays behave outside their developer's hands.Independent evaluation
Domoradzki T, Grochowski P, Jaśkiewicz A, Pająk BUltrasensitive analysis of genetic instability related to chemical exposureJournal of Applied Genetics, 2022; 63:305-313DOI 10.1007/s13353-021-00677-6 (external)A research application that ran several panels on the MassARRAY Analyzer 4, which shows the platform in research use years after its launch.Application study
Williams GR and colleaguesJournal of Applied Laboratory Medicine, 2020; 5:454-466DOI 10.1093/jalm/jfaa018 (external)How much work one laboratory reported in validating a multi-gene iPLEX panel: the authors describe significant development time, a modified manufacturer protocol and manual review of every assay plate.Validation effort

Reference materials feed the plan. See the assay validation roadmap and template T-02 Validation plan.

General information; not legal, regulatory or clinical advice.

Sources in reduced form

These sources are cited by identifier, journal and year only. This page gives none of their figures.

Sources

  1. [1]
    LAB-13

    CLSI MM17 (2nd edition, 31 May 2018) covers validation and verification of qualitative multiplex nucleic acid assays for genotyping and pathogen detection, as good-practice recommendations rather than requirements; CLSI QMS02 (7th edition, 11 March 2024) is a guideline on developing and managing laboratory documents.

    CLSI MM17: Validation and Verification of Multiplex Nucleic Acid Assays; CLSI QMS02 (external)

    Checked on 24 November 2026Third party

  2. [2]
    LAB-04

    CLIA 42 CFR 493.1253: for an unmodified FDA-cleared test a laboratory verifies accuracy, precision and reportable range and confirms reference intervals; for modified tests and tests not subject to FDA clearance it also establishes analytical sensitivity and specificity (including interfering substances) and any other needed characteristic before reporting patient results.

    42 CFR 493.1253 Standard: Establishment and verification of performance specifications (eCFR) (external)

    Checked on 24 November 2026Regulator

  3. [3]
    DAT-10

    Published validations describe typical QC practice: reserved negative-control wells, reference samples of known genotype, duplicates to check discordance, an identity or sex marker, and second-person review. Acceptance numbers come from each laboratory's own validation; no numeric limit is published as a standard.

    SNPflow quality-control article (PLoS ONE, DOI 10.1371/journal.pone.0059508); Journal of Applied Laboratory Medicine 2020;5(3):454-466 (external)

    Checked on 24 November 2026Peer-reviewedmedium confidence

  4. [4]
    LAB-05

    CLIA 42 CFR 493.1256: daily minimum controls are one positive and one negative for qualitative tests, two controls for tests with an extraction step (one able to detect extraction errors), and for molecular amplification two controls plus an inhibition control where inhibition is a significant cause of false negatives.

    42 CFR 493.1256 Standard: Control procedures (eCFR) (external)

    Checked on 24 November 2026Regulator

  5. [5]
    LAB-14

    CDC's GeT-RM page (10 September 2024) lists consensus genotypes for 363 samples across 9 studies covering 34 genes or loci, plus a list of 546 hard-to-detect pathogenic variants in 84 genes; DNA comes from Coriell cell lines. Characterization studies covered 107 DNAs for five loci (2010), 137 DNAs across 28 pharmacogenetic genes (2016) and 179 DNAs focused on CYP2D6 (2019).

    CDC Genetic Testing Reference Material Coordination Program (GeT-RM): reference materials (external)

    Checked on 24 November 2026Regulator

  6. [6]
    LAB-12

    The ISO 15189:2022 transition period ended in December 2025. UKAS flags clinical involvement in reviewing verification reports, and periodic review of examination methods (clauses 7.3.1 and 7.3.2), as common nonconformities.

    ISO 15189 transition update (UKAS, 6 June 2024) (external)

    Checked on 24 November 2026Third partymedium confidence

  7. [7]
    LAB-15

    Horizon (Revvity) HD200-IVD multiplex FFPE reference standard: 11 variants in 6 genes at about 0.9% to 24.5% allele fraction, fractions confirmed per batch by ddPCR, labeled IVD and only available where licensed. It covers only some variants, so check it against the panel.

    Quantitative Multiplex FFPE Reference Standard (IVD) product page (Horizon), horizondiscovery.com/en/reference-standards/products/quantitative-multiplex-ffpe-reference-standard-ivd

    Checked on 24 November 2026Third partymedium confidence

  8. [8]
    SUP-07

    In April 2016 Agena and Horizon Discovery announced iPLEX reference standards for verifying oncology assay performance on the MassARRAY System, saying robust performance needs upfront analytical verification and validation. Current availability is unknown.

    Agena Bioscience and Horizon Discovery to launch iPLEX reference standards for the MassARRAY System (SelectScience, 18 April 2016), selectscience.net/article/agena-bioscience-and-horizon-discovery-to-launch-iplex-r-reference-standards-for-the-massarray-r-system

    Checked on 24 November 2026Manufacturer-reportedmedium confidence