Validation and SOPs

Maintenance, training and receiving

Outlines for the procedures that sit beside the assay SOP: maintenance and troubleshooting, training and competency, and receiving and lot-change checks, with no acceptance criteria.

Checked on 24 November 2026
On this page

The assay procedure, the run record and change control are on SOPs.

Maintenance and troubleshooting procedure

An outline for your own maintenance and troubleshooting procedure, filled from your manufacturer's documents.

Who does what

The most useful page of this SOP is a single table that says who may do each task. Everything else hangs from it.

The manufacturer’s warranty names periodic maintenance and calibration, replacement of named hardware components, data backup, networking and waste disposal as customer responsibilities, and excepts the laser.[1] Your SOP copies the assignment from your own manufacturer guide, not from this table.

Responsibility matrix for maintenance and troubleshooting
Task or areaWho does itThe general reason
Periodic maintenance and calibration the manufacturer's guide assigns to the customerOperator or named role, as your SOP saysThe manufacturer's warranty lists these among customer responsibilities.
Replacement of customer-side hardware components the guide namesOperator or named role, as your SOP saysNamed in the same warranty list; the guide says which components.
Data backupOperator or named role, as your SOP saysA customer responsibility in the warranty, and the part that saves you after a failure.
NetworkingOperator or named role, as your SOP saysA customer responsibility in the warranty.
Waste handlingOperator or named role, as your SOP saysA customer responsibility in the warranty, under your biosafety rules.
Work inside coversEngineer onlyCovers and interlocks guard against injury and against damage to the instrument.
Anything involving the laser, high voltage or vacuum partsEngineer onlyThe instrument contains a laser, high-voltage electronics and a vacuum system, and each carries a risk of injury.
Any other work inside the instrumentEngineer onlyInternal parts can injure people and are easily damaged, so none of it is a customer task.

Covers stay closed. Work inside an instrument is for a qualified service engineer only. Safety statement

The task table, with the intervals left blank

Tasks and intervals come from your manufacturer's current guide. Copy them in, name the guide and its version, and say who does each one.

Maintenance task table
Task (from your guide)Interval (from your guide)Performed byRecord locationNotesRow actions

Your entries are kept in this browser only. Nothing is sent to us.

One example framework has a laboratory perform and document the manufacturer’s maintenance and function checks at least as often as the manufacturer specifies.[2] The manufacturer’s service plans also include preventive maintenance in their lower tier.[3] Neither source gives the tasks, and this site does not guess at them.

Escalation: what you check, who checks, when you stop

Write the first check from your manufacturer's guide, who does it, what to record, and the point at which you hand the problem on.

Escalation table
SymptomFirst check from your manufacturer's guideWho checksWhat to recordEscalate toWhen to escalate (your rule)Row actions

Your entries are kept in this browser only. Nothing is sent to us.

Safety, records and backup

Refer to your site biosafety procedure and the safety data sheets for spills. Keep the maintenance log, the service reports and the backup record where an assessor can find them in minutes, and name who checks that backups can be restored.

See also the safety statement and the T-05 preventive-maintenance log.

Get the outline

Training and competency record

A structure for recording who has been trained on what, and when competency was last reviewed.

Trained is not the same as competent

A training record shows someone was taught. A competency record shows someone was later seen to do it right. Keep both, and keep them apart.

The record is per person and per topic. Each row says what was taught, who taught it, how competence was assessed, who assessed, what the criteria were and when it is next assessed. The criteria are set by your laboratory, so the template leaves that column blank.

Topics to cover

A starting list. Add your own, and split any topic that is too big to assess in one go.

  • Instrument operation, as the manufacturer's guide describes it
  • Contamination control
  • Data review
  • Safety and waste
  • The SOP set: each procedure the person works to, read and understood

For instrument topics, the manufacturer’s warranty names periodic maintenance and calibration, replacement of named hardware components, data backup, networking and waste disposal as customer responsibilities, so these are the tasks your operators need to be competent at.[1] Work inside the instrument is not an operator topic. It is for a qualified service engineer.

Three ways to assess

Use more than one for a task that matters.

Assessment methods
MethodWhat it showsSuits
Observed practiceThe person performs the task correctly, in front of an assessor, with the real equipment.Hands-on tasks: plate set-up, handling consumables, starting a run.
Written assessmentThe person knows the rules and the reasons: what a control is for, what to do when one fails.Safety, contamination control and procedure knowledge.
Case reviewThe person reads a past run, plot or spectrum and reaches the decision your SOP expects.Data review and interpretation.

Name the assessor on every row. Your SOP decides whether a trainer may also assess and how you handle that.

Fill in a record

One row per person per topic. Export it as a CSV, or print it for the file.

Training and competency record
TraineeTopicDateTrainerAssessment methodAssessorCriteria (set by your laboratory)OutcomeNext re-assessment dueRow actions

Your entries are kept in this browser only. Nothing is sent to us.

Receiving and lot-change checks

A checklist for receiving consumables and for the checks to run when a lot changes.

At the door: receiving checks

Most lot problems are found at receiving or at the first run. Receiving is the cheaper place to find them.

Receiving checklist
ItemSupplierLotExpiryReceived onReceived byPackaging intactLabel and lot recordedStored as the supplier's insert statesIn quarantine until checkedRow actions

Your entries are kept in this browser only. Nothing is sent to us.

Storage conditions come from the supplier’s insert for each item, not from this page. Anything that fails a receiving check stays in quarantine until someone with the authority your SOP names decides what happens to it.

When the lot changes

A new lot is a change. The check is proportionate: your SOP says how much.

  1. Step 1: Prepare a parallel run

    Run the new lot beside the current one with known controls, as your SOP describes.
  2. Step 2: Take the criteria from your SOP and the insert

    Write them down before the run. This page sets none.
  3. Step 3: Record the result against the criteria

    Put the numbers and the comparison in the record, not just 'OK'.
  4. Step 4: Decide

    The person your SOP names decides whether the lot enters use, and signs.
  5. Step 5: Link to a change record

    If your procedure treats a lot change as a change, open the record and cross-reference it.

One example framework sets minimum daily controls for qualitative and molecular tests, so a lot check normally runs the same kind of controls your routine runs use.[4] Which controls, in what number and against which limits is for your SOP and your director to say.

Records and links

A lot change usually belongs in change control. Reagent and lot logs also appear as an appendix to the assay procedure, and lot numbers go on every run record.

General information; not legal, regulatory or clinical advice.

Sources

  1. [1]
    WAR-03

    The warranty page makes these customer responsibilities: periodic maintenance and calibration, replacing hardware components such as dispense pins, liquid-handler tips and clean resin tools, data backup, networking and waste disposal. Lasers are excepted, so laser replacement is not a customer task.

    Agena Bioscience: System Warranty, agenabio.com/support/system-warranty/

    Checked on 24 November 2026Manufacturer-reportedmedium confidence

  2. [2]
    LAB-06

    CLIA 42 CFR 493.1254: for unmodified manufacturer-supported equipment the laboratory performs and documents the manufacturer's maintenance and function checks at least as often as the manufacturer specifies, with results within the manufacturer's limits before patient testing; modified or non-manufacturer equipment needs the laboratory's own written protocols.

    42 CFR 493.1254 Standard: Maintenance and function checks (eCFR) (external)

    Checked on 24 November 2026Regulator

  3. [3]
    SUP-01

    Agena's service-plan page lists an Essential tier (preventive maintenance, on-site applications support, fast response) and a Plus tier that adds applications training, monthly performance verification kits, next-day response and qualification services. Both cover the system and the Chip Prep Module; no prices, hours or definitions are given.

    Agena Bioscience: Service Plans, agenabio.com/support/service-plans/

    Checked on 24 November 2026Manufacturer-reportedmedium confidence

  4. [4]
    LAB-05

    CLIA 42 CFR 493.1256: daily minimum controls are one positive and one negative for qualitative tests, two controls for tests with an extraction step (one able to detect extraction errors), and for molecular amplification two controls plus an inhibition control where inhibition is a significant cause of false negatives.

    42 CFR 493.1256 Standard: Control procedures (eCFR) (external)

    Checked on 24 November 2026Regulator