Workflow and data
Tools
Browser tools that run on your device: a plate and chip layout builder, a run-time planner and a floor-plan sketch for contamination control.
- Plate and chip layoutPlan a plate map with controls and duplicates, then export it to CSV or print it.
- Run-time plannerAdd up stage times for a run, with the arithmetic shown so that you can check it.
- Floor-plan sketchKeep pre- and post-amplification work apart: one-way flow, controls, and a floor-plan sketch tool.
Plate and chip layout
Plan a plate map with controls and duplicates, then export it to CSV or print it.
A MassARRAY run starts from a 96-well or 384-well reaction plate and ends on a chip. If the plate map is not the same at every step, a good run can still give wrong answers. The map below is for planning and for the run log; it keeps nothing outside your browser.
Plate formats: the 24-well system was released in February 2015, and the manufacturer’s current system page lists 96- and 384-well formats only.[1]
The two grids
96-well grid, rows A to H and columns 1 to 12, with a generic chip view. Example only; reserve and number controls according to your SOP.
Text version of this figure
A grid of 8 rows (A to H) and 12 columns (1 to 12), with legend tiles for sample, negative control, reference, duplicate, identity or sex marker and empty wells.
384-well grid, rows A to P and columns 1 to 24. Generic grid, not to scale; the pad geometry of a specific chip is not stated here.
Text version of this figure
A grid of 16 rows (A to P) and 24 columns (1 to 24), with the same legend.
What to reserve
Published work describes the same set of reserved positions. No numeric rule is published as a standard: how many of each you need comes from your own validation.
| Reserve | What it tells you | Where it is described |
|---|---|---|
| Negative-control (no-template) wells | Whether signal appears with no DNA added, which points to carry-over or a reagent problem. | Published validations[2] |
| Positive or reference sample of known genotype | Whether the run reproduces a result you already trust. | Published validations[2] |
| Duplicates | Whether the same sample gives the same call twice. | Published validations[2], SNPflow discordance checks[3] |
| Identity or sex marker | Whether the sample on the plate is the sample you meant. | Published validations[2], SNPflow sex-determination module[3] |
| Empty wells | Open positions you can use for extra negative controls or a rerun. | One core facility's rule: at least two open wells per 384-well plate[7] |
One map from PCR to chip
Draw the plate map once, before the run, and keep it unchanged. The PCR plate, the extension plate, the chip pads and the software layout should all read from it. Any change after pipetting starts goes in the run log with the reason.
- Write the plate orientation on the map: which corner is A1 on the physical plate, and how the plate goes onto the dispenser or into the chip-prep unit.
- Put controls where a local problem would show: more than one negative control, spread across the plate.
- When a plate is split across chips or a chip carries more than one plate, number each chip and keep the mapping from plate position to chip pad on the same sheet.
The map feeds the run log template, and the run record SOP page shows where the map is checked.
Plate map builder
Stays in this browser.
Click or drag across wells to assign the selected type. With the keyboard, use the arrow keys to move and Enter or Space to assign. Row and column headers assign a whole row or column.
- Sample
- Negative (no-template) control
- Positive or reference sample
- Duplicate
- Identity or sex marker
- Empty
Category check
Not assigned yet on this plate:
- Sample
- Negative (no-template) control
- Positive or reference sample
- Duplicate
- Identity or sex marker
A missing category is a prompt to check your SOP, not an error.
Run-time planner
Add up stage times for a run, with the arithmetic shown so that you can check it.
Choose a plate format and the number of plates. The planner adds the manufacturer’s published stage times and shows each step of the sum, so that you can check it against the source and against your own experience.
Manufacturer-reported (2026) For a 96-well plate the manufacturer gives about 5.5 h for PCR, SAP and extension (about 23 min hands-on) and about 100 min for detection (about 5 min hands-on). For a 384-well plate it gives about 5.75 h (about 45 min) and about 150 min (about 5 min). Data analysis takes 5 to 10 min of hands-on time. These times vary with chemistry and acquisition settings.[4]
The manufacturer prints no total. Adding its stage times gives about 7.2 h per 96-well plate and 8.25 h per 384-well plate, with about 55 to 60 min of hands-on time for 384-well.[5]
Plan a run
Stage times are the manufacturer’s published values.
Plan: 1 × 96-well plate, run one after another
1. Time on the instruments, one plate
5.50 h + 100 min (1.67 h) = 7.17 h = 7 h 10 min
PCR, SAP and extension, then detection. Published times; the manufacturer prints no total, so the sum is ours.
2. Hands-on time, one plate
23 + 5 + 5 to 10 = 33 min to 38 min
PCR, SAP and extension, plus detection, plus data analysis, in minutes.
3. 1 plate, no overlap assumed
1 × 7.17 h = 7.17 h = 7 h 10 min
Hands-on: 1 × (33 min to 38 min) = 33 min to 38 min
| Plate | Starts | PCR, SAP and extension end | Detection ends |
|---|---|---|---|
| 1 | +0 min | +5 h 30 min | +7 h 10 min |
Times vary with chemistry and acquisition settings. This plan leaves out reagent preparation, instrument readiness, waiting between steps and any overlap between plates, so it describes plates that run back to back without a pause.
What the plan leaves out
- Reagent preparation, plate sealing and spin steps, and any wait before you start a stage.
- Instrument readiness, calibration and performance verification under your own SOP.
- Overlap between plates. The planner runs plates back to back and assumes none. Whether you can overlap them depends on your instruments and staffing, and the manufacturer’s stage times come with no rule about it.
- The time to review and sign off results, which is on top of the 5 to 10 min of hands-on analysis the manufacturer gives.
Planning a batch also depends on how many plates your chip-prep unit or dispenser can take in a day. Check that against your manufacturer documents.
The stages are explained on the stages page, and your own times can go straight into the planner with the checkbox on the left. Keep your own times in the run log so the next plan starts from evidence.
The planner runs in your browser. It saves nothing and sends nothing.
Where the time goes
The workflow swimlane with the published stage times.
Text version of this figure
- Stage 0: Intake and plate map
- Stage 1: DNA input and QC
- Stage 2: Assay design
- Stage 3: Multiplex PCR
- Stage 4: SAP treatment
- Stage 5: Extension
- Stage 6: Resin clean-up
- Stage 7: Dispense onto chip
- Stage 8: MALDI-TOF acquisition
- Stage 9: Typer analysis and review
- Stage 10: Reporting and archive
Floor-plan sketch
Keep pre- and post-amplification work apart: one-way flow, controls, and a floor-plan sketch tool.
Drag the three areas and the start marker to match your space, then read the check. You can export the sketch as a PNG for a procedure or a request to us.
General good practice, not a manufacturer requirement.
Drag the areas and the start marker, or choose one below and use the move buttons. With the keyboard, tab to an area and use the arrow keys; hold Shift to move farther.
4 of 4 checks pass.
One-way flow check
- Pass: The three areas do not overlap.
- Pass: Each area lies farther from the start than the one before it, so the path runs one way.
- Pass: The post-amplification area is farther from the reagent area than the sample set-up area is.
- Pass: Everything sits inside the plan.
Simple geometry only. It does not assess airflow, doors, people movement or equipment, and it states no pressure values.
The guidance behind the check is in contamination control.
Sources
[1]
MAN-04The MassARRAY 24-Well System was released on 24 February 2015, in a 24-well format upgradable to 96 or 384 wells. Agena's current system page lists only 96- and 384-well formats.
Agena releases MassARRAY 24-Well System for mass spectrometry-based genetic testing (News-Medical, 24 February 2015), news-medical.net/news/20150224/Agena-releases-MassARRAY-24-Well-System-for-mass-spectrometry-based-genetic-testing.aspx
Checked on 24 November 2026Manufacturer-reportedmedium confidence
[2]
DAT-10Published validations describe typical QC practice: reserved negative-control wells, reference samples of known genotype, duplicates to check discordance, an identity or sex marker, and second-person review. Acceptance numbers come from each laboratory's own validation; no numeric limit is published as a standard.
Checked on 24 November 2026Peer-reviewedmedium confidence
[3]
WFL-11The SNPflow QC tool for iPLEX output computes call rate, expected versus observed replicates, minor allele frequency, discordant replicates, discordance rate and a Hardy-Weinberg test, checks genotype frequencies against reference data and has a sex-determination SNP module. It sets no numeric cut-offs.
Checked on 24 November 2026Peer-reviewedmedium confidence
[4]
WFL-02Agena's stage times, which vary with chemistry and acquisition settings: 96-well PCR, SAP and extension about 5.5 h (about 23 min hands-on) and detection about 100 min (about 5 min hands-on); 384-well about 5.75 h (about 45 min hands-on) and detection about 150 min (about 5 min hands-on); data analysis 5-10 min hands-on.
Agena Bioscience: MassARRAY System product page, agenabio.com/products/massarray-system/
Checked on 24 November 2026Manufacturer-reportedmedium confidence
[5]
WFL-03Adding Agena's stage times gives about 7.2 h of run time per 96-well plate (5.5 h plus 100 min) and 8.25 h per 384-well plate (5.75 h plus 150 min), with about 55-60 min of hands-on time for 384-well. The page prints no total.
Agena Bioscience: MassARRAY System product page (stage times added by this site), agenabio.com/products/massarray-system/
Checked on 24 November 2026Derived
[6]
WFL-05The Chip Prep Module (announced 24 August 2015) automates dispensing and desalting of samples onto SpectroCHIP Arrays and loading them onto the analyzer. It processes eight 96-pad arrays in under a day, has on-board cooling and storage for two more, runs largely unattended and is tied to the 96-well system.
Agena Bioscience introduces front-end automation for its MassARRAY System for genetic analysis (press release, 24 August 2015); Technology Networks report, agenabio.com/company/press-release/agena-bioscience-introduces-front-end-automation-for-its-massarray-system-for-genetic-analysis/
Checked on 24 November 2026Manufacturer-reported
[7]
WFL-07One university core facility's own submission rules, not Agena specifications: elute DNA in water rather than TE because the instrument is salt-sensitive; 260/280 about 1.7-2.0; 1 uL at 10 ng/uL per reaction; at least two open wells per 384-well plate for negative controls; up to 36-plex (25-30 typical); amplicons about 100 bp; three unlabeled primers of 15-25 bases per SNP; no known polymorphism within 20 bp of the SNP.
SNP genotyping by MassARRAY (university genetics core resource page); Methods in Molecular Biology chapter on iPLEX Gold genotyping (2011), cores.arizona.edu/facilities/azgc/resources/snp-genotyping-mass-array
Checked on 24 November 2026Third partylow confidence
Research-use context. No diagnostic or clinical claims. Read the research-use notice