Workflow and data
Workflow
The MassARRAY workflow from sample to data: an overview and each stage in order.
Overview
The MassARRAY workflow from sample to data, with run-time planning and layout tools you can use in the browser.
The manufacturer presents the MassARRAY system as an open platform in 96-well and 384-well formats: end-point PCR with single-base extension, automated sample transfer and MALDI-TOF detection, then analysis and reporting in Typer software.[1]
This page follows one plate from intake to archive. It is written for the person who plans runs and the engineer who needs to know what happened upstream of a symptom. Every published figure carries its source, and every figure that looks like instrument output is synthetic.
Go deeper
- StagesEach stage with inputs, outputs, published parameters and contamination notes.
- MALDI-TOF and assay chemistryThe science, with a computed mass example you can change.
- Contamination controlOne-way flow, controls and a floor-plan sketch tool.
- Plate and chip layoutA plate map builder with CSV export.
- Run-time plannerAdd up stage times for a batch of plates.
- Data outputWhat the software produces and how laboratories review it.
The run in one picture
Eleven stages, numbered 0 to 10, across the people and machines that carry them out. Stages 0 to 3 sit in the pre-PCR area and stages 4 to 10 in the post-PCR area. That split is general good practice, not a manufacturer requirement.[2]
Workflow swimlane: who and what does each stage, with the manufacturer's published stage times for 96- and 384-well formats.
Text version of this figure
- Stage 0, Intake and plate map (pre-PCR area): Log the samples, assign every well, reserve the control wells.
- Stage 1, DNA input and QC (pre-PCR area): Check and normalize the DNA that goes into the reaction.
- Stage 2, Assay design (pre-PCR area): Choose the variants and design the primers so every product fits the mass window.
- Stage 3, Multiplex PCR (pre-PCR area): Amplify each locus in a multiplex reaction.
- Stage 4, SAP treatment (post-PCR area): Neutralize leftover nucleotides so they cannot take part in extension.
- Stage 5, Extension (post-PCR area): Extend a primer by one base; the mass of the product shows the allele.
- Stage 6, Resin clean-up (post-PCR area): Remove salts before spotting, because salts distort spectra.
- Stage 7, Dispense onto chip (post-PCR area): Transfer the cleaned reactions onto the SpectroCHIP array.
- Stage 8, MALDI-TOF acquisition (post-PCR area): The analyzer reads each spot and records a mass spectrum.
- Stage 9, Typer analysis and review (post-PCR area): The software makes calls; a person reviews them and the controls.
- Stage 10, Reporting and archive (post-PCR area): Record the outcome, keep the evidence and archive the run.
Stage times
Manufacturer-reported (2026) The manufacturer prints the stage times below. It prints no total, so the run-time and hands-on totals are our sums of its published parts, with the arithmetic shown. Times vary with chemistry and acquisition settings.[3][4]
| Format | PCR, SAP and extension | Hands-on | Detection | Hands-on | Analysis, hands-on | Run time (our sum) | Hands-on total (our sum) |
|---|---|---|---|---|---|---|---|
| 96-well | about 5.5 h | about 23 min | about 100 min | about 5 min | 5 to 10 min | 5.5 h + 100 min (1.67 h) = 7.17 h, about 7.2 h | 23 + 5 + (5 to 10) = 33 to 38 min |
| 384-well | about 5.75 h | about 45 min | about 150 min | about 5 min | 5 to 10 min | 5.75 h + 150 min (2.5 h) = 8.25 h | 45 + 5 + (5 to 10) = 55 to 60 min |
Want to plan several plates or your own times? Use the run-time planner.
Stage explorer
Pick a stage for a short summary. The full cards, with inputs, outputs and controls, are on the stages page.
Stage 0: Intake and plate map
Log the samples, give every well an owner and reserve the control wells before anything is pipetted. Keep one plate map from PCR to chip.
One core facility's rule asks for at least two open wells on each 384-well plate for negative controls.[5]
Two paths from reaction plate to chip
Path 1: dispenser. The cleaned reaction plate goes to a stand-alone dispenser, which loads the matrix-prepared pads of the chip. The chip is then loaded into the analyzer. We do not describe the dispensing mechanism.
Path 2: chip-prep unit. Manufacturer-reported (2015) The Chip Prep Module, announced on 24 August 2015, automates dispensing and desalting onto SpectroCHIP arrays and loads them onto the analyzer. It processes eight 96-pad arrays in under a day, has cooled storage for two more, runs largely unattended and is tied to the 96-well system.[9] That is a 2015 figure and should not be read as current capacity.
The same release says the module is designed to reduce cross-contamination from manual sample handling.[10] It is a manufacturer claim from 2015, with no performance figure.
Replacing consumable hardware such as dispense pins and liquid-handler tips, and disposing of waste, appear on the manufacturer’s list of customer responsibilities.[11] Work inside the instrument is for a service engineer.
Input and scale
How much DNA you need depends on the panel; there is no single platform-wide figure.[6]
Manufacturer-reported (2017) Application notes from 2015 and 2017 say iPLEX genotyping runs in 24-, 96- or 384-well formats, with up to 40 SNPs per well in about one day, and that the platform can interrogate 10 to 400 SNPs per sample from blood, plasma, buccal swab and FFPE material, from as little as 5 ng of DNA. Plex limits depend on the chemistry.[14]
The annual report of the manufacturer’s owner gives up to 50 DNA variants per PCR reaction and up to 384 samples per SpectroCHIP array as stated capacity.[15]
A 2011 chapter describes medium-throughput iPLEX Gold as suited to up to 36 markers in 384 individuals, on pre-spotted arrays.[16]
The 24-well format was released in February 2015; the manufacturer’s current system page lists only 96- and 384-well formats.[17]
Controls and contamination at a glance
Published validations describe the same quality-control habits: reserved negative-control wells, reference samples of known genotype, duplicates to check for discordance, an identity or sex marker, and review by a second person. No numeric limit is published as a standard; each laboratory takes its limits from its own validation.[12]
Plan the controls when you draw the plate map, not after the run. A control that was never given a well cannot be read back later. The plate and chip layout page has a map builder.
Keep work before PCR and after PCR apart. APHL guidance recommends one-way workflow and separate areas, and CDC’s MMWR describes one-way workflow with separate areas for specimen preparation, amplification and detection for open amplification systems.[2] See contamination control.

Stages
Each stage of the workflow in order, what it produces and where things most often go wrong.
Eleven stages, in order. Each card says what the stage is for, what goes in and what comes out, the published parameters with their attribution, the controls that belong at that point, and the contamination rule for the area it sits in. Choose a chemistry to see how the stage differs for it.
Stages 0 to 3 sit in the pre-PCR area and stages 4 to 10 in the post-PCR area, which is general good practice and not a manufacturer requirement.[2] Hazards: covers on the instruments stay closed. Work inside an instrument is for a qualified service engineer.
- Stage 00pre-PCR area
Intake and plate map
- Purpose
- Give every sample a well and every control a well before work starts, so that every later stage uses the same map.
- Goes in
- Sample list, extraction records, the run request.
- Comes out
- One plate map, kept unchanged from PCR to chip, and a run log opened for the plate.
- Published
One core facility's rule asks for at least two open wells on each 384-well plate for negative controls.[5]
- Controls
Reserve wells for negative controls, a reference sample of known genotype, duplicates and an identity or sex marker.[12] Use the plate map builder.
- Contamination
Pre-PCR area. Reagents and samples are handled here and no amplified material comes back in.
- Your kit
Take volumes and cycling from your current kit instructions. This page prints none.
iPLEXGeneric step
Same.
- Stage 01pre-PCR area
DNA input and QC
- Purpose
- Confirm that the DNA is good enough and set the amount per reaction, so a weak result can be traced to input and not guessed at.
- Goes in
- Extracted DNA, including whole-genome-amplified DNA where your design allows it.
- Comes out
- DNA at the concentration your method calls for, in the plate layout of the plate map.
- Published
The manufacturer quotes roughly 10 to 40 ng of extracted nucleic acid. Published workflows use 5 to 10 ng (iPLEX HS, OncoFOCUS v3.0), 10 ng (UltraSEEK breast panel) and 40 ng (UltraSEEK Oncogene panel, 2014), so the amount depends on the panel.[6]
One core facility's rule asks for DNA eluted in water rather than TE, because the instrument is sensitive to salt, a 260/280 ratio of about 1.7 to 2.0, and 1 µL at 10 ng/µL per reaction.[5]
A PLoS ONE study found freshly prepared whole-genome-amplified DNA performed like genomic DNA, while WGA DNA stored frozen for 18 months gave worse efficiency and accuracy.[18]
- Controls
Include the reference sample of known genotype.[12]
- Contamination
Pre-PCR area. Reagents and samples are handled here and no amplified material comes back in. Open one DNA plate at a time.
- Your kit
Take volumes and cycling from your current kit instructions. This page prints none.
iPLEXGeneric step
Same.
- Stage 02pre-PCR area
Assay design
- Purpose
- Choose the variants and design the primers so that every expected product lands inside the mass window and apart from its neighbors.
- Goes in
- Variant list, reference sequence, panel size you want.
- Comes out
- A multiplex design: one PCR primer pair and one extension primer for each variant.
- Published
The manufacturer states a read window of about 4,500 to 9,000 Da for one chemistry and says the system distinguishes analytes 16 Da apart.[7]
One core facility's rule accepts up to 36-plex (25 to 30 typical), amplicons of about 100 bp, three unlabeled primers of 15 to 25 bases per variant, and no known polymorphism within 20 bp of the variant.[5]
A 2019 study used Assay Design Suite v2.0, a 10-nucleotide tag on the primary PCR primers to move unextended primers outside the window, and extra bases on some extension primers to keep allele peaks apart.[20]
- Controls
Check on paper that no two expected products fall within 16 Da of each other. The chemistry page has a mass calculator for this kind of check.
- Contamination
Pre-PCR area. Reagents and samples are handled here and no amplified material comes back in. Keep primer stocks and working dilutions in the reagent area.
- Your kit
Take volumes and cycling from your current kit instructions. This page prints none.
iPLEXGeneric step
Same.
- Stage 03pre-PCR area
Multiplex PCR
- Purpose
Locus-specific PCR amplifies each target region.[8]
- Goes in
- DNA, multiplex primer mix, PCR reagents, the plate map.
- Comes out
- A sealed reaction plate of amplified product.
- Published
Manufacturer-reported (2026) PCR, SAP and extension together take about 5.5 h on a 96-well plate (about 23 min hands-on) and about 5.75 h on a 384-well plate (about 45 min hands-on).[3]
- Controls
Carry the no-template wells through PCR with everything else.[12]
- Contamination
Set the reaction up in the pre-PCR area. Once the plate is sealed and in the cycler, that plate has amplified product and does not return to the pre-PCR area.
- Your kit
Take volumes and cycling from your current kit instructions. This page prints none.
iPLEXGeneric step
Same.
- Stage 04post-PCR area
SAP treatment
- Purpose
Shrimp alkaline phosphatase neutralizes unused nucleotides left from PCR, so that they cannot compete in the next step.[8]
- Goes in
- The sealed PCR plate, SAP reagent.
- Comes out
- A treated reaction plate.
- Controls
No-template and reference wells stay on the plate and receive the same treatment.
- Contamination
Post-PCR area. Amplified product is present from here on; keep plates and tools in this area. This is the first stage that opens a plate holding amplified product, so it is the stage where care with seals and tips matters most.
- Your kit
Take volumes and cycling from your current kit instructions. This page prints none.
iPLEXGeneric step
Same.
- Stage 05post-PCR area
Extension
- Purpose
A primer that anneals directly next to the variant is extended by a single base using mass-modified dideoxynucleotide terminators. The mass of the extended primer shows which allele was there.[8]
- Goes in
- The treated plate, extension primer mix, enzyme and terminators from your kit.
- Comes out
- A plate of extension products ready for clean-up.
- Published
Terminator structures and mass shifts are not shown on this site. Take them from your manufacturer documents.
- Controls
Controls travel through extension with the samples. A no-template well should show the unextended primer and little else.
- Contamination
Post-PCR area. Amplified product is present from here on; keep plates and tools in this area.
- Your kit
Take volumes and cycling from your current kit instructions. This page prints none.
iPLEXGeneric step
Single-base extension with mass-modified terminators.
- Stage 06post-PCR area
Resin clean-up
- Purpose
Remove salts before spotting. Salt adducts and incomplete desalting can distort spectra and cause wrong calls.[13]
- Goes in
- The extension plate, clean-up material from your kit.
- Comes out
- A desalted plate ready to spot.
- Published
The manufacturer describes salts being removed before spotting, with desalting automated by the chip-prep unit.[13] This site gives no material, amount or time for the clean-up.
- Controls
Look at the spectra of the first spots for the distortion that incomplete desalting causes before you commit the rest of the plate.
- Contamination
Post-PCR area. Amplified product is present from here on; keep plates and tools in this area.
- Your kit
Take volumes and cycling from your current kit instructions. This page prints none.
iPLEXGeneric step
Same.
- Stage 07post-PCR area
Dispense onto chip
- Purpose
- Move each cleaned reaction to its own pad on the chip, in the order of the plate map.
- Goes in
- The desalted plate, a SpectroCHIP array.
- Comes out
- A loaded chip, ready for the analyzer.
- Published
Manufacturer-reported (2015) The Chip Prep Module automates dispensing and desalting onto SpectroCHIP arrays and loading them onto the analyzer. It processes eight 96-pad arrays in under a day, has cooled storage for two more, and is tied to the 96-well system.[9]
Stated capacity is up to 384 samples per array.[15] A 2011 chapter describes iPLEX Gold on pre-spotted arrays.[16]
- Controls
Check the orientation of plate to chip before dispensing, and keep the plate map with the chip.
- Contamination
Post-PCR area. Amplified product is present from here on; keep plates and tools in this area. The manufacturer says its chip-prep unit is designed to reduce cross-contamination from manual handling; that is a 2015 claim with no figure.[10]
- Your kit
Take volumes and cycling from your current kit instructions. This page prints none.
iPLEXGeneric step
Same.
- Stage 08post-PCR area
MALDI-TOF acquisition
- Purpose
- A laser pulse ionizes the sample on each pad and the analyzer measures how long the ions take to reach the detector. The result is a mass spectrum for each spot.
- Goes in
- The loaded chip and your validated acquisition method.
- Comes out
- A raw spectrum for every spot.
- Published
Manufacturer-reported (2026) Detection takes about 100 min for a 96-well plate and about 150 min for a 384-well plate, with about 5 min hands-on.[3]
The manufacturer states a read window of about 4,500 to 9,000 Da for one chemistry.[7] Acquisition settings are set in your own validated method; none are printed here.
- Controls
Run performance verification as your own SOP says. See performance verification.
- Contamination
The chip holds desalted product. Handle it in the post-PCR area and keep the analyzer covers closed: work inside the instrument is for a service engineer.
- Your kit
Take volumes and cycling from your current kit instructions. This page prints none.
iPLEXGeneric step
Same.
- Stage 09post-PCR area
Typer analysis and review
- Purpose
Typer software turns spectra into calls and shows qualitative results as Detected or Not Detected.[1] A person then reviews controls, calls and plots.
- Goes in
- Raw spectra, the plate map, your assay definitions.
- Comes out
- Calls with confidence labels, plate and cluster views, and a reviewed result.
- Published
The software appears in papers as Typer, SpectroTYPER v4.0, Typer v4.0.26.74 and TyperAnalyzer. This site treats these as aliases of one product line.[22]
How cluster plots, spectra and plate views are read, and why a person reviews them, is set out on the calls and confidence page.
The SNPflow tool for iPLEX output computes call rate, expected against observed replicates, minor allele frequency, discordance and a Hardy-Weinberg test, and sets no numeric cut-offs.[23]
- Controls
Second-person review of results.[12] See manual review.
- Contamination
Electronic only. Keep result files separate from edited copies.
- Your kit
Take volumes and cycling from your current kit instructions. This page prints none.
iPLEXGeneric step
Same.
- Stage 10post-PCR area
Reporting and archive
- Purpose
- Record the outcome and keep the evidence that supports it, in a form someone else can read years later.
- Goes in
- The reviewed result, the plate map, the run log.
- Comes out
- A signed run record and an archived set of files.
- Published
Data backup is listed as a customer responsibility in the manufacturer’s warranty terms.[11]
- Controls
Use the run log template and the records checklist.
- Contamination
Not applicable to the digital record. Plates and chips follow your waste and storage rules.
- Your kit
Take volumes and cycling from your current kit instructions. This page prints none.
iPLEXGeneric step
Same.
Continues on Chemistry and contamination control: the assay chemistry and how to keep pre- and post-amplification work apart.
Sources
[1]
WFL-01Agena presents the MassARRAY System as an open platform in 96- and 384-well formats: end-point PCR with single-base extension, automated sample transfer and MALDI-TOF detection, then analysis and reporting in Typer software, which shows qualitative results as Detected / Not Detected.
Agena Bioscience: MassARRAY System product page, agenabio.com/products/massarray-system/
Checked on 24 November 2026Manufacturer-reportedmedium confidence
[2]
LAB-19APHL guidance on molecular laboratory space recommends unidirectional workflow separating pre-PCR stages, at least three enclosed rooms (reagents; sample preparation and assay setup; PCR and post-amplification) or physical separation using dead-air boxes or PCR workstations, and positive air pressure in reagent rooms. CDC's MMWR says CLIA requires unidirectional workflow with separate areas for specimen preparation, amplification and detection for open amplification systems; closed systems are exempt.
APHL: Molecular Workflow (newborn screening resources); CDC MMWR Recommendations and Reports 58(RR-6), aphl.org/programs/newborn_screening/Pages/Molecular-Workflow.aspx
Checked on 24 November 2026Third partymedium confidence
[3]
WFL-02Agena's stage times, which vary with chemistry and acquisition settings: 96-well PCR, SAP and extension about 5.5 h (about 23 min hands-on) and detection about 100 min (about 5 min hands-on); 384-well about 5.75 h (about 45 min hands-on) and detection about 150 min (about 5 min hands-on); data analysis 5-10 min hands-on.
Agena Bioscience: MassARRAY System product page, agenabio.com/products/massarray-system/
Checked on 24 November 2026Manufacturer-reportedmedium confidence
[4]
WFL-03Adding Agena's stage times gives about 7.2 h of run time per 96-well plate (5.5 h plus 100 min) and 8.25 h per 384-well plate (5.75 h plus 150 min), with about 55-60 min of hands-on time for 384-well. The page prints no total.
Agena Bioscience: MassARRAY System product page (stage times added by this site), agenabio.com/products/massarray-system/
Checked on 24 November 2026Derived
[5]
WFL-07One university core facility's own submission rules, not Agena specifications: elute DNA in water rather than TE because the instrument is salt-sensitive; 260/280 about 1.7-2.0; 1 uL at 10 ng/uL per reaction; at least two open wells per 384-well plate for negative controls; up to 36-plex (25-30 typical); amplicons about 100 bp; three unlabeled primers of 15-25 bases per SNP; no known polymorphism within 20 bp of the SNP.
SNP genotyping by MassARRAY (university genetics core resource page); Methods in Molecular Biology chapter on iPLEX Gold genotyping (2011), cores.arizona.edu/facilities/azgc/resources/snp-genotyping-mass-array
Checked on 24 November 2026Third partylow confidence
[6]
WFL-04Agena quotes roughly 10 to 40 ng of extracted nucleic acid; published workflows use 5-10 ng (iPLEX HS, OncoFOCUS v3.0), 10 ng (UltraSEEK breast panel) and 40 ng (UltraSEEK Oncogene panel, 2014), so input depends on the panel and is not one platform-wide range.
Agena Bioscience: MassARRAY System product page; PubMed Central article PMC9022865; Agena UltraSEEK Oncogene panel release (News-Medical, 31 May 2014), agenabio.com/products/massarray-system/
Checked on 24 November 2026Manufacturer-reportedmedium confidence
[7]
CHM-02Agena states the iPLEX Pro read window as about 4,500 to 9,000 Da and says the system distinguishes analytes 16 Da apart; a 2006 review gives gaps of at least 16 Da between the four terminator products and the same window.
Agena Bioscience: Understanding Different MassARRAY Chemistries; 2006 review of MALDI-TOF DNA analysis (PubMed Central PMC1523240), agenabio.com/resources/articles/understanding-different-massarray-chemistries/
Checked on 24 November 2026Manufacturer-reported
[8]
CHM-01iPLEX is locus-specific PCR, then shrimp alkaline phosphatase treatment to neutralize leftover nucleotides, then single-base extension of a primer annealing directly next to the SNP using mass-modified dideoxynucleotide terminators, then MALDI-TOF. The mass of the extended primer shows which allele was present.
Checked on 24 November 2026Peer-reviewed
[9]
WFL-05The Chip Prep Module (announced 24 August 2015) automates dispensing and desalting of samples onto SpectroCHIP Arrays and loading them onto the analyzer. It processes eight 96-pad arrays in under a day, has on-board cooling and storage for two more, runs largely unattended and is tied to the 96-well system.
Agena Bioscience introduces front-end automation for its MassARRAY System for genetic analysis (press release, 24 August 2015); Technology Networks report, agenabio.com/company/press-release/agena-bioscience-introduces-front-end-automation-for-its-massarray-system-for-genetic-analysis/
Checked on 24 November 2026Manufacturer-reported
[10]
WFL-14Agena's 24 August 2015 Chip Prep Module release says the module is designed to reduce cross-contamination from manual sample handling and lets the system run largely unattended. Manufacturer claim from 2015; no performance figure is given.
Agena Bioscience introduces front-end automation for its MassARRAY System for genetic analysis (press release, 24 August 2015), agenabio.com/company/press-release/agena-bioscience-introduces-front-end-automation-for-its-massarray-system-for-genetic-analysis/
Checked on 24 November 2026Manufacturer-reportedmedium confidence
[11]
WAR-03The warranty page makes these customer responsibilities: periodic maintenance and calibration, replacing hardware components such as dispense pins, liquid-handler tips and clean resin tools, data backup, networking and waste disposal. Lasers are excepted, so laser replacement is not a customer task.
Agena Bioscience: System Warranty, agenabio.com/support/system-warranty/
Checked on 24 November 2026Manufacturer-reportedmedium confidence
[12]
DAT-10Published validations describe typical QC practice: reserved negative-control wells, reference samples of known genotype, duplicates to check discordance, an identity or sex marker, and second-person review. Acceptance numbers come from each laboratory's own validation; no numeric limit is published as a standard.
Checked on 24 November 2026Peer-reviewedmedium confidence
[13]
DIA-02Salt adducts and incomplete desalting can distort spectra and cause wrong calls (published patent, see CHM-05). Agena describes salts being removed before spotting, with desalting automated by the Chip Prep Module.
Checked on 24 November 2026Third partymedium confidence
[14]
WFL-06Agena application notes from 2015 and 2017 say iPLEX genotyping runs in 24-, 96- or 384-well formats with up to 40 SNPs per well in about one day, and that the platform can interrogate 10 to 400 SNPs per sample, handle a few to hundreds of samples per day, take blood, plasma, buccal swab and FFPE material and work from as little as 5 ng DNA. Plex limits depend on the chemistry.
Single nucleotide polymorphism detection with the iPLEX assay and the MassARRAY System (Agena application note, SelectScience 2015 and The Pathologist 2017), selectscience.net/resource/single-nucleotide-polymorphism-detection-with-the-iplex-r-assay-and-the-massarray-r-system
Checked on 24 November 2026Manufacturer-reportedmedium confidence
[15]
WFL-13Mesa Labs' annual report names MassARRAY applications as inherited disease, pharmacogenetics, oncology, infectious disease and doping/toxicology, with customers mainly clinical research laboratories and contract research organizations; stated capacity is up to 50 DNA variants per PCR reaction and up to 384 samples per SpectroCHIP array.
Mesa Laboratories, Inc. Form 10-K for the fiscal year ended 31 March 2025, sec.gov/Archives/edgar/data/724004/000143774925018586/mlab20250331_10k.htm
Checked on 24 November 2026Manufacturer-reportedmedium confidence
[16]
WFL-12Medium-throughput iPLEX Gold is described as suited to up to 36 markers in 384 individuals, using locus-specific PCR, shrimp alkaline phosphatase treatment, single-base extension with mass-modified terminators and MALDI-TOF on pre-spotted SpectroCHIP arrays (2011 book chapter, Sequenom era).
Checked on 24 November 2026Peer-reviewedmedium confidence
[17]
MAN-04The MassARRAY 24-Well System was released on 24 February 2015, in a 24-well format upgradable to 96 or 384 wells. Agena's current system page lists only 96- and 384-well formats.
Agena releases MassARRAY 24-Well System for mass spectrometry-based genetic testing (News-Medical, 24 February 2015), news-medical.net/news/20150224/Agena-releases-MassARRAY-24-Well-System-for-mass-spectrometry-based-genetic-testing.aspx
Checked on 24 November 2026Manufacturer-reportedmedium confidence
[18]
WFL-10A PLoS ONE study (174 SNPs, iPLEX Gold, 17- to 36-plex) found freshly prepared whole-genome-amplified DNA performed like genomic DNA, while WGA DNA stored frozen for 18 months gave worse efficiency and accuracy.
Checked on 24 November 2026Peer-reviewedmedium confidence
[19]
CHM-09Published EpiTYPER protocol: bisulfite conversion, PCR with a T7 promoter tag, shrimp alkaline phosphatase, T7 transcription, RNase A cleavage after U positions (T-reaction), a clean-up step, then MALDI-TOF. A methylated CpG unit gives a fragment about 16 Da heavier; percent methylation is the methylated peak area over methylated plus unmethylated area.
Checked on 24 November 2026Peer-reviewedmedium confidence
[20]
WFL-08A 2019 open-access study of Staphylococcus aureus isolates used Assay Design Suite v2.0, a 10-nucleotide tag on primary PCR primers to move unextended primers outside the 4500-9000 Da window, and extra bases on some extension primers to keep allele peaks apart. Per-SNP call rates ran 81% to 98.3% (about 89% average over nine usable SNPs), and the listed steps add up to 11.25 h.
Open-access S. aureus MassARRAY genotyping study (PubMed Central PMC7471782) (external)
Checked on 24 November 2026Peer-reviewedmedium confidence
[21]
CHM-10EpiTYPER examples: an unmethylated fragment at 4484.85 Da and the same fragment with two methylated CpGs at 4516.84 Da (difference 31.99 Da, about 2 x 16); amplicons 100-600 bp (typically 250-450); default analysis window 1.5-7 kDa; 126 triplicate measurements per 384-well plate; example QC cut-offs of 0.1 replicate SD and 80% CpG-unit success. Window and thresholds are lab or software defaults.
Checked on 24 November 2026Peer-reviewedmedium confidence
[22]
DAT-07Typer naming: Sequenom announced TYPER 4.0 on 18 September 2007 alongside iPLEX Gold. Papers print Typer software, SpectroTYPER v4.0, Typer v4.0.26.74, MassARRAY Typer Analyzer v4.0.26.73 and TyperAnalyzer, which this site treats as aliases.
Sequenom launches TYPER 4.0 genotyping software (BioSpace); PubMed Central article PMC9022865, biospace.com/sequenom-inc-launches-typer-4-0-genotyping-software
Checked on 24 November 2026Third partymedium confidence
[23]
WFL-11The SNPflow QC tool for iPLEX output computes call rate, expected versus observed replicates, minor allele frequency, discordant replicates, discordance rate and a Hardy-Weinberg test, checks genotype frequencies against reference data and has a sex-determination SNP module. It sets no numeric cut-offs.
Checked on 24 November 2026Peer-reviewedmedium confidence
Research-use context. No diagnostic or clinical claims. Read the research-use notice